Copyright
©The Author(s) 2017.
World J Radiol. Feb 28, 2017; 9(2): 79-84
Published online Feb 28, 2017. doi: 10.4329/wjr.v9.i2.79
Published online Feb 28, 2017. doi: 10.4329/wjr.v9.i2.79
Table 1 Imaging protocol for both 3T and 7T magnetic resonance imaging
| Parameter | 7T T2 SPACE | 3T T2 SPACE | 7T TOF | 3T TOF |
| TE | 101 | 77 | 2.81 | 3.23 |
| TR | 1500 | 1600 | 60 | 21 |
| Slice thickness | 625.00 μm | 630.00 μm | 160.00 mm | 1.00 mm |
| Pixel size | 0.600/0.600 | 0.625/0.625 | 0.3125/0.3125 | 0.234/0.234 |
| Number of averages | 2 | 4 | 1 | 1 |
| Field of view | 136 × 160 | 100 × 160 | 160 × 160 | 107 × 120 |
Table 2 Data from 3 subjective analyses of image quality with average and standard deviation
| Subject | B mode | 3T | 7T |
| 1 | 3.00 | 3.33 | 3.00 |
| 2 | 3.00 | 2.00 | Not available |
| 3 | 3.00 | 4.33 | 4.67 |
| 4 | 4.00 | 2.00 | 4.33 |
| Average | 3.25 | 2.92 | 4.00 |
| Standard deviation | 0.50 | 1.13 | 0.882 |
Table 3 RM one-way ANOVA and Tukey’s multiple comparisons test
| Vessel wall measurements | P value |
| Lumen area | 0.3152 |
| Wall area | 0.0628 |
| Total vessel area | 0.0573 |
| Normalized wall index | 0.0362a |
| Wall thickness | 0.0519 |
- Citation: Pruzan AN, Kaufman AE, Calcagno C, Zhou Y, Fayad ZA, Mani V. Feasibility of imaging superficial palmar arch using micro-ultrasound, 7T and 3T magnetic resonance imaging. World J Radiol 2017; 9(2): 79-84
- URL: https://www.wjgnet.com/1949-8470/full/v9/i2/79.htm
- DOI: https://dx.doi.org/10.4329/wjr.v9.i2.79