Efficacy and safety of disposable digital single-operator cholangioscopy: A noninferiority, multicenter, randomized, open-label, blinded endpoint trial
Yuan Xie, Zhi-Liang Tang, Department of Hepato-Biliary-Pancreatic-Splenic Surgery, The First People's Hospital of Zhaoqing, Zhaoqing 526000, Guangdong Province, China
Jin-Yu Qiu, Department of Clinical Trial Institution, Zhujiang Hospital, Southern Medical University, Guangzhou 510280, Guangdong Province, China
Min Ju, Department of Ear, Nose and Throat, The People’s Hospital of Gaoan, Gaoan 330800, Jiangxi Province, China
Shao-Ru Cai, Gao-Fei Kang, Ke-Bo Zhong, Department of Hepatobiliary Surgery, Zhujiang Hospital, Southern Medical University, Guangzhou 510280, Guangdong Province, China
Pi Guo, Department of Preventive Medicine, Shantou University Medical College, Shantou 515041, Guangdong Province, China
Pei-Hua Cao, Clinical Research Center, Zhujiang Hospital, Southern Medical University, Guangzhou 510280, Guangdong Province, China
ORCID number: Yuan Xie (0000-0001-9080-7358); Jin-Yu Qiu (0009-0002-5988-0498); Min Ju (0009-0008-5983-3361); Shao-Ru Cai (0000-0002-1197-6813); Gao-Fei Kang (0009-0007-6547-8346); Pi Guo (0000-0002-0990-6461); Zhi-Liang Tang (0009-0006-9298-2591); Pei-Hua Cao (0000-0003-4269-3899); Ke-Bo Zhong (0000-0001-7146-9461).
Co-first authors: Yuan Xie and Jin-Yu Qiu.
Co-corresponding authors: Pei-Hua Cao and Ke-Bo Zhong.
Author contributions: Xie Y wrote and revised the manuscript; Qiu JY performed data management and quality control; Xie Y and Qiu JY were co-first authors and contributed equally to this work; Ju M, Cai SR, and Kang GF helped to collect the data; Guo P proposed revision suggestions; Tang ZL helped supervise the trial; Cao PH participated in the statistical analysis; Zhong KB designed the study; Cao PH and Zhong KB are co-corresponding authors and contributed equally to this project. All authors have read and approved the final manuscript.
AI contribution statement: No AI tool was used to generate research data, interpret results, or formulate conclusions.
Supported by Guangdong S&T Program, No. 2025B1111110001; National Key R&D Program of China, No. 2022YFA1104900; and 2022 Artificial Liver Special Fund of the Beijing Liver and Gallbladder Health Public Welfare Foundation, No. iGandanF-1082022-RGG053.
Institutional review board statement: The study was reviewed and approved by the Medical Ethics Committee of Zhujiang Hospital, Southern Medical University, No. 2022-QX-004-01.
Clinical trial registration statement: This trial was registered at the Chinese Clinical Trial Registry website (https://www.chictr.org.cn), No. ChiCTR2600127409.
Informed consent statement: All study participants, or their legal representatives, provided written informed consent prior to study enrollment.
Conflict-of-interest statement: All the authors report no relevant conflicts of interest for this article.
CONSORT 2010 statement: The authors have read the CONSORT 2010 Statement, and the manuscript was prepared and revised according to the CONSORT 2010 Statement.
Data sharing statement: The technical appendix, statistical code, and dataset are available from the corresponding author upon a reasonable request.
Corresponding author: Ke-Bo Zhong, MD, Department of Hepatobiliary Surgery, Zhujiang Hospital, Southern Medical University, No. 253 Gongye Middle Avenue, Haizhu District, Guangzhou 510280, Guangdong Province, China. zhongkb@smu.edu.cn
Received: July 6, 2026
Revised: August 3, 2026
Accepted: September 9, 2026
Published online: January 7, 2027
Processing time: 126 Days and 23.3 Hours

Abstract
BACKGROUND

Cholangioscopy is an important tool used for minimally invasive biliary surgery, and image quality plays a crucial role in the diagnosis and treatment of biliary lesions. Currently, both digital cholangioscopy and fiberoptic cholangioscopy are widely used in clinical settings. However, from the perspective of clinical diagnostic requirements, few head-to-head studies have compared the two devices.

AIM

To evaluate the efficacy and safety of disposable digital single-operator cholangioscopy (DSOC) in terms of diagnostic and therapeutic procedures.

METHODS

A total of 114 participants were enrolled from January 2022 to December 2022. The participants were randomized at a ratio of 1:1 to the test and control groups. Disposable DSOCs were employed in the test group, whereas reusable fiberoptic single-operator cholangioscopies (FSOCs) were used in the control group. The primary outcome was measured by assessing the image quality compliance rate for clinical diagnostic requirements. The secondary outcome was assessed by score-based device performance.

RESULTS

The image quality compliance rates for clinical diagnostic requirements were observed to be 98.28% in the DSOC group, and 98.21% in the FSOC group (rate difference: 0.0615%; 95% confidence interval: -7.4994% to 7.8518%). 92.9% of the images were graded as excellent in the DSOC group, whereas 16.07% in the FSOC group (P < 0.05). DSOC performed better in terms of the smoothness of insertion and withdrawal, as well as the sealability of the instrument channel and flexibility of the bending section. Additionally, the accessory could cooperate closely with the instrument channel to complete the procedure in the DSOC group (P < 0.05). The overall rate of reported adverse events was 46.6% (27/58) in the DSOC group compared with 51.8% (29/56) in the FSOC group (P > 0.05).

CONCLUSION

DSOC was significantly superior to FSOC in terms of image quality, smoothness, sealability and flexibility. Digital cholangioscopy represents a growing trend in minimally invasive biliary surgery.

Key Words: Digital single-operator cholangioscopy; Fiberoptic single-operator cholangioscopy; Randomized controlled trial; Noninferiority trial; Comparative study

Core Tip: This randomized study compared digital cholangioscopy with fiberoptic cholangioscopy, from the perspective of clinical diagnostic requirements. The results revealed that digital cholangioscopy was significantly superior to fiberoptic cholangioscopy in terms of image quality, smoothness, sealability and flexibility, although the image quality compliance rates for clinical diagnostic requirements were comparable between the procedures. Additionally, the accessory could cooperate better with the instrument channel to complete the digital cholangioscopy procedure. Digital cholangioscopy represents a growing trend in minimally invasive biliary surgery.



INTRODUCTION

Biliary diseases (including biliary strictures and bile duct stones) are commonly encountered and can occasionally present challenges for expert physicians. Cholangioscopy is an important tool used for minimally invasive biliary surgery, and image quality plays a crucial role in the diagnosis and treatment of biliary lesions. Fiberoptic cholangioscopy was first introduced in the 1970s[1] but had limited applications because of low-quality images and time-consuming manipulation[2]. Two expert endoscopists were needed at that time to perform the procedure. In 2007, the first fiberoptic single-operator cholangioscopy (FSOC, SpyGlassTM Direct Visualization System, Boston Scientific Corp, Natick, MA, United States) was introduced to clinical practice. The emergence of FSOC enabled a single endoscopist to complete cholangioscopic procedures and resulted in the creation of several apparatuses, such as biopsy forceps, laser lithotripters and electrohydraulic lithotripters; these innovations allowed physicians to obtain access to hepatic biliary trees under direct visualization, thus yielding favorable results in both diagnostic and therapeutic areas[3-7]. In 2015, the first digital single-operator cholangioscopy (DSOC, SpyGlassTM Digital-Imaging System, Boston Scientific Corp, Natick, MA, United States) was released for public use and involved the utilization of a digital probe instead of a fiberoptic probe, thus further increasing image resolution and promoting operation efficiency[8-11]. However, from the perspective of clinical diagnostic requirements, few head-to-head studies have compared DSOC and FSOC, despite the fact that DSOC can improve cholangioscopic procedures. Thus, we performed a randomized multicenter study to evaluate the efficacy and safety of DSOC compared with FSOC in terms of diagnostic and therapeutic procedures.

MATERIALS AND METHODS
Study design and participants

This trial was a noninferiority, multicenter, randomized, open-label and blinded endpoint clinical study (Figure 1). It was conducted in 3 teaching hospitals in China. The participants were enrolled from January 2022 to December 2022 and followed up from February 2022 to January 2023. The trial protocol was approved and monitored by the Medical Ethics Committee of Zhujiang Hospital, Southern Medical University, No. 2022-QX-004-01. Written informed consent was obtained from each participant or legal representative. We followed the CONSORT reporting guidelines.

Figure 1
Figure 1 Participant flow diagram. DSOC: Digital single-operator cholangioscopy; FSOC: Fiberoptic single-operator cholangioscopy.
Inclusion and exclusion criteria

The following inclusion criteria for study eligibility were utilized: (1) Participants aged 18 to 75 years, regardless of sex; (2) Participants who needed cholangioscopic examinations or therapies; and (3) Participants who voluntarily participated in the trial. The following detailed exclusion criteria were utilized: (1) Participants with liver cirrhosis ascites; (2) Participants with bleeding tendencies; (3) Participants with unstable or complicated diseases, including acute biliary infections, organ failure, malignant tumors or hypoproteinemia (albumin < 30 g/L); (4) Pregnant or lactating women; (5) Participants with mental illness; and (6) Participants who had participated in other trials 30 days before enrollment in the study.

