Basic Study
Copyright ©The Author(s) 2022.
World J Hepatol. Jul 27, 2022; 14(7): 1357-1364
Published online Jul 27, 2022. doi: 10.4254/wjh.v14.i7.1357
Table 1 Tissue pad degradation of H-HD1000i and H-1100 after 300 repeated activations

H-HD1000i
H-1100
1st experiment (mm)0.590.06
2nd experiment (mm)0.420.09
Mean (mm)0.510.08
Table 2 Vessel sealing performance after 300 repeated activations. Values are presented as median (range) or n (%)

H-HD1000i
H-1100
P value
Vessel sealing spead (sec.)5.1 (3.9-5.9)4.9 (3.2-5.2)0.21
Vessel burst pressure < 700 mmHg4 (40)0 (0)0.09