BPG is committed to discovery and dissemination of knowledge
Retrospective Study
Copyright: ©Author(s) 2026.
World J Clin Oncol. Apr 24, 2026; 17(4): 115851
Published online Apr 24, 2026. doi: 10.5306/wjco.v17.i4.115851
Table 1 Control limits, tolerance limits, and action limits for Individual-chart and Exponentially Weighted Moving Average analysis of patient-specific quality assurance gamma passing rates in head-and-neck radiotherapy plans across four gamma criteria (3%/3 mm, 2%/3 mm, 3%/2 mm, and 2%/2 mm)
CriteriaIndividual charts
Moving-range charts
Exponentially Weighted Moving Average charts
Tolerance limit
Action limit
Upper control limit
Central line
Lower control limit
Upper control limit
Central line
Lower control limit
Upper control limit
Central line
Lower control limit
3%, 3 mm99.80198.83597.8691.1870.363099.156998.83598.513197.869496.4844
3%, 2 mm98.43797.41596.3931.2550.384097.755697.41597.074496.393692.2299
2%, 3 mm96.595.5994.681.1180.342095.893395.5995.286794.680386.7638
2%, 2 mm95.12894.493.6720.8940.274094.642694.494.157493.671283.1989