Table 1 Patients’ clinical characteristics
| Clinical characteristic | n | % |
| Mean age (yr) | 62.8 ± 9.5 | |
| Male | 104 | 77.0 |
| Female | 31 | 23.0 |
| BMI (m/kg2) | 26.2 ± 2.9 | |
| Creatinine/(μmol/L) | ||
| Previous | 77.5 ± 27.2 | |
| Postoperation | 82.8 ± 30.1 | |
| Risk factors | ||
| Hypertension | 102 | 75.6 |
| Diabetes mellitus | 59 | 43.7 |
| Hypercholesterolemia | 14 | 10.4 |
| Angiography indication | ||
| Unstable angina | 94 | 69.6 |
| STEMI | 25 | 18.5 |
| NSETMI | 16 | 11.9 |
Table 2 Lesion position and selection of guide catheter
| Target vessel | n (%) | Type of GC | n (%) |
| RCA | 58 (43.0) | JR | 52 (38.5) |
| SAL | 6 (4.4) | ||
| LAD | 45 (33.3) | JL | 9 (6.7) |
| EBU | 36 (26.7) | ||
| LCX | 32 (23.7) | JL | 5 (3.7) |
| EBU | 27 (20.0) |
Table 3 Summary of procedural characteristics
| Procedural characteristic | n (%) or mean ± SD |
| Procedural results | |
| Technique success | 128 (94.8) |
| PCI success | 125 (92.6) |
| Stent implantation | 124 (91.9) |
| DCB therapy | 1 (0.7) |
| Artery path | |
| Right radial artery | 128 (94.8) |
| Left radial artery | 5 (3.7) |
| Distal radial artery | 2 (1.5) |
| Indication for using a rapid exchange extension catheter | |
| Difficult balloon delivery | 35 (25.9) |
| Difficult stent delivery | 96 (71.1) |
| Difficult posterior expansion balloon | 4 (3.0) |
| Preprocess lesion before delivering the rapid exchange catheter | |
| Multiple predilatation | 77 (57.0) |
| Buddy wire technique | 45 (33.4) |
| Balloon anchoring technique | 8 (5.9) |
| Cutting balloon modification | 5 (3.7) |
| Depth of extension catheter intubation (mm) | 25.3 ± 15.4 |
| Rapid exchange catheter service time1 (min) | 18.9 ± 9.7 |
| PCI operation time2 (min) | 35.8 ± 16.6 |
| Mean length of stents (mm) | 33.5 ± 14.4 |
| Mean diameter of stents (mm) | 2.75 ± 0.45 |
| Length of DCB (mm) | 30.0 |
| Diameter of DCB (mm) | 3.0 |
| Contrast dose (mL) | 162.2 ± 38.3 |
| Perioperative dual antiplatelet | |
| Aspirin + clopidogrel | 77 (57.0) |
| Aspirin + ticagrelor | 58 (43.0) |
| Perioperative anticoagulant | |
| Unfractionated heparin | 98 (72.6) |
| Bivalirudin | 37 (27.4) |
Table 4 Adverse events and major adverse cardiovascular events
| Adverse event | n (%) |
| Intraoperative complication | |
| Perforation | 0 |
| Dissection | 9 (6.7) |
| Thrombosis | 0 |
| Air embolism | 0 |
| Pressure damping | 27 (20.0) |
| Stent deformation or stripping | 6 (4.4) |
| Extension catheter fracture | 0 |
| Perioperative complication | |
| Bleeding | 5 (3.7) |
| Major bleeding | 0 |
| Stent thrombosis | 0 |
| MACEs during hospitalization | |
| Recurrence of myocardial infarction | 0 |
| Repeated revascularization | 0 |
| All-cause death | 0 |
| MACEs in the one-month follow-up | |
| Recurrence of myocardial infarction | 0 |
| Repeated revascularization | 0 |
| All-cause death | 0 |
- Citation: Wang HC, Lu W, Gao ZH, Xie YN, Hao J, Liu JM. Application of a rapid exchange extension catheter technique in type B2/C nonocclusive coronary intervention via a transradial approach. World J Clin Cases 2021; 9(12): 2751-2762