Table 1 Characteristics of the tested video capsule endoscopes
| Video capsule endoscope | Target organ | Cameras | Permanent magnet | Weight (g) | Length (mm) | Length of internal metal parts (mm) | Width (mm) |
| PillCam SB3 | Small bowel | 1 | No | 2.97 | 26.0 | 18.22 | 11.4 |
| OMOM HD | Small bowel | 1 | No | 3.188 | 26.1 | 19.53 | 11.5 |
| MiroCam 1200 | Small bowel | 1 | No | 3.351 | 24.2 | 18.82 | 11.3 |
| CapsoCam Plus | