World J Hepatol. Oct 27, 2026; 18(10): 121142
Published online Oct 27, 2026. doi: 10.4254/wjh.121142
Computed tomography volumetry vs three-dimensional surface scanning for liver graft size estimation using explant liver mass
La Ode Irman Jaya, Pawel Rykowski, Mateusz Bartkowiak, Emilia Kruk, Maria Chodurska, Maciej Krasnodębski, Michał Grąt, Department of General, Transplant, and Liver Surgery, Medical University of Warsaw, University Clinical Centre, Warsaw 02-091, Mazowieckie, Poland
Zbigniew Lewandowski, Department of Epidemiology and Biostatistics, Medical University of Warsaw, Center for Simulation and Medical Innovation, Warsaw 02-106, Mazowieckie, Poland
ORCID number: La Ode Irman Jaya (0009-0005-4753-8603); Pawel Rykowski (0000-0002-0519-2932); Mateusz Bartkowiak (0000-0003-4066-9801); Maria Chodurska (0009-0006-8274-7205); Zbigniew Lewandowski (0000-0002-9963-2723); Maciej Krasnodębski (0000-0002-6234-615X); Michał Grąt (0000-0003-3372-3072).
Author contributions: Jaya LOI and Rykowski P designed the study; Jaya LOI, Bartkowiak M, Kruk E, and Chodurska M collected the data; Jaya LOI and Lewandowski Z performed the statistical analysis; Krasnodębski M and Grąt M supervised the study; Jaya LOI drafted the manuscript. All authors revised the manuscript and approved the final version.
AI contribution statement: AI tools, specifically ChatGPT and Grammarly, were used during manuscript preparation solely for language polishing and grammatical correction to improve clarity and readability. The authors are non-native English speakers, and these tools were used to assess the linguistic quality of the manuscript. The manuscript, including all scientific content such as the study design, data analysis, results, and interpretation, was entirely developed by the authors, and no part of the scientific content was generated by AI. Lastly, no image or figures included in this manuscript were generated by AI.
Institutional review board statement: This study was reviewed by the Bioethics Committee of the Medical University of Warsaw.
Informed consent statement: Patients were not required to give informed consent to the study because the analysis used anonymous clinical data that were obtained after each patient agreed to treatment by written consent.
Conflict-of-interest statement: All the authors report no relevant conflicts of interest for this article.
STROBE statement: The authors have read the STROBE Statement-checklist of items, and the manuscript was prepared and revised according to the STROBE Statement-checklist of items.
Data sharing statement: No additional data is available.
Corresponding author: Pawel Rykowski, Department of General, Transplant, and Liver Surgery, Medical University of Warsaw, University Clinical Centre, Stefana Banacha 1a, Warsaw 02-091, Mazowieckie, Poland. pawel.rykowski@wum.edu.pl
Received: March 17, 2026
Revised: April 30, 2026
Accepted: September 10, 2026
Published online: October 27, 2026
Processing time: 218 Days and 13.2 Hours

Abstract
BACKGROUND

Accurate estimation of graft size is essential in liver transplantation to ensure adequate functional liver mass. Computed tomography (CT) volumetry is widely used for this purpose; however, alternative imaging approaches, such as three-dimensional (3D) surface scanning, may provide complementary advantages.

AIM

To compare CT volumetry and 3D surface scanning for liver graft size estimation, using explant liver mass as the reference standard.

METHODS

A total of 19 liver explants were analyzed. CT volumetry and 3D surface-derived volumetry were quantitatively compared with the explant liver mass. Correlation analysis, regression-based calibration, absolute error, and reliability [intraclass correlation coefficient (ICC)] were assessed.

RESULTS

Both methods showed strong correlation with explant liver mass (r = 0.94 for 3D surface volumetry and r = 0.938 for CT volumetry, both P < 0.001). Regression analysis demonstrated proportional calibration for both methods, with slopes of 1.01 (R2 = 0.761) for 3D surface volumetry and 1.04 (R2 = 0.820) for CT volumetry. CT volumetry showed significantly lower absolute estimation error compared with 3D surface volumetry (P = 0.011). Reliability analysis revealed excellent agreement between CT volumetry and explant liver mass (ICC = 0.90) and good-to-excellent agreement for 3D surface volumetry (ICC = 0.87).

CONCLUSION

Both CT volumetry and 3D surface scanning provide quantitatively consistent estimates of liver graft size. However, CT volumetry remains more precise, while 3D surface scanning may serve as a complementary method in this pilot methodological study.

Key Words: Computed tomography volumetry; Three-dimensional scanning; Liver transplantation; Liver graft; Correlation; Imaging

Core Tip: Liver transplantation requires accurate estimation of graft size to optimize postoperative outcomes, with computed tomography (CT) volumetry currently the clinical standard. In this pilot methodological validation study, CT volumetry was compared with three-dimensional surface scanning using explant liver mass as the reference standard. While CT volumetry demonstrated higher precision, three-dimensional surface scanning showed strong correlation and good measurement reliability, supporting its potential role as a complementary method for ex vivo liver volumetry.



INTRODUCTION

Liver transplantation is a life-saving treatment for patients with acute and chronic end-stage liver disease, offering substantial gains in survival and quality of life[1-3]. Recent syntheses estimate that liver disease causes approximately 2 million deaths annually - about 4% of all deaths worldwide - largely from cirrhosis and hepatocellular carcinoma, confirming its status as a major global cause of mortality[4-6]. In addition, chronic liver diseases, such as hepatitis B virus infection, hepatitis C virus infection, metabolic dysfunction-associated steatosis liver disease, and cirrhosis, remain highly prevalent. Data from the Global Burden of Disease Study 2021 demonstrate age-standardized prevalence rates per 100000 population of 3584 for hepatitis B virus, 1718 for hepatitis C virus, 15018 for metabolic dysfunction-associated steatosis liver disease, and 20303 for cirrhosis[4,7]. Furthermore, contemporary registry and cohort studies consistently demonstrate favorable short- and medium-term outcomes after deceased-donor liver transplantation, with overall patient survival rates approaching 90% at 1 year and 75%-80% at 5 years, even among high-acuity recipients[2,8-10].

In liver transplantation, computed tomography (CT)-based volumetry is the gold standard for preoperative estimation of liver graft volume, as numerous studies have established its strong correlation with actual graft weight and its role in predicting perioperative complications associated with graft size[11,12]. However, CT volumetry is sensitive to imaging parameters and segmentation practices, including contrast phase, slice thickness, and operator experience, and may be unavailable when donors are referred from centers without routine abdominal CT imaging, thereby limiting pre-procurement volumetric assessment. These limitations have motivated the exploration of alternative volumetric approaches[13-15]. A recent systematic evaluation of mobile three-dimensional (3D) scanning applications, including models generated using Polycam (Polycam, Inc., San Francisco, CA, United States), has reported substantial variability in reconstruction accuracy across anatomical models. For example, reconstruction of a breast phantom showed that the 3D scanner application exhibited the highest mean error (21.4 mm) and the greatest variability among the tested platforms[16]. In practical workflows, reconstructed surface meshes are routinely post-processed using Blender (Blender Foundation, Amsterdam, the Netherlands) to remove artifacts, apply smoothing, standardize geometry, and improve measurement consistency[17]. Preliminary studies using Light Detection and Ranging (known as LiDAR)-based 3D scanning for ex vivo liver grafts have demonstrated high correlation with actual graft weight, with a mean error of 17.3 g and a mean accuracy of 98%[18].

Accurate assessment of graft size is critical in liver transplantation. However, most previous studies have focused on the correlation between imaging-derived liver volumes and graft weight, which does not fully capture clinical agreement, calibration, or measurement reliability. Comparative evaluations of CT-based volumetry methods against intraoperative graft weight have not only demonstrated good correlations but also revealed systematic error, underscoring the need for more comprehensive measurement frameworks[15,19,20]. A clinically critical setting in which accurate volumetric estimation appears to be essential is the large-for-size (LFS) problem, now recognized as an important determinant of early post-transplant outcomes in selected recipients. LFS is not simply “a large graft” but a mismatch between graft volume and the recipient’s abdominal domain and hemodynamic tolerance, potentially predisposing to graft compression, portal vein thrombosis, and early graft dysfunction[21,22].

Given the clinical importance of accurate volumetry, 3D surface-scanning technologies validated in other anatomical regions may provide a foundation for assessing liver grafts. The system has demonstrated high accuracy, low median volume error, and excellent reliability in applications such as upper-limb volumetry for breast cancer-related lymphedema and hand surgery, when benchmarked against water displacement, tape measurements, and high-precision scanners. Furthermore, modern 3D cameras and laser scanners have exhibited sub-millimeter geometric precision and the ability to detect small volumetric changes (approximately 0.5-4 mL) with high repeatability and strong agreement relative to established volumetric reference standards, supporting their broader use in soft-tissue volumetry[23-26].

This study addressed this gap by quantitatively comparing 3D liver volumetry with CT volumetry, using explant liver mass as the reference standard. Given the relatively small sample size, this study should be considered a pilot methodological validation study, and the findings should be interpreted accordingly.

MATERIALS AND METHODS
Study setting and design

This retrospective observational study was conducted at the Department of General, Transplantation, and Liver Surgery, University Clinical Center of the Medical University of Warsaw, Poland. Liver explants obtained during routine liver transplantation from 30 November 2024 to 6 February 2026 were prospectively weighed and underwent 3D surface scanning according to a standardized protocol, while preoperative CT volumetry data were retrieved from routine clinical imaging. Nineteen consecutive liver explants obtained during routine liver transplantation procedures were included. No formal exclusion criteria were predefined, and no specimens required exclusion owing to technical issues (e.g., damage during handling or incomplete imaging data). This study was reviewed by the Bioethics Committee of the Medical University of Warsaw.

Liver explant mass measurement

Immediately after procurement, each liver explant was gently dried with sterile gauze and weighed using a calibrated surgical scale. Explant mass served as the reference standard for all volumetric analyses, based on the assumption that liver tissue density approximates 1 g/cm3, thereby allowing mass to serve as a clinically relevant surrogate for organ volume. However, this approximation may not fully account for variations in tissue composition, perfusion status, and post-explant changes, potentially introducing systematic differences between mass- and volume-based measurements, particularly given the in vivo vs ex vivo conditions under which these measurements were obtained.

3D surface scanning procedure and volumetry measurements

Following mass measurement, each liver explant underwent 3D surface scanning using the Polycam photogrammetry mobile application (iOS/Android) (Polycam, Inc., San Francisco, CA, United States). Multiple circumferential image sweeps were performed to ensure complete surface coverage. Image acquisition was performed under standardized lighting and background conditions. Mesh construction, refinement, and volumetric analysis were performed by an investigator who was blinded to the CT volumetry and explant liver mass measurements. The scans were exported in STL format and reconstructed using Blender version 5.0.0. Mesh refinement included surface smoothing, artifact removal, gap closure, and dimensional calibration. To ensure measurement reliability, volumetric estimates for each 3D mesh were obtained twice per explant (n = 2), and the results were averaged to yield a single representative 3D volume for each sample. This approach minimized operator-dependent variability in mesh selection and region-of-interest definition.

CT volumetry acquisition

Preoperative CT volumetry was obtained from contrast-enhanced abdominal CT scans performed as part of the transplant evaluation protocol. Volumetric measurements were performed using AW Server 3.2 EXT 4.0 (GE Healthcare, Chicago, IL, United States) with the Hepatic VCAR module, which provides semi-automated liver segmentation and quantitative volumetric analysis. Manual corrections were applied when necessary by an experienced radiologist.

Statistical analysis

Continuous variables are presented as mean ± SD or median with interquartile range (IQR), as appropriate. Explant liver mass was used as the reference standard for all analyses. Imaging-based graft size estimates obtained from CT volumetry and 3D surface scanning were initially evaluated for their association with explant liver mass using Pearson correlation analysis. Assumptions underlying the analysis, including linearity and distribution of residuals, were assessed before analysis. Calibration of imaging-based graft size estimates was assessed using linear regression analysis, with explant liver mass as the dependent variable and imaging-derived estimates as independent variables. Regression coefficients, coefficients of determination (R2), and intercepts were evaluated to assess the degree of correspondence and potential systematic offset between estimated graft size and explant mass.

Agreement between imaging-derived estimates and explant liver mass was further evaluated using Bland-Altman analysis. Mean bias and 95% limits of agreement were calculated to assess the magnitude and distribution of differences across the range of graft sizes. Absolute estimation error for each modality was defined as the absolute difference between imaging-based estimates and explant liver mass. Differences in absolute error between CT volumetry and 3D surface volumetry were compared using the Wilcoxon signed-rank test for paired samples.

Reliability and agreement between imaging-derived estimates and explant liver mass were quantified using intraclass correlation coefficients (ICCs). ICC (3, 1) values derived from a two-way fixed-effects model with absolute agreement were reported. This model was selected as it reflects absolute agreement between measurement methods applied to the same subjects in a fixed-effects framework. All statistical analyses were performed using R version 4.5.2 (R Foundation for Statistical Computing, Vienna, Austria). A two-sided P < 0.05 was considered statistically significant.

RESULTS
Baseline characteristics of the study population

In total, 19 liver explants (13 males and 6 females) were included in the study, with baseline demographics summarized in Table 1. The cohort had a median (IQR) age of 54.0 (40.0-63.5) years, height of 172.0 (163.0-181.0) cm, and body weight of 80.0 (65.5-96.0) kg, corresponding to a median (IQR) body mass index of 26.3 (22.7-32.0) kg/m2. Median (IQR) thoracic and abdominal circumferences were 93.0 (85.0-103.5) cm and 97.0 (86.5-107.5) cm, respectively, reflecting baseline demographic and anthropometric characteristics typical of liver transplant candidates. All liver explants were cirrhotic, and all patients underwent elective liver transplantation.

Table 1 Patient demographics, median (interquartile range).
Parameters
Study group, n = 19
Sex
Male13
Female6
Age in years54.0 (40.0-63.5)
Height in cm172.0 (163.0-181.0)
Weight in kg80.0 (65.5-96.0)
BMI in kg/m226.3 (22.7-32.0)
Thoracic circumference93.0 (85.0-103.5)
Abdominal circumference97.0 (86.5-107.5)
Correlation coefficient between 3D and CT volumetry with liver explant mass reference

Explant liver mass demonstrated a strong correlation with 3D surface-derived volumetry [r = 0.94, 95% confidence interval (CI): 0.84-0.98, P < 0.001; n = 19] and with CT-based volumetry (r = 0.938, 95%CI: 0.84-0.98, P < 0.001; n = 19). Corresponding scatter plots are presented in Figure 1.

Figure 1
Figure 1 Comparison of computed tomography volumetry, three-dimensional reconstruction workflow, and corresponding correlation analyses. A: Representative axial computed tomography (CT) image; B: Three-dimensional (3D) liver model reconstructed from CT imaging using semi-automated segmentation and volumetric analysis (AW Server 3.2 EXT 4.0, GE Healthcare; Hepatic VCAR application); C: 3D reconstructed explant model generated from Polycam (photogrammetry-based scanning application) and refined in Blender, representing the surface-based anatomical morphology of the liver specimen; D: Correlation between 3D surface-derived volumetry and liver explant mass; E: Correlation between CT volumetry and explant liver mass. Values are expressed in grams (g) for mass and cubic centimeters (cm3) for volumetric measurements. CT: Computed tomography; 3D: Three-dimensional.
Clinical calibration of imaging-based recipient liver volume estimates

Linear regression analysis was performed with explant liver mass as the dependent variable and 3D surface volumetry as the independent variable. The model demonstrated good fit (R2 = 0.775; adjusted R2 = 0.761), with a residual SE of 248.9 g (n = 19; df = 17). The regression coefficient was β = 1.01 (95%CI: 0.73-1.30; P < 0.001), indicating proportional agreement between 3D surface-derived volumetry and explant liver mass (Table 2).

Table 2 Calibration model of 3D surface volumetry against explant liver mass.
Variable
β coefficient (95%CI)
SE
P value
Intercept299.7 (-129.40 to 728.77)203.380.16
3D surface estimate1.01 (0.73 to 1.29)0.13< 0.001

Linear regression analysis was performed using explant liver mass as the dependent variable and CT volumetry as the independent variable. The model demonstrated good fit (R2 = 0.830; adjusted R2 = 0.820), with a residual SE of 216.4 g (n = 19; df = 17). The estimated regression coefficient was β = 1.04 (95%CI: 0.79-1.29; P < 0.001), indicating proportional agreement between CT-derived volumetry and explant liver mass (Table 3).

Table 3 Calibration model of computed tomography volumetry against explant liver mass.
Variable
β coefficient (95%CI)
SE
P value
Intercept221.86 (-157.08 to 600.79)179.600.234
CT volumetry1.04 (0.80 to 1.28)0.11< 0.001
Agreement between imaging-derived estimates and explant liver mass

Bland-Altman plots comparing imaging-based graft size estimates with explant liver mass are shown in Figure 2. For CT volumetry, the mean bias was -284.1 g, with 95% limits of agreement ranging from -697.9 g to 129.8 g (Figure 2A). For 3D surface-derived volumetry, the mean bias was -315.7 g, with 95% limits of agreement ranging from -789.9 g to 158.4 g (Figure 2B).

Figure 2
Figure 2 Agreement between imaging-based graft size estimates and explant liver mass. Bland-Altman plots showing the differences between imaging-based graft size estimates and explant liver mass plotted against the mean of the measurements. A: Computed tomography volumetry vs explant liver mass; B: Three-dimensional surface-derived volumetry vs explant liver mass. Dashed lines indicate the mean bias and 95% limits of agreement. Values are expressed in grams (g). CT: Computed tomography; 3D: Three-dimensional.
Comparison of absolute estimation error between imaging modalities

Absolute error distributions for CT volumetry and 3D surface-derived volumetry relative to explant liver mass are shown in Figure 3. The median absolute percentage error was 18.45% (IQR: 9.72%) for CT volumetry and 21.26% (IQR: 12.87%) for 3D surface-derived volumetry. The median paired difference in absolute percentage error (3D minus CT) was 2.44% (95%CI: 0.00%-3.9%; Wilcoxon signed-rank test, P = 0.011), indicating significantly greater error for 3D surface-derived volumetry (Table 4).

Figure 3
Figure 3 Absolute error of graft size estimation relative to explant liver mass for computed tomography volumetry and three-dimensional surface scanning. A: Three-dimensional (3D) surface reconstruction of a liver explant; B: Boxplot distribution of absolute percentage error (%) for computed tomography volumetry and 3D surface-derived volumetry; C: 3D reconstruction of the liver from computed tomography imaging used for volumetric analysis. CT: Computed tomography; 3D: Three-dimensional.
Table 4 Accuracy of graft size estimation relative to explant liver mass, n = 19.
Method
Median absolute error (%)
IQR (%)
CT volumetry18.459.72
3D surface-derived volumetry21.2612.87
Paired comparison (3D-CT)
StatisticValue
    Median paired difference (%)2.44
    95%CI0.00-3.9
    Wilcoxon signed-rank P value0.011
ICCs assessed reliability and agreement

ICCs for agreement between imaging-based graft size estimates and explant liver mass are summarized in Table 5. The ICC (3, 1) was 0.90 (95%CI: 0.76-0.96; n = 19) for CT volumetry vs explant mass and 0.87 (95%CI: 0.7-0.95; n = 19) for 3D surface-derived volumetry vs explant liver mass.