Randomization and masking

Participants were randomized at a 1:1 ratio to the test and control groups according to the inclusion and exclusion criteria specified in the trial protocol. Single-use DSOC (RP-H-C21, Redpine, China) was used in the test group for photography, observation, diagnosis and treatment, whereas reusable FSOC (CHF-P60, Olympus, Japan) was used in the control group (Figure 2).

Figure 2
Figure 2  Digital single-operator cholangioscopy used in the test group (left) and fiberoptic single-operator cholangioscopy used in the control group (right).

Randomization was performed by a professional statistician using SAS software. Randomized assignment tables were generated by using the PLAN process of SAS software, which was simulated via a computer. This clinical trial adopted an open design because the cholangioscopies used in the DSOC and FSOC groups differed in appearance.

Study procedures

Participants were randomized to the DSOC and FSOC groups. During the cholangioscopic operations, images were obtained for each participant (including the luminal mucosa and lesions of the biliary tracts) when the operator considered the images to have achieved the optimal resolution. These images were classified as excellent, good, ordinary, poor, or very poor according to the evaluation criteria, with a rigid hierarchy being utilized (Supplementary Table 1). The image quality was evaluated by an independent team that included three hepato-biliary surgery doctors who did not perform cholangioscopic operations on the participants. Given that the evaluation criteria were inevitably subject to a certain degree of subjective influence from the evaluators, the final determination of image quality was based on the assessments upon which two evaluators concurred. If the assessments of the three evaluators were mutually discordant, then a fourth evaluator was introduced, and the definitive judgment was rendered in accordance with the assessments that were consistent between any two evaluators. If each image of one participant was rated as being excellent or good, the image qualities of the participant were considered to have met the clinical diagnostic requirements. If any one of the images was rated as being ordinary, poor or very poor, the image qualities of the participant were considered to have not met the clinical diagnostic requirements. Moreover, both DSOC and FSOC were scored by the operators according to their specific performances (Supplementary Table 2).

Outcomes

The primary outcome was measured by assessing the image quality compliance rate for clinical diagnostic requirements. The following formula was used to calculate the rate: Image quality compliance rate for clinical diagnostic requirements = (the number of participants whose image quality met clinical diagnostic requirements/the total number of participants who underwent cholangioscopic operations) × 100%. The secondary outcome was assessed by evaluating the device performance, which was scored based on insertion, withdrawal, sealing and angulation for cholangioscopy, in addition to accessory insertion, withdrawal and compatibility.

Adverse events

Adverse events in this trial could have possibly involved bleeding, infection, regurgitation cholangitis, acute pancreatitis and perforation of the biliary tract. When the abovementioned adverse events occurred, active measures were performed to treat the participants and we obtained detailed medical records.

Sample size calculation

In this trial, the estimated image quality compliance rate for clinical diagnostic requirements was 97%, with a predefined difference between the DSOC and FSOC groups of D = 0. When considering possible deviations in actual clinical practice, the investigators and statisticians set a noninferiority margin of Δ = -10% (DSOC group-FSOC group) for this trial. The following parameters were utilized: Δ = 0.05 (two-sided) and power (1-β) = 0.8. We calculated a sample size of 92 participants (46 participants per group) by using the sample size calculation formula for a noninferiority trial. On the basis of an anticipated dropout rate of approximately 20% during the trial, along with the requirements for central sample allocation and block randomization, the sample size per group was increased to 58 participants, thus resulting in a total sample size of 116 participants.

Statistical analysis

Continuous variables were presented as the mean ± SD, and the data were analyzed via the independent two-sample t-test. Categorical data were presented as frequencies, and the data were analyzed using the χ2 or Fisher’s exact test. All P values were 2-sided and considered to be statistically significant when the value was less than 0.05. Statistical analyses were performed by using SAS 9.3 software.

RESULTS
Participants

A total of 114 participants were enrolled at 3 university teaching hospitals in China, of whom 58 participants were in the DSOC group, with an average age of 57.09 ± 12.86 years; moreover, 56 participants were in the FSOC group, with an average age of 57.50 ± 13.46 years. The DSOC and FSOC groups were matched at baseline, and no significant differences were observed in sex, age, weight, pulse or blood pressure among the groups. Blood test results did not markedly differ between the groups, including blood and coagulation routine tests, as well as liver and renal function measurements (P > 0.05, Table 1). Overall, the clinical characteristics of the enrolled participants were well balanced, and they were observed to be homogeneous between the groups before the cholangioscopic operations. The participants underwent subsequent cholangioscopic operations via laparoscopic choledochendysis or T-tube sinus tract in both the DSOC and FSOC groups, and significant differences were not observed in terms of how the cholangioscopy could obtain access to the common bile duct; furthermore, no differences were observed with respect to the conditions in which the stone retrieval baskets were used (P > 0.05; Table 2).

