Assessment of quality control system by sigma metrics and quality goal index ratio: A roadmap towards preparation for NABL
Monica Verma, Kiran Dahiya, Veena Singh Ghalaut, Vasudha Dhupper
Monica Verma, Kiran Dahiya, Veena Singh Ghalaut, Vasudha Dhupper, Department of Biochemistry, Pt. B.D. Sharma, University of Health Sciences, Rohtak 124001, Haryana, India
Author contributions: Verma M designed the research, interpreted the results, and drafted the manuscript; Dahiya K and Ghalaut VS helped in statistical analysis, critical revision, and final approval of the article; Ghalaut VS and Dhupper V helped in making critical revisions and final approval of the article.
Conflict-of-interest statement: None exists.
Corresponding author to: Monica Verma, DNB, MBBS, MD, Assistant Professor, Doctor, Department of Biochemistry, Pt. B.D. Sharma, University of Health Sciences, House No. 315, 3rd Floor, Model Town, Rohtak 124001, Haryana, India. monisoni26@gmail.com
Telephone: +91-98-13334543
Received: July 31, 2018
Peer-review started: July 31, 2018
First decision: August 20, 2018
Revised: September 4, 2018
Accepted: October 17, 2018
Article in press: October 17, 2018
Published online: November 29, 2018
Processing time: 121 Days and 4.6 Hours
Core Tip

Core tip: As the majority of tests take place in the biochemistry laboratory, it plays a major role in patient care. Therefore, it is necessary to follow a proper quality management system to provide accurate and precise reports to patients. National Accreditation Board for Testing and Calibration of Laboratories accreditation is an important benchmark for “A” grade quality. Sigma metrics is also a well-known self-assessment tool to guide quality control strategy design. On the basis of sigma metrics and quality goal index ratio, it may be concluded that the Clinical Biochemistry Laboratory, PGIMS, Rohtak was able to achieve satisfactory results with world class performance for many analytes one year preceding accreditation by the National Accreditation Board for Testing and Calibration of Laboratories.