Table 5 Intraclass correlation coefficients for agreement between imaging-based graft size estimates and explant liver mass.
Method
ICC (3, 1)
95%CI
n
CT volumetry vs explant liver mass0.900.76-0.9619
3D surface-derived volumetry vs explant liver mass0.870.70-0.9519
DISCUSSION

The study cohort consisted of adult transplant recipients with moderate body size and normal variability in body measurements, reflecting patients typically encountered in routine clinical practice. This is particularly relevant to LFS syndrome analysis, because graft size matching inherently depends on the recipient’s anatomy and the available intra-abdominal space. Therefore, volumetric estimates should be interpreted in the context of the recipient’s anatomical constraints and physiological tolerance, underscoring the need for consistent and reliable techniques for assessing graft size. Although the cohort reflects a typical transplant population, given the small sample size, the findings should be considered exploratory and primarily methodological.

CT volumetry exhibited a strong association with liver explant mass, consistent with its established role as the reference standard for clinical graft volumetry. Prior studies have shown that 3D CT-based volumetry not only improves the accuracy of graft size estimation compared with traditional 2D techniques but also provides more consistent information for size matching in liver transplantation[27]. The observed strong performance of 3D surface-derived volumetry in correlating with explant mass emphasizes its potential as an accessible complementary method[28]. This approach may be particularly advantageous within multimodal 3D planning frameworks that integrate internal CT-derived anatomy with external surface models, thereby supporting surgical simulation, preoperative planning, and education.

Calibration analyses indicated that both approaches maintained appropriate proportional scaling with explant liver mass, as evidenced by regression slopes approaching 1.0 and minimal systematic bias. This behavior suggests that both techniques yield consistent quantitative estimates across the evaluated graft-size spectrum. From a clinical perspective, preservation of proportional calibration is critical, because even modest systematic deviations at the extremes of graft size may influence graft selection and perioperative planning. Collectively, these results support the use of both imaging modalities for consistent graft size estimation, extending their utility beyond simple relative comparisons.

Measurement differences relative to explant liver mass showed moderate deviation for both modalities. CT-based estimates exhibited narrower dispersion, while surface-based measurements displayed greater variability, particularly at higher volumes (Figure 2). Notably, error magnitude did not systematically increase with graft size, arguing against the presence of substantial proportional bias. From a clinical perspective, dispersion rather than mean bias often determines the utility of a measurement method for individual patients. Accordingly, the narrower limits of agreement observed with CT-based volumetry indicate greater measurement precision compared with surface-based methods. At the same time, the lack of progressive divergence across the graft-size range demonstrates that surface-based scanning yields stable, size-independent estimates, supporting its use as a complementary or confirmatory tool rather than indicating inconsistent performance. This pattern is consistent with findings from CT-based liver volumetry studies, where CT methods demonstrate strong agreement with actual graft weight but retain non-negligible dispersion on Bland-Altman analysis, with differences in precision often being more clinically relevant than mean bias.

For example, Mohapatra et al[15] reported similar IQR percentage errors for conventional CT and 3D liver volumetry, with slightly narrower error distribution observed for 3D methods, thereby improving overall accuracy without eliminating residual variability. Likewise, Park et al[29] demonstrated minimal mean bias between deep learning-assisted CT volumetry and intraoperative graft weight, with clinical performance largely determined by the width of the 95% limits of agreement rather than the average difference. In addition, previous studies have reported that 3D surface-based liver models can achieve clinically acceptable agreement with CT-derived liver volumes. Witowski et al[30] reported that surface-reconstructed, 3D printed liver models yield volumetric measurements consistent with CT volumetry, with Bland-Altman analysis showing no significant systematic bias and only modest volume differences, supporting their reliability for routine preoperative assessment.

CT volumetry demonstrated lower variability and narrower error dispersion compared with 3D surface scanning (Figure 3). Conceptually, these two modalities assess the liver under fundamentally different physiological conditions. CT volumetry represents the in vivo organ, with its apparent volume affected by perfusion, venous filling, and the timing of image acquisition, whereas 3D surface scanning is usually performed ex vivo on a graft that is no longer perfused and may have been flushed and drained. As a result, systematic differences between these modalities can be anticipated a priori, even when segmentation and reconstruction are performed with high technical accuracy. Although CT volumetry exhibited slightly lower variability compared with 3D surface scanning, the magnitude of disagreement between techniques remained within a range that may be acceptable for routine graft size estimation. Thus, the key consideration is not only the strength of correlation but also whether the level of agreement is sufficient to support specific clinical decisions.

Both CT and 3D surface volumetry demonstrated high reliability for estimating explant mass, as summarized in Table 3. While CT volumetry achieved slightly superior agreement, 3D surface scanning yielded consistent and clinically acceptable results. These findings support the integration of 3D surface scanning as a complementary modality to CT for quantitative graft evaluation. Accurate volumetric assessment is essential for graft-recipient size matching and perioperative planning, and the observed concordance reinforces its role in multimodal 3D surgical workflows.