Table 1 Demographic and clinical characteristics at baseline, n (%)/mean ± SD.
Characteristics
DSOC group
FSOC group
P value
Sex0.5763
Male27 (46.55)29 (51.79)
Female31 (53.45)27 (48.21)
Age (years)57.09 ± 12.8657.50 ± 13.460.8683
Weight (kg)57.79 ± 10.9559.46 ± 11.500.4284
Pulse (bpm)76.67 ± 12.5276.09 ± 13.140.8087
SBP (mmHg)123.90 ± 19.74125.71 ± 20.910.6340
DBP (mmHg)76.95 ± 11.7677.32 ± 11.070.8619
RBC (1012/L)4.13 ± 0.634.31 ± 0.630.1351
WBC (109/L)6.80 ± 3.756.74 ± 2.830.9150
PLT (109/L)213.12 ± 80.13226.86 ± 78.710.3580
PT (seconds)12.42 ± 1.2712.25 ± 1.180.4639
APTT (seconds)31.46 ± 4.0530.66 ± 3.960.2927
Alb (g/L)41.00 ± 4.5441.55 ± 4.980.5373
ALT (U/L)77.58 ± 115.0188.00 ± 112.270.6256
Cr (μmol/L)76.49 ± 28.3371.49 ± 15.360.2463
BUN (mmol/L)4.81 ± 2.094.39 ± 1.440.2195
Table 2 Surgical approaches and concomitant instruments for the cholangioscopy operation, n (%).
Surgical approaches
DSOC group
FSOC group
P value
Laparoscopic choledochendysis24 (41.38)28 (50.00)0.3556
T-tube sinus tract34 (58.62)28 (50.00)
Stone retrieval basket
Without45 (77.59)43 (76.79)0.9189
With13 (22.41)13 (23.21)
Primary outcome

In this trial, a total of 169 images were collected from 58 participants in the DSOC group and 168 images were collected from 56 participants in the FSOC group. The locations of the collected images included the common bile duct, the first- and second-order biliary branches, and the confluence of the second-order biliary branches. The contents of the abovementioned images included biliary stones and the mucosa, as well as stone retrieval baskets and holmium laser fibers (Table 3). After evaluation by an independent team, we observed that the image quality of the DSOC group was superior to that of the FSOC group. Overall, 92.9% of the images were graded as excellent in the DSOC group, whereas 16.07% of the images were excellent in the FSOC group (P < 0.05; Table 4). Almost all of the images in both groups met the clinical diagnostic requirements. The image quality compliance rates for clinical diagnostic requirements were 98.28% in the DSOC group and 98.21% in the FSOC group (rate difference: 0.0615%; 95% confidence interval: -7.4994% to 7.8518%). Noninferiority was established, as the lower confidence limit exceeded the prespecified margin of -10% (Table 5).

Table 3 Image acquisition data for cholangioscopy, n (%).
Participant numbers based on image contents
DSOC group
FSOC group
P value
Common bile duct1.0000
Absent2 (3.45)1 (1.79)
Present56 (96.55)55 (98.21)
First-order biliary branches0.4957
Absent56 (96.55)56 (100.00)
Present2 (3.45)0 (0.00)
Second-order biliary branches0.2436
Absent55 (94.83)56 (100.00)
Present3 (5.17)0 (0.00)
Confluence of second-order biliary branches0.0550
Absent0 (0.00)4 (7.14)
Present58 (100.00)52 (92.86)
Stone0.8488
Absent29 (50.00)29 (51.79)
Present29 (50.00)27 (48.21)
Stone retrieval basket/holmium laser fiber0.1339
Absent33 (56.90)24 (42.86)
Present25 (43.10)32 (57.14)
Biliary mucosa
Absent0 (0.00)0 (0.00)
Present58 (100.00)56 (100.00)
Image numbers based on biliary locations
Common bile duct100 (59.17)110 (65.48)0.0411
Second-order biliary branches5 (2.96)0 (0.00)
First-order biliary branches2 (1.18)0 (0.00)
Confluence of second-order biliary branches62 (36.69)58 (34.52)
Image numbers based on captured contents
Stone29 (17.16)27 (16.07)0.6765
Stone retrieval basket/holmium laser fiber27 (15.98)33 (19.64)
Biliary mucosa113 (66.86)108 (64.29)
Table 4 Image quality evaluation for cholangioscopy, n (%).
Image quality evaluation
DSOC group
FSOC group
P value
Excellent157 (92.90)27 (16.07)< 0.0001
Good11 (6.51)140 (83.33)
Ordinary1 (0.59)1 (0.60)
Very poor0 (0.00)0 (0.00)
Poor0 (0.00)0 (0.00)
Table 5 Non-inferiority test concerning the image quality compliance rate for clinical diagnostic requirements, n (%).
Study group
n
Image quality compliance rate for clinical diagnostic requirements
Rate difference (95%CI)
DSOC group5857 (98.28)0.0615% (-7.4994% to 7.8518%)
FSOC group5655 (98.21)
Secondary outcome

