Randomized Controlled Trial
Copyright ©The Author(s) 2024.
World J Gastroenterol. Aug 21, 2024; 30(31): 3680-3688
Published online Aug 21, 2024. doi: 10.3748/wjg.v30.i31.3680
Table 1 Endoscopic working channel cleanliness scale
Working channel findings
Endoscope inspection ratings
Dark colored debris0 (none): No dark colored debris
1 (mild): Small dark particles or linear debris in up to 3 locations
2 (moderate): Small dark particles or debris in > 3 locations or larger debris in 1 location
3 (severe): Larger dark particles or debris in > 1 location
Light colored debris0 (none): No light colored/reflective debris
1 (mild): Small light-colored/reflective particles or residue in up to 3 locations
2 (moderate): Small light-colored/reflective particles or residue in > 3 locations, or larger debris/residue in 1
3 (severe): Larger light-colored/reflective particles or debris in > 1 location or present in continuous large
Opaque fluid0 (none): No opaque fluid
1 (mild): Up to 5 individual drops of opaque fluid within channel
2 (moderate): 6 to 10 individual drops of opaque fluid within channel
3 (severe): Greater than 10 individual drops of opaque fluid or pooling of fluid within channel
Clear fluid0 (none): No clear fluid
1 (mild): Up to 5 individual drops of clear fluid within channel
2 (moderate): 6 to 10 individual drops of clear fluid within channel
3 (severe): Greater than 10 individual drops of clear fluid or pooling of fluid within channel
Table 2 Teflon tube damage scale
Damage
Endoscope inspection ratings
Scratch characteristics0 (none): No scratches visualized
1 (mild): Superficial, with regular margins, < 1 image field in length
2 (moderate): Deeper, with irregular margins or > 1 image field in length
3 (severe): Deep with irregular margins
Scratch severity0 (none): No scratches visualized
1 (mild): < 10 scratches visualized
2 (moderate): 10-20 scratches visualized
3 (severe): > 20 scratches visualized
Adherent peel0 (none): No adherent peel visualized
1 (mild): Localized to 1 small (< 1 cm) site
2 (moderate): Present in 2–3 locations or 1-2 cm site
3 (severe): Present in > 3 locations or single > 2 cm site
Table 3 Comparison of details of cleanliness scores of endoscopic working channels, n (%)
Working channel findings
Control group (n = 30)
Experimental group (n = 30)
χ2
P value
Dark colored debris4 (13.3)2 (6.7)0.740.671
Light colored debris20 (66.7)13 (43.3)3.300.069
Residual debris22 (73.3)14 (46.7)4.440.035
Opaque fluid2 (6.7)3 (10.0)0.221.00
Clear fluid14 (46.7)18 (60.0)1.070.438
Table 4 Comparison of damage scoring details for Teflon tubes
Working channel findings
Control group (n = 30)
Experimental group (n = 30)
Z
P value
Scratch characteristics1 (1, 2)1 (1, 2)-1.4730.141
Scratch severity1 (1, 2)1 (1, 1)-1.6290.103
Adherent peel1.5 (1, 2)1 (1, 2)-1.7770.076