Published online Jun 15, 2024. doi: 10.4239/wjd.v15.i6.1254
Revised: February 2, 2024
Accepted: March 19, 2024
Published online: June 15, 2024
Processing time: 179 Days and 6.3 Hours
The FreeStyle Libre flash glucose monitoring (FGM) system entered the Chinese market in 2017 to complement the self-monitoring of blood glucose. Due to its increased usage in clinics, the number of studies investigating its accuracy has increased. However, its accuracy has not been investigated in highland popu-lations in China.
To evaluate measurements recorded using the FreeStyle Libre FGM system compared with capillary blood glucose measured using the enzyme electrode method in patients with type 2 diabetes (T2D) who had migrated within 3 mo from highlands to plains.
Overall, 68 patients with T2D, selected from those who had recently migrated from highlands to plains (within 3 mo), were hospitalized at the Department of Endocrinology from August to October 2017 and underwent continuous glucose monitoring (CGM) with the FreeStyle Libre FGM system for 14 d. Throughout the study period, fingertip capillary blood glucose was measured daily using the enzyme electrode method (Super GL, China), and blood glucose levels were read from the scanning probe during fasting and 2 h after all three meals. Moreover, the time interval between reading the data from the scanning probe and collecting fingertip capillary blood was controlled to < 5 min. The accuracy of the FGM system was evaluated according to the CGM guidelines. Subsequently, the factors influencing the mean absolute relative difference (MARD) of this system were analyzed by a multiple linear regression method.
Pearson’s correlation analysis showed that the fingertip and scanned glucose levels were positively correlated (R = 0.86, P = 0.00). The aggregated MARD of scanned glucose was 14.28 ± 13.40%. Parker's error analysis showed that 99.30% of the data pairs were located in areas A and B. According to the probe wear time of the FreeStyle Libre FGM system, MARD1 d and MARD2–14 d were 16.55% and 14.35%, respectively (t = 1.23, P = 0.22). Multiple stepwise regression analysis showed that MARD did not correlate with blood glucose when the largest amplitude of glycemic excursion (LAGE) was < 5.80 mmol/L but negatively correlated with blood glucose when the LAGE was ≥ 5.80 mmol/L.
The FreeStyle Libre FGM system has good accuracy in patients with T2D who had recently migrated from highlands to plains. This system might be ideal for avoiding the effects of high hematocrit on blood glucose monitoring in populations that recently migrated to plains. MARD is mainly influenced by glucose levels and fluctuations, and the accuracy of the system is higher when the blood glucose fluctuation is small. In case of higher blood glucose level fluctuations, deviation in the scanned glucose levels is the highest at extremely low blood glucose levels.
Core Tip: Although blood glucose monitoring is the standard and most established approach for informing diabetes therapy, continuous glucose monitoring has seen a rapid increase in the number of users in recent years. Abnormal glucose metabolism in highland populations is a major health concern worldwide. This study focused on the accuracy of the FreeStyle Libre FGM system in highland populations.