Furthermore, the device performances of the DSOC and FSOC groups were scored by the operator, with focuses on insertion, withdrawal, sealing and angulation for the cholangioscopy, in addition to accessory insertion, withdrawal and compatibility. We observed that the DSOC performed better in terms of the smoothness of insertion and withdrawal, as well as the sealability of the instrument channel, and flexibility of the bending section. Additionally, the accessory could cooperate closely with the instrument channel to complete the procedure in the DSOC (P < 0.05; Table 6).

Table 6 Cholangioscopy performance scored by operator, n (%).
Performance metrics
DSOC group
FSOC group
P value
Insertion< 0.0001
10 (0.00)0 (0.00)
20 (0.00)0 (0.00)
30 (0.00)0 (0.00)
414 (24.14)43 (76.79)
544 (75.86)13 (23.21)
Withdrawal< 0.0001
10 (0.00)0 (0.00)
20 (0.00)0 (0.00)
30 (0.00)0 (0.00)
411 (18.97)41 (73.21)
547 (81.03)15 (26.79)
Sealing0.0090
10 (0.00)0 (0.00)
23 (5.17)0 (0.00)
35 (8.62)15 (26.79)
450 (86.21)41 (73.21)
Angulation< 0.0001
10 (0.00)0 (0.00)
20 (0.00)0 (0.00)
33 (5.17)8 (14.29)
419 (32.76)37 (66.07)
536 (62.07)11 (19.64)
Accessory insertion and withdrawal0.2923
10 (0.00)0 (0.00)
20 (0.00)0 (0.00)
32 (6.25)2 (5.71)
414 (43.75)22 (62.86)
516 (50.00)11 (31.43)
Accessory compatibility0.0016
10 (0.00)0 (0.00)
20 (0.00)0 (0.00)
34 (12.50)2 (5.71)
49 (28.12)25 (71.43)
519 (59.38)8 (22.86)
Adverse events

The overall rate of reported adverse events was 46.6% (27/58) in the DSOC group compared with 51.8% (29/56) in the FSOC group; however, marked differences were not observed in the proportions of adverse events between the two groups, and the adverse events were not correlated with the type of cholangioscopies after careful investigation (P > 0.05; Table 7). Major adverse events included fever, abdominal pain, nausea and vomiting. Serious adverse events occurred in both groups. One participant suffered from peritonitis in the DSOC group, and another participant experienced hemorrhagic shock in the FSOC group; however, the two participants recovered after the relevant treatments (Table 8).

Table 7 Adverse event occurrence status of the participants, n (%).
Adverse events
DSOC group
FSOC group
P value
Occurrence0.576
No31 (53.4)27 (48.2)
Yes27 (46.6)29 (51.8)
Serious adverse events
Occurrence0.98
No57 (98.28)55 (98.21)
Yes1 (1.72)1 (1.79)
Table 8 Adverse event profiles of the participants, n (%).
Nonserious events
DSOC group
FSOC group
P value
Fever5 (8.62)6 (10.71)0.629
Nausea and vomiting4 (6.90)2 (3.57)
Abdominal pain17 (29.31)20 (35.71)
Serious events
Shock01 (1.79)1.000
Peritonitis1 (1.72)0
DISCUSSION