Several limitations of this study should be acknowledged. First, the sample size was relatively small, reflecting the exploratory and pilot nature of this analysis; therefore, the findings require validation in larger cohorts and with alternative scanning methods. Second, explant liver mass was used as the reference standard rather than true graft volume; however, it represents a direct and clinically meaningful surrogate for graft size assessment (Figure 2). Finally, the analysis focused on methodological performance and did not evaluate post-transplant clinical outcomes, which will be important to address in future studies to determine the clinical impact of these measurement approaches.

CONCLUSION

Accurate graft size estimation remains a critical component of liver transplantation. Both CT volumetry and 3D surface scanning provided graft size estimates consistent with explant liver mass, demonstrating quantitative consistency within the studied cohort. CT volumetry exhibited greater precision and consistency, supporting its established role as the reference standard for assessing graft size. Nevertheless, 3D surface scanning showed reliable performance and acceptable agreement, suggesting its potential utility as a complementary approach in selected transplant settings. These findings provide a methodological basis for future studies evaluating imaging-based assessment of graft size within the liver transplantation workflow; however, they should be interpreted with caution, given the limited sample size and the exploratory nature of the study.

ACKNOWLEDGEMENTS

The authors thank the transplant surgical team and radiology staff at the University Clinical Center of the Medical University of Warsaw for their support in data acquisition and technical assistance. The authors also acknowledge the contributions of all personnel involved in graft handling and data management.

