Characteristic | Mean | SD | Minimum | Maximum |
Age (yr) | 53.20 | 8.75 | 34.00 | 81.00 |
Course (yr) | 7.59 | 5.78 | 0.17 | 30.00 |
BMI (kg/m2) | 23.81 | 2.74 | 19.00 | 32.20 |
HbA1c (%) | 9.22 | 2.40 | 5.70 | 13.80 |
Hb (g/L) | 149.68 | 15.88 | 115 | 184 |
HCT (%) | 43.99 | 4.42 | 35.5 | 54.2 |
Fingertip blood glucose level (mmol/L) | 9.78 | 3.75 | 2.08 | 24.12 |
Scanned blood glucose level (mmol/L) | 8.82 | 3.70 | 2.20 | 24.60 |
Group | n | MAD (mean ± SD, mmol/L) | MARD (mean ± SD, %) |
< 2.77 | 1 | 1.32 | 63.46 |
2.77-4.44 | 42 | 0.85 ± 0.82 | 21.85 ± 21.91 |
4.45-6.66 | 171 | 1.13 ± 0.94 | 20.56 ± 18.60 |
6.67-11.10 | 508 | 1.10 ± 0.94 | 12.72 ± 10.75 |
11.11-16.65 | 252 | 1.53 ± 1.14 | 11.66 ± 8.78 |
16.66-22.20 | 35 | 2.74 ± 3.24 | 14.71 ± 17.49 |
> 22.20 | 4 | 3.11 ± 2.33 | 13.25 ± 9.61 |
- Citation: Sun ZM, Du YZ, Wang SY, Sun SY, Ye Y, Sun XP, Li MX, He H, Long WC, Zhang CH, Yao XY, Fan WY, Wang L, Wu YH. Accuracy of FreeStyle Libre flash glucose monitoring in patients with type 2 diabetes who migrated from highlands to plains. World J Diabetes 2024; 15(6): 1254-1262