Currently, both digital and fiberoptic cholangioscopy have been widely used for the diagnosis and treatment of biliary diseases. It appears that digital cholangioscopy can provide higher-resolution images for doctors in contrast with fiberoptic cholangioscopy; however, from the perspective of clinical diagnostic requirements, to our knowledge, head-to-head studies concerning comparisons between the two devices are scarce. Thus, we designed a non-inferiority, multicenter, randomized, open-label, blinded endpoint trial to evaluate the efficacy and safety of DSOC in terms of diagnostic and therapeutic procedures. As expected, our clinical study indicated that the image quality of digital cholangioscopy was superior to that of fiberoptic cholangioscopy although both of the techniques met the clinical diagnostic requirements. Moreover, digital cholangioscopy performed better in terms of smoothness, sealability and flexibility, and the accessory could cooperate more closely with the instrument channel to complete the procedure in the digital cholangioscopy approach. In this trial, digital cholangioscopy employed complementary metal oxide semiconductor image sensors, whereas fiberoptic cholangioscopy utilized charge coupled device sensors. The complementary metal oxide semiconductor image sensors offered greater sensitivity and higher resolution compared with their charge coupled device counterparts. In addition, digital cholangioscope had a thinner diameter (4.4 mm) than that of fiberoptic cholangioscope (5.2 mm), thus digital cholangioscope could readily access the biliary or sinus tract, and could be operated more easily in the narrow space compared with the fiberoptic version. The abovementioned cases might explain how they translated into better participant outcomes.

A previous cohort study demonstrated that the digital cholangioscopy approach could provide enhanced diagnostic yield, shorter procedure time and less radiation exposure in biliary stone diseases and indeterminate biliary strictures in comparison with fiberoptic cholangioscopy[12]. In another study, digital cholangioscopy was demonstrated to be a feasible and effective option for the management of difficult cholelithiasis, as well as for visually evaluating and obtaining histological samples for indeterminate biliary strictures[13]. However, the referred studies were either retrospective or descriptive in nature, and it was speculated that better digital visualization might explain the abovementioned positive results. To date, no studies have verified the noninferiority of digital cholangioscopy in comparison to fiberoptic approach in terms of image quality. This is a pending issue and a concrete answer is needed. By introducing an independent image quality evaluation team, our randomized study revealed that the image quality of digital cholangioscopy was superior to that of fiberoptic cholangioscopy from the perspective of clinical diagnostic requirements. However, image quality itself is not the ultimate clinical goal. Theoretically, higher image quality could potentially affect important clinical outcomes, such as improved biopsy targeting, shortened operative time, reduced surgical complexity, more accurate characterization of biliary tract lesions, enhanced confidence in difficult stone removal procedures and increased success rates in treating anatomically complex biliary tract diseases, but these results were not directly assessed in this trial.

Clinical studies are scarce concerning the discrepancy between digital and fiberoptic cholangioscopy methods in operational performance, although results obtained in a bench model revealed that digital cholangioscopy was significantly superior to fiberoptic cholangioscopy in several aspects, including the identification of target lesions, the duration of the procedure and the ease of use[14]. In real clinical settings, our study revealed that digital cholangioscopy performed better than fiberoptic cholangioscopy, regardless of the smoothness, sealability, flexibility and compatibility with which the accessory coordinated with the cholangioscopy approach (according to the feedback obtained from the operators). In parallel with previous studies[15,16], our data revealed that there was no significant difference in the rate of adverse events between the two types of cholangioscopy approaches. Most of the adverse events were mild, and they mainly involved fever, abdominal pain, nausea and vomiting. Digital cholangioscopy demonstrated a favorable safety profile for diagnostic and therapeutic procedures.

Originally introduced to clinical practice in the 1970s, fiberoptic cholangioscopy has evolved into digital cholangioscopy with several decades of modifications and refinements. Notably, the digital cholangioscopy approach used in our study involved a single-use endoscopic catheter and a small main unit. The endoscopic catheter was disposable, and the main unit was portable, which could reduce cross-infections and maintenance costs in hospitals (to a certain extent). The advent of endoscopy marked the transition of biliary surgery from the era of open surgery to minimally invasive surgery, thereby broadening the approach to obtain access to hepatic biliary trees, such as percutaneous transhepatic cholangioscopy. The miniaturization of endoscopy may further decrease the luminal size required for interventional tools, thus facilitating the resolution of difficult lesions such as cystic duct calculi[17].

Our study has several limitations. First, the operators were not blinded to the cholangioscopies because the features of the digital cholangioscopies clearly distinguished them from those of the fiberoptic versions. Second, participants with biliary tumors were not included in our study; thus, images for biliary tumors were absent, and comparisons between digital and fiberoptic cholangioscopy approaches were lacking in terms of image quality for biliary tumors, which inevitably limited the general applicability of the conclusions. Therefore, it should be stated that the current conclusions primarily apply to benign biliary tract diseases and that the findings should not be overgeneralized to malignant diseases. Finally, the device performance was assessed by a single operator in both the test and control groups at a single center, which may have been biased in terms of manipulation, although the operators were experienced users of both digital and fiberoptic cholangioscopy methods.

CONCLUSION

Disposable digital cholangioscopy was significantly superior to fiberoptic cholangioscopy in terms of image quality, smoothness, sealability and flexibility. Digital cholangioscopy represents a growing trend in minimally invasive biliary surgery.

