Copyright: ©Author(s) 2027.
World J Gastroenterol. Jan 7, 2027; 33(1): 124775
Published online Jan 7, 2027. doi: 10.3748/wjg.124775
Published online Jan 7, 2027. doi: 10.3748/wjg.124775
Table 1 Parameter comparison of mainstream biodegradable pancreatic and biliary stents
| Stent (platform) | Material and design | Diameter | Length | Degradation cycle | Approval scope | Principal indications |
| ARCHIMEDES BPS | PDX (fast and medium variants); alternative polymer (slow variant); helical, dual-channel | 6-10 Fr | 40-125 mm | Fast approximately 12 days; medium approximately 20 days; slow approximately 11 weeks | CE-marked (Europe); FDA 510(k) (K243412), July 2025, for pancreatic duct drainage | PEP prophylaxis; pancreatic drainage; benign biliary and cystic-duct leaks |
| UNITY-B | Balloon-expandable magnesium alloy, polymer-coated | 8-10 mm | Manufacturer range | Approximately 1-6 months | CE-marked (Europe); investigational in United States | Benign biliary strictures |
| PDX biliary stent (e.g., SX-ELLA-BD)1 | PDX monofilament, self-expanding | Approximately 8-10 mm | 30-80 mm | approximately 6 months to full resorption | CE-marked (Europe) | Benign biliary strictures; percutaneous and endoscopic use |
Table 2 Summary of existing clinical studies across the four major indications
| Ref. | Design | n | Stent type | Key results | Notes |
| PEP prophylaxis | |||||
| Anderloni et al[10], 2020 | Prospective pilot | 38 (9 for PEP) | ARCHIMEDES | 100% technical success | 1 mild PEP; no cholangitis |
| Pérez-Cuadrado Robles et al[15], 2022 | Prospective multicenter feasibility | 53 | ARCHIMEDES | 100% technical; 77.9% clinical success | All AEs mild |
| Maieron et al[16], 2023 | Observational multicenter (Austria) | 60 patients/63 stents | Biodegradable | 94% technical; 90% clinical; no follow-up in 97% | 2 reinterventions |
| McGarrigle et al[14], 2026 | Retrospective multicenter (Australia and New Zealand) | 111 ERCPs | Biodegradable pancreatic | 98% technical; PEP 11% (all mild); AUD $393/patient saving | Preferred by 75% of endoscopists |
| Abbasi et al[17], 2026 | Health-economic model (United Kingdom NHS) | Model-based | Biodegradable pancreatic | Cost-effective; ICER approximately £9317/QALY | Model-based |
| POPF | |||||
| Bakheet et al[19], 2020 | Preclinical (porcine) | Animal | Tubular BDS | Reduced PJ anastomotic stricture | Preclinical |
| Mazzola et al[18], 2022 | Prospective technical series | High-risk PJ | Internal BDS | CR-POPF 13.3% | No significant stent-related complications |
| González-Abós et al[21], 2025 | Propensity-matched cohort | 80 high-risk | Biodegradable | 0% POPF vs 10%; P = 0.039 | Consistent matched/unmatched |
| Farooqui et al[20], 2025 | Systematic review and meta-analysis | 8 studies/155 | Biodegradable | Pooled POPF approximately 7% | Observational only |
| Farooqui et al[22], 2026 | Blinded RCT (interim) | 50 of 310 planned | ARCHIMEDES (fast) | CR-POPF 11.5% vs 25.0% (P = 0.20); biochemical leak 0% vs 12.5% (P = 0.01) | Comparable major complications/mortality |
| PSC/altered anatomy | |||||
| Ponsioen et al[25], 2018 | Multicenter RCT (balloon vs plastic) | 65 | Plastic (comparator) | Equivalent recurrence; higher AEs with stenting (45% vs 6.7%) | Harm of conventional stenting in PSC |
| Lindström et al[24], 2020 | Case series (DBE) | 7 (4 PSC) | ARCHIMEDES | 100% technical; resolution 3/3 anastomotic | 2/4 PSC recurrent cholangitis |
| Bile leaks | |||||
| Laukkarinen et al[26], 2007 | Prospective animal study | Animal | Biodegradable biliary | Reduced drainage; earlier drain removal; no retrieval | Preclinical |
| Siiki et al[23], 2018 | Comparative cohort (3.5 years) | 32 (8 vs 24) | PDX | No repeat endoscopy in BDS; lower total output | Similar 30-day AEs |
| Mohanka et al[29], 2021 | Prospective series (liver transplant) | 20 | ARCHIMEDES | No bile leaks/bilomas/stent-related complications | 1 cholangitis (migration) |
| Salamanca-Bustos et al[30], 2016 | Randomized pilot (LT) | 10 | Biodegradable | 100% patency; no leaks; complete resorption | First RCT pilot |
