AIM: To conduct a bacterial culture study for monitoring decontamination of automated endoscope reprocessors (AERs) after high-level disinfection (HLD).
METHODS: From February 2006 to January 2011, authors conducted randomized consecutive sampling each month for 7 AERs. Authors collected a total of 420 swab cultures, including 300 cultures from 5 gastroscope AERs, and 120 cultures from 2 colonoscope AERs. Swab cultures were obtained from the residual water from the AERs after a full reprocessing cycle. Samples were cultured to test for aerobic bacteria, anaerobic bacteria, and mycobacterium tuberculosis.
RESULTS: The positive culture rate of the AERs was 2.0% (6/300) for gastroscope AERs and 0.8% (1/120) for colonoscope AERs. All the positive cultures, including 6 from gastroscope and 1 from colonoscope AERs, showed monofloral colonization. Of the gastroscope AER samples, 50% (3/6) were colonized by aerobic bacterial and 50% (3/6) by fungal contaminations.
CONCLUSION: A full reprocessing cycle of an AER with HLD is adequate for disinfection of the machine. Swab culture is a useful method for monitoring AER decontamination after each reprocessing cycle. Fungal contamination of AERs after reprocessing should also be kept in mind.