ACKNOWLEDGEMENTS

We thanked Tang JM (Jian-Ming Tang) and Li N (Nan Li) for offering disposable digital single-operator cholangioscopes in this trial.

References
1.  Nakajima M, Akasaka Y, Fukumoto K, Mitsuyoshi Y, Kawai K. Peroral cholangiopancreatosocopy (PCPS) under duodenoscopic guidance. Am J Gastroenterol. 1976;66:241-247.  [PubMed]  [DOI]
2.  Karagyozov P, Boeva I, Tishkov I. Role of digital single-operator cholangioscopy in the diagnosis and treatment of biliary disorders. World J Gastrointest Endosc. 2019;11:31-40.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in CrossRef: 41]  [Cited by in RCA: 34]  [Article Influence: 4.9]  [Reference Citation Analysis (3)]
3.  Draganov PV, Chauhan S, Wagh MS, Gupte AR, Lin T, Hou W, Forsmark CE. Diagnostic accuracy of conventional and cholangioscopy-guided sampling of indeterminate biliary lesions at the time of ERCP: a prospective, long-term follow-up study. Gastrointest Endosc. 2012;75:347-353.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 208]  [Cited by in RCA: 171]  [Article Influence: 12.2]  [Reference Citation Analysis (1)]
4.  Ramchandani M, Reddy DN, Gupta R, Lakhtakia S, Tandan M, Darisetty S, Sekaran A, Rao GV. Role of single-operator peroral cholangioscopy in the diagnosis of indeterminate biliary lesions: a single-center, prospective study. Gastrointest Endosc. 2011;74:511-519.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 159]  [Cited by in RCA: 130]  [Article Influence: 8.7]  [Reference Citation Analysis (6)]
5.  Maydeo A, Kwek BE, Bhandari S, Bapat M, Dhir V. Single-operator cholangioscopy-guided laser lithotripsy in patients with difficult biliary and pancreatic ductal stones (with videos). Gastrointest Endosc. 2011;74:1308-1314.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 120]  [Cited by in RCA: 99]  [Article Influence: 6.6]  [Reference Citation Analysis (1)]
6.  Hartman DJ, Slivka A, Giusto DA, Krasinskas AM. Tissue yield and diagnostic efficacy of fluoroscopic and cholangioscopic techniques to assess indeterminate biliary strictures. Clin Gastroenterol Hepatol. 2012;10:1042-1046.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 75]  [Cited by in RCA: 62]  [Article Influence: 4.4]  [Reference Citation Analysis (0)]
7.  Laleman W, Verraes K, Van Steenbergen W, Cassiman D, Nevens F, Van der Merwe S, Verslype C. Usefulness of the single-operator cholangioscopy system SpyGlass in biliary disease: a single-center prospective cohort study and aggregated review. Surg Endosc. 2017;31:2223-2232.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 77]  [Cited by in RCA: 66]  [Article Influence: 7.3]  [Reference Citation Analysis (2)]
8.  Navaneethan U, Hasan MK, Kommaraju K, Zhu X, Hebert-Magee S, Hawes RH, Vargo JJ, Varadarajulu S, Parsi MA. Digital, single-operator cholangiopancreatoscopy in the diagnosis and management of pancreatobiliary disorders: a multicenter clinical experience (with video). Gastrointest Endosc. 2016;84:649-655.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 177]  [Cited by in RCA: 166]  [Article Influence: 16.6]  [Reference Citation Analysis (2)]
9.  Lenze F, Bokemeyer A, Gross D, Nowacki T, Bettenworth D, Ullerich H. Safety, diagnostic accuracy and therapeutic efficacy of digital single-operator cholangioscopy. United European Gastroenterol J. 2018;6:902-909.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 53]  [Cited by in RCA: 52]  [Article Influence: 6.5]  [Reference Citation Analysis (4)]
10.  Brewer Gutierrez OI, Bekkali NLH, Raijman I, Sturgess R, Sejpal DV, Aridi HD, Sherman S, Shah RJ, Kwon RS, Buxbaum JL, Zulli C, Wassef W, Adler DG, Kushnir V, Wang AY, Krishnan K, Kaul V, Tzimas D, DiMaio CJ, Ho S, Petersen B, Moon JH, Elmunzer BJ, Webster GJM, Chen YI, Dwyer LK, Inamdar S, Patrick VB, Attwell A, Hosmer A, Ko C, Maurano A, Sarkar A, Taylor LJ, Gregory MH, Strand DS, Raza A, Kothari S, Harris JP, Kumta NA, Manvar A, Topazian MD, Lee YN, Spiceland CM, Trindade AJ, Bukhari MA, Sanaei O, Ngamruengphong S, Khashab MA. Efficacy and Safety of Digital Single-Operator Cholangioscopy for Difficult Biliary Stones. Clin Gastroenterol Hepatol. 2018;16:918-926.e1.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 60]  [Cited by in RCA: 82]  [Article Influence: 10.3]  [Reference Citation Analysis (1)]