| Lopez-Lopez et al[31], 2026 | Comparative cohort (LT) | 158 | Self-expandable absorbable | Early biliary complications 2.6% vs 23.8% (P < 0.001); cost €6988 vs €36364 | Propensity-matched |
| Wu et al[28], 2025 | Case series (post-choledochotomy) | 4 | ARCHIMEDES | No complications when positioned; no pancreatitis | Learning curve |
| Venkatachalapathy et al[27], 2025 | Prospective multicenter (United Kingdom/Ireland) | Cohort | ARCHIMEDES | 17-fold QoL improvement; cost saving | Observational |
| Cross-cutting | |||||
| Chen et al[34], 2025 | Systematic review (percutaneous) | 7 studies/323 | Biodegradable biliary | Technical success 98%-100% | Hemobilia most frequent |
| Peristeri et al[35], 2026 | Systematic review (PRISMA) | 15 studies/467 | Biodegradable (mostly PDX) | Technical 99.4%; clinical 73.3% | Cholangitis most frequent; reintervention mostly planned |
| Almeida and Donato[32], 2020 | Systematic review and meta-analysis | Benign strictures | BDS vs multiple plastic | Resolution 83% vs 84%; 1 vs 3 interventions | Cholangitis 24.1% vs 6.1% |
| Dreifuss et al[33], 2025 | Systematic review and meta-analysis | 12 studies/384 | Absorbable biliary | 100% technical; 81% clinical; 3% migration | No mortality |
Table 3 Grading of Recommendations Assessment, Development and Evaluation assessment of evidence preventive role of biodegradable pancreatobiliary stents across four indications
| Indication and principal outcome | Initial certainty | Risk of bias | Inconsistency | Indirectness | Imprecision | Publication bias | Final certainty |
| PEP prophylaxis: Incidence and severity of PEP | Low | Serious | Not serious | Serious | Serious | Suspected | Very low |
| Basis: Predominantly uncontrolled or retrospective feasibility cohorts, often with mixed indications. No adequately powered randomized comparison of biodegradable vs conventional pancreatic stents or no stent has evaluated PEP incidence | |||||||
| PEP prophylaxis: Avoidance of follow-up imaging and endoscopic retrieval | Low | Not serious | Not serious | Not serious | Not serious | Not detected | Moderate (rise) large effect |
| Basis: Avoidance of surveillance imaging and endoscopic retrieval follows almost deterministically from in-situ degradation. Across cohorts essentially no biodegradable-stent recipient required retrieval, compared with repeat endoscopy in approximately two-thirds of plastic-stent patients. Rated up one level for large magnitude of effect. This outcome reflects procedural burden, not prevention of PEP itself | |||||||
| Postoperative pancreatic fistula: Clinically relevant POPF after PD | High | Not serious | Not serious | Not serious | Very serious | Not detected | Low |
| Basis: One blinded randomized trial provides direct evidence, but only 50 of the planned 310 participants were included in the interim analysis, and few events leave substantial uncertainty around the effect on clinically relevant POPF. The interim, incomplete-accrual limitation was accounted for under imprecision rather than risk of bias to avoid double-counting; risk of bias is therefore rated not serious | |||||||
| PSC-related strictures: Stricture resolution and prevention of recurrent cholangitis | Low | Serious | Not assessable | Very serious | Very serious | Suspected | Very low |
| Basis: Direct evidence is limited to four PSC patients within a seven-patient case series involving surgically altered anatomy. Evidence of harm from conventional plastic stenting does not establish biodegradable-stent safety or effectiveness | |||||||
| Bile leaks: Clinical closure without recurrence or reintervention | Low | Serious | Not serious | Serious | Serious | Suspected | Very low |
| Basis: Direct evidence for established post-cholecystectomy bile leaks is based mainly on a small comparative cohort with eight biodegradable-stent recipients. Remaining evidence largely concerns transplant prophylaxis, post-choledochotomy use, or benign strictures | |||||||
- Citation: Jan STA, Abbasi A, Menon S. Preventive role of biodegradable pancreatobiliary stents. World J Gastroenterol 2027; 33(1): 124775
- URL: https://www.wjgnet.com/1007-9327/full/v33/i1/124775.htm
- DOI: https://dx.doi.org/10.3748/wjg.124775