Table 1 Means, standard deviations, calculated difference between the measured serum creatinine level (mg/dL) using StatSensor™ creatinine and the laboratory reference, Pearson's correlation coefficients, paired t-test and Maloney-Rastogi test, subdivided by study centre, pre- and post-offset correction
Centre 1 | Centre 2 | Mean (centre 1 +2) | Pearson correlation coefficient | P value (paired t-test) | P value (Maloney-Rastogi-test) | |||
StatSensor™ | ||||||||
cSCr | Pre offset | Device A | 0.80 ± 0.34 | 0.89 ± 0.27 | 0.85 ± 0.31 | Inter-device evaluation (Device A vs B) | ||
Device B | 0.81 ± 0.34 | 0.89 ± 0.26 | 0.85 ± 0.30 | Center 1 = 0.97 | Center 1 = 0.404 | Center 1 = 0.480 | ||
Difference between Device A vs B | -0.01 ± 0.09 | 0.00 ± 0.09 | Center 2 = 0.95 | Center 2 = 0.689 | Center 2 = 0.521 | |||
Mean (Device A + B) | 0.80 ± 0.34 | 0.89 ± 0.26 | 0.85 ± 0.31 | |||||
Post offset | Device A | 0.97 ± 0.50 | ||||||
Device B | 0.96 ± 0.50 | |||||||
Laboratory reference | ||||||||
vSCr | 0.96 ± 0.42 | 1.02 ± 0.26 | 0.99 ± 0.35 | Inter-method evaluation (cSCr vs vSCr) | ||||
Difference | ||||||||
cSCr - vSCr | Pre offset | Mean (Device A + B) | -0.16 ± 0.16a | -0.14 ± 0.15b | -0.15 ± 0.15c | < 0.0001a,b,c | < 0.0001a,c; 0.9962b | |
Device A | -0.16 ± 0.16 | 0.93 | < 0.0001 | < 0.0001 | ||||
Device B | -0.15 ± 0.17 | 0.92 | < 0.0001 | < 0.0001 | ||||
Post offset | Device A | 0.01 ± 0.19 | 0.93 | 0.638 | < 0.0001 | |||
Device B | 0.00 ± 0.20 | 0.92 | 0.955 | < 0.0001 |
Table 2 Pearson's correlation coefficients for creatinine values measured using the laboratory reference (venous serum creatinine) and StatSensor™ creatinine, subdivided by study centre, renal function and gender
vSCr (laboratory) vs | cSCr (StatSensor™) | Overall | |
Device A | Device B | ||
Study centre | |||
Centre 1 | 0.93 | 0.92 | 0.93 |
Centre 2 | 0.85 | 0.82 | 0.84 |
Overall | 0.90 | 0.88 | 0.89 |
Renal function | |||
vSCr < 1.2 mg/dL | 0.66 | 0.59 | 0.63 |
vSCr > 1.2 mg/dL | 0.91 | 0.90 | 0.91 |
Gender | |||
Male | 0.91 | 0.89 | 0.90 |
Female | 0.81 | 0.78 | 0.80 |
Table 3 Results of the re-analysed data from centre 1 (n = 201), shown pre- and post-implementation of the curve offset, separated for the two devicespre- and post-offset correction
Device | Slope | Intercept (mg/dL) | r2 | Mean difference (mg/dL) (method-difference) | STDV of differences (method-difference) | % bias | |
Pre offset | A | 0.765 | 0.065 | 0.870 | -0.16 | 0.16 | -15.54 |
B | 0.745 | 0.089 | 0.845 | -0.16 | 0.17 | -14.76 | |
Post offset | A | 1.110 | -0.099 | 0.870 | 0.01 | 0.19 | 0.39 |
B | 1.088 | -0.084 | 0.845 | 0.00 | 0.20 | -0.02 |
Table 4 Sensitivity, specificity, positive predictive value, negative predictive value and mean values for the cut-off value of 1.2 mg/dL, of the re-analysed data from centre 1 (n = 201), shown pre- and post-implementation of the curve offset, separated for the two devices (%)
Device | Sensitivity | Specificity | PPV | NPV | |
Pre offset | A | 35.48 | 99.41 | 91.67 | 89.42 |
B | 41.94 | 99.41 | 92.86 | 90.37 | |
Post offset | A | 80.65 | 98.26 | 89.29 | 95.57 |
B | 70.97 | 94.12 | 68.75 | 94.67 | |
Mean (A + B) | 77.42 | 94.71 | 72.37 | 95.83 |