11.  Wong JC, Tang RS, Teoh AY, Sung JJ, Lau JY. Efficacy and safety of novel digital single-operator peroral cholangioscopy-guided laser lithotripsy for complicated biliary stones. Endosc Int Open. 2017;5:E54-E58.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 33]  [Cited by in RCA: 41]  [Article Influence: 4.6]  [Reference Citation Analysis (0)]
12.  Mizrahi M, Khoury T, Wang Y, Cohen J, Sheridan J, Chuttani R, Berzin TM, Sawhney MS, Pleskow DK. "Apple Far from the Tree": comparative effectiveness of fiberoptic single-operator cholangiopancreatoscopy (FSOCP) and digital SOCP (DSOCP). HPB (Oxford). 2018;20:285-288.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 18]  [Cited by in RCA: 20]  [Article Influence: 2.5]  [Reference Citation Analysis (1)]
13.  Al Lehibi A, Aljahdali E, Al Balkhi A, Almasoudi T, Al Ghamdi A, Al Sayari K, Al Khathlan A, Qutub A, AlEid A, Ahmad S, Al Otaibi N, Al Mtawa A. The utility of digital cholangioscopy (SpyGlass DS) in biliary and pancreatic diseases: A clinical feasibility study at two tertiary care centers in Saudi Arabia (with Videos). Arab J Gastroenterol. 2020;21:49-53.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 1]  [Cited by in RCA: 6]  [Article Influence: 1.0]  [Reference Citation Analysis (0)]
14.  Shah RJ, Neuhaus H, Parsi M, Reddy DN, Pleskow DK. Randomized study of digital single-operator cholangioscope compared to fiberoptic single-operator cholangioscope in a novel cholangioscopy bench model. Endosc Int Open. 2018;6:E851-E856.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 23]  [Cited by in RCA: 20]  [Article Influence: 2.5]  [Reference Citation Analysis (0)]
15.  Dimas ID, Vardas E, Papastergiou V, Fragaki M, Velegraki M, Mpitouli A, Voudoukis E, Theodoropoulou A, Giannikaki E, Chlouverakis G, Paspatis GA. Comparison of digital versus fiberoptic cholangioscopy in patients requiring evaluation of bile duct disease or treatment of biliary stones. Ann Gastroenterol. 2019;32:199-204.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 2]  [Cited by in RCA: 7]  [Article Influence: 1.0]  [Reference Citation Analysis (0)]
16.  Fugazza A, Gabbiadini R, Tringali A, De Angelis CG, Mosca P, Maurano A, Di Mitri R, Manno M, Mariani A, Cereatti F, Bertani H, Sferrazza S, Donato G, Tarantino I, Cugia L, Aragona G, Cantù P, Mazzocchi A, Canfora ML, Venezia L, Bendia E, Maroni L, Zulli C, Conte E, Soriani P, Ligresti D, Vilardo E, Penagini R, Benedetti A, Arcidiacono PG, Khalaf K, Troncone E, Costamagna G, Repici A, Anderloni A. Digital single-operator cholangioscopy in diagnostic and therapeutic bilio-pancreatic diseases: A prospective, multicenter study. Dig Liver Dis. 2022;54:1243-1249.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 11]  [Cited by in RCA: 17]  [Article Influence: 4.3]  [Reference Citation Analysis (2)]
17.  Feng Y, Liang Y, Liu Y, Zhang Y, Huang S, Ren L, Ma X, Zhou A, Shi R. Digital cholangioscopy-assisted, direct visualization-guided, radiation-free, endoscopic retrograde intervention for cholelithiasis: technical feasibility, efficacy, and safety. Surg Endosc. 2024;38:1637-1646.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 2]  [Cited by in RCA: 5]  [Article Influence: 2.5]  [Reference Citation Analysis (0)]
Footnotes

Peer review: Externally peer reviewed.

Peer-review model: Single blind

Specialty type: Gastroenterology and hepatology

Country of origin: China

Peer-review report’s classification

Scientific quality: Grade B, Grade B

Novelty: Grade C, Grade C

Creativity or innovation: Grade B, Grade C

Scientific significance: Grade B, Grade C

P-Reviewer: Rhodin KE, Associate Professor, United States; Tokumaru S, PhD, Japan S-Editor: Wu S L-Editor: A P-Editor: Zhao YQ

Write to the Help Desk