References
1.  Grąt M, Wronka KM, Stypułkowski J, Bik E, Krasnodębski M, Masior Ł, Lewandowski Z, Grąt K, Patkowski W, Krawczyk M. The Warsaw Proposal for the Use of Extended Selection Criteria in Liver Transplantation for Hepatocellular Cancer. Ann Surg Oncol. 2017;24:526-534.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 25]  [Cited by in RCA: 27]  [Article Influence: 3.0]  [Reference Citation Analysis (1)]
2.  Jiménez-Romero C, Marcacuzco A, Caso O, Manrique A, García-Sesma A, Calvo J, Nutu A, Moreno-González E, Bernaldo de Quirós M, Justo Alonso I. Long-Term Outcomes of Liver Transplant Recipients. What Do Patients Die From? World J Surg. 2025;49:1632-1642.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 4]  [Reference Citation Analysis (4)]
3.  Karvellas CJ, Leventhal TM, Rakela JL, Zhang J, Durkalski V, Reddy KR, Fontana RJ, Stravitz RT, Lake JR, Lee WM, Parekh JR. Outcomes of patients with acute liver failure listed for liver transplantation: A multicenter prospective cohort analysis. Liver Transpl. 2023;29:318-330.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 71]  [Cited by in RCA: 64]  [Article Influence: 21.3]  [Reference Citation Analysis (2)]
4.  Feng G, Yilmaz Y, Valenti L, Seto WK, Pan CQ, Méndez-Sánchez N, Ye F, Sookoian S, Targher G, Byrne CD, Chan WK, Treeprasertsuk S, Yu HH, Kim SU, George J, Xu WJ, Sebastiani G, Ocama P, Ryan JD, Lupșor-Platon M, Ghazinyan H, Hamid S, Perera N, Alswat K, Isakov V, Pan Q, Sarin SK, Schattenberg JM, Sotoudeheian M, Wong YJ, Sharara AI, Al-Busafi SA, Opio CK, Chai J, Fouad Y, Shi Y, Al-Mahtab M, Zhang S, Pirola CJ, Wong VW, Zheng MH. Global Burden of Major Chronic Liver Diseases in 2021. Liver Int. 2025;45:e70058.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 16]  [Cited by in RCA: 23]  [Article Influence: 23.0]  [Reference Citation Analysis (0)]
5.  Asrani SK, Devarbhavi H, Eaton J, Kamath PS. Burden of liver diseases in the world. J Hepatol. 2019;70:151-171.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 3074]  [Cited by in RCA: 2639]  [Article Influence: 377.0]  [Reference Citation Analysis (10)]
6.  Gan C, Yuan Y, Shen H, Gao J, Kong X, Che Z, Guo Y, Wang H, Dong E, Xiao J. Liver diseases: epidemiology, causes, trends and predictions. Signal Transduct Target Ther. 2025;10:33.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 290]  [Cited by in RCA: 204]  [Article Influence: 204.0]  [Reference Citation Analysis (2)]
7.  Devarbhavi H, Asrani SK, Arab JP, Nartey YA, Pose E, Kamath PS. Global burden of liver disease: 2023 update. J Hepatol. 2023;79:516-537.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 1738]  [Cited by in RCA: 1644]  [Article Influence: 548.0]  [Reference Citation Analysis (19)]
8.  Evans MD, Diaz J, Adamusiak AM, Pruett TL, Kirchner VA, Kandaswamy R, Humphreville VR, Leventhal TM, Grosland JO, Vock DM, Matas AJ, Chinnakotla S. Predictors of Survival After Liver Transplantation in Patients With the Highest Acuity (MELD ≥40). Ann Surg. 2020;272:458-466.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 26]  [Cited by in RCA: 19]  [Article Influence: 3.2]  [Reference Citation Analysis (0)]
9.  Kaltenmeier C, Jorgensen D, Dharmayan S, Ayloo S, Rachakonda V, Geller DA, Tohme S, Molinari M. The liver transplant risk score prognosticates the outcomes of liver transplant recipients at listing. HPB (Oxford). 2021;23:927-936.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 13]  [Cited by in RCA: 15]  [Article Influence: 3.0]  [Reference Citation Analysis (2)]
10.  Bernal W, Bittermann T, Sheth R, Trovato F, McPhail MJ, Khan P, Calvert S, Pirani T, Patel S, Loveridge R, Willars C, Auzinger G, Wendon JA. Survival With and Without Liver Transplantation in Critically Ill Patients With Cirrhosis: A 20-Year Experience. Clin Gastroenterol Hepatol. 2026;24:432-442.e8.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 4]  [Reference Citation Analysis (1)]
11.  Choi E, Kim SH. Optimizing accuracy: a comparative analysis of preoperative liver volumetry in living donor liver transplantation from a surgeon's perspective - a retrospective cohort study. Int J Surg. 2025;111:8149-8158.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 1]  [Cited by in RCA: 1]  [Article Influence: 1.0]  [Reference Citation Analysis (0)]
12.  Tamulevicius M, Oezcelik A, Koitka S, Theysohn JM, Hoyer DP, Farzaliyev F, Haubold J, Nensa F, Treckmann J, Malamutmann E. Preoperative Computed Tomography Volumetry and Graft Weight Estimation of Left Lateral Segment in Pediatric Living Donor Liver Transplant. Exp Clin Transplant. 2023;21:831-836.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 2]  [Reference Citation Analysis (2)]
13.  Çelik H, Odaman H, Altay C, Ünek T, Özbilgin M, Egeli T, Ağalar C, Astarcıoğlu İK, Barlık F. Manual and semi-automated computed tomography volumetry significantly overestimates the right liver lobe graft weight: a single-center study with adult living liver donors. Diagn Interv Radiol. 2024;30:3-8.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 4]  [Cited by in RCA: 6]  [Article Influence: 3.0]  [Reference Citation Analysis (0)]
14.  Kalshabay Y, Zholdybay Z, Di Martino M, Medeubekov U, Baiguissova D, Ainakulova A, Doskhanov M, Baimakhanov B. CT volume analysis in living donor liver transplantation: accuracy of three different approaches. Insights Imaging. 2023;14:82.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 17]  [Cited by in RCA: 15]  [Article Influence: 5.0]  [Reference Citation Analysis (4)]
15.  Mohapatra N, Gurumoorthy Subramanya Bharathy K, Kumar Sinha P, Vasantrao Sasturkar S, Patidar Y, Pamecha V. Three-Dimensional Volumetric Assessment of Graft Volume in Living Donor Liver Transplantation: Does It Minimise Errors of Estimation? J Clin Exp Hepatol. 2020;10:1-8.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 18]  [Cited by in RCA: 21]  [Article Influence: 3.5]  [Reference Citation Analysis (2)]
