Case Report
Copyright ©The Author(s) 2022.
World J Clin Cases. May 16, 2022; 10(14): 4594-4600
Published online May 16, 2022. doi: 10.12998/wjcc.v10.i14.4594
Figure 1
Figure 1 The residual end is repaired with an intravenous needle during surgery.
Figure 2
Figure 2 Repair steps of trachea duct inflatable tube fracture. A: Rupture of the endotracheal tube inflatable tube was found; B: Retraction of the needle by 1 mm prior to its insertion; C: A 22G indwelling needle is inserted into the cut end of the inflatable tube and the needle is removed. Sealing the injection end of the needle with a valve plug; D: Sealing the injection end of the needle with a valve plug; E: Use a pressure gauge to adjust the pressure inside the cuff; F: Marking and fixing the repaired end.
Figure 3
Figure 3 The method to repair the inflatable tube that was completely detached from it is root.
Figure 4
Figure 4 Protective measure should be taken by properly attaching the pilot balloon assembly to the tracheal tube wall.