16.  Botti E, Jansen B, Ballen-Moreno F, Kapila A, Brahimetaj R. Evaluating the Accuracy and Repeatability of Mobile 3D Imaging Applications for Breast Phantom Reconstruction. Sensors (Basel). 2025;25:4596.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in RCA: 4]  [Reference Citation Analysis (0)]
17.  Kimura T, Takiguchi K, Tsukita S, Muto M, Chiba H, Sato N, Kofunato Y, Ishigame T, Kenjo A, Tanaka H, Marubashi S. Development of anatomically accurate digital organ models for surgical simulation and training. PLoS One. 2025;20:e0320816.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 7]  [Cited by in RCA: 5]  [Article Influence: 5.0]  [Reference Citation Analysis (1)]
18.  Katsanos G, Karakasi KE, Karolos IA, Kofinas A, Antoniadis N, Tsioukas V, Tsoulfas G. Volumetric assessment of hepatic grafts using a light detection and ranging system for 3D scanning: Preliminary data. World J Hepatol. 2022;14:1504-1511.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 1]  [Cited by in RCA: 3]  [Article Influence: 0.8]  [Reference Citation Analysis (2)]
19.  Bozkurt B, Emek E, Arikan T, Ceyhan O, Yazici P, Sahin T, Mammadov E, Serin A, Gurcan NI, Yuzer Y, Tokat Y. Liver Graft Volume Estimation by Manual Volumetry and Software-Aided Interactive Volumetry: Which is Better? Transplant Proc. 2019;51:2387-2390.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 13]  [Cited by in RCA: 11]  [Article Influence: 1.6]  [Reference Citation Analysis (1)]
20.  Frericks BB, Caldarone FC, Nashan B, Savellano DH, Stamm G, Kirchhoff TD, Shin HO, Schenk A, Selle D, Spindler W, Klempnauer J, Peitgen HO, Galanski M. 3D CT modeling of hepatic vessel architecture and volume calculation in living donated liver transplantation. Eur Radiol. 2004;14:326-333.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 63]  [Cited by in RCA: 56]  [Article Influence: 2.5]  [Reference Citation Analysis (1)]
21.  Kostakis ID, Raptis DA, Davidson BR, Iype S, Nasralla D, Imber C, Sharma D, Pissanou T, Pollok JM. Donor-Recipient Body Surface Area Mismatch and the Outcome of Liver Transplantation in the UK. Prog Transplant. 2023;33:61-68.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 8]  [Reference Citation Analysis (1)]
22.  Nakayama T, Akabane M, Imaoka Y, Esquivel CO, Melcher ML, Sasaki K. Revisiting the Prognostic Influences of Donor-Recipient Size Mismatch in Deceased Donor Liver Transplantation. Transplant Direct. 2024;10:e1722.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in RCA: 11]  [Reference Citation Analysis (3)]
23.  Cutti AG, Santi MG, Hansen AH, Fatone S; Residual Limb Shape Capture Group. Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics. Sensors (Basel). 2024;24:2350.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 7]  [Reference Citation Analysis (1)]
24.  de Sire A, Losco L, Cigna E, Lippi L, Gimigliano F, Gennari A, Cisari C, Chen HC, Fusco N, Invernizzi M. Three-dimensional laser scanning as a reliable and reproducible diagnostic tool in breast cancer related lymphedema rehabilitation: a proof-of-principle study. Eur Rev Med Pharmacol Sci. 2020;24:4476-4485.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 23]  [Reference Citation Analysis (2)]
25.  Mastick J, Smoot BJ, Paul SM, Kober KM, Cooper BA, Madden LK, Conley YP, Dixit N, Hammer MJ, Fu MR, Piper M, Cate SP, Shepherd J, Miaskowski C. Assessment of Arm Volume Using a Tape Measure Versus a 3D Optical Scanner in Survivors with Breast Cancer-Related Lymphedema. Lymphat Res Biol. 2022;20:39-47.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 5]  [Cited by in RCA: 11]  [Article Influence: 2.8]  [Reference Citation Analysis (1)]
26.  White BN, Lu IM, Kao LS, Dixon JB, Weiler MJ, Frank ND, Binkley J, Subhedar P, Okoli J, Buhariwalla K, Suarez-Ligon A, Gabram-Mendola SGA. An infrared 3D scanning device as a novel limb volume measurement tool in breast cancer patients. World J Surg Oncol. 2020;18:278.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 6]  [Cited by in RCA: 11]  [Article Influence: 1.8]  [Reference Citation Analysis (1)]
27.  Banchini F, Capelli P, Hasnaoui A, Palmieri G, Romboli A, Giuffrida M. 3-D reconstruction in liver surgery: a systematic review. HPB (Oxford). 2024;26:1205-1215.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in RCA: 12]  [Reference Citation Analysis (1)]
28.  Portnoy Y, Koren J, Khoury A, Factor S, Dadia S, Ran Y, Benady A. Three-dimensional technologies in presurgical planning of bone surgeries: current evidence and future perspectives. Int J Surg. 2023;109:3-10.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 3]  [Cited by in RCA: 35]  [Article Influence: 11.7]  [Reference Citation Analysis (3)]
29.  Park R, Lee S, Sung Y, Yoon J, Suk HI, Kim H, Choi S. Accuracy and Efficiency of Right-Lobe Graft Weight Estimation Using Deep-Learning-Assisted CT Volumetry for Living-Donor Liver Transplantation. Diagnostics (Basel). 2022;12:590.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Full Text (PDF)]  [Cited by in Crossref: 18]  [Cited by in RCA: 19]  [Article Influence: 4.8]  [Reference Citation Analysis (2)]
30.  Witowski J, Wake N, Grochowska A, Sun Z, Budzyński A, Major P, Popiela TJ, Pędziwiatr M. Investigating accuracy of 3D printed liver models with computed tomography. Quant Imaging Med Surg. 2019;9:43-52.  [RCA]  [PubMed]  [DOI]  [Full Text]  [Cited by in Crossref: 24]  [Cited by in RCA: 39]  [Article Influence: 5.6]  [Reference Citation Analysis (3)]
Footnotes

Peer review: Externally peer reviewed.

Peer-review model: Single blind

Specialty type: Gastroenterology and hepatology

Country of origin: Poland

Peer-review report’s classification

Scientific quality: Grade B, Grade C, Grade C

Novelty: Grade B, Grade C, Grade C

Creativity or innovation: Grade B, Grade C, Grade C

Scientific significance: Grade B, Grade C, Grade C

P-Reviewer: Arumugam VA, PhD, Professor, India; Yao QH, Assistant Professor, PhD, China S-Editor: Hu XY L-Editor: A P-Editor: Yang YQ

Write to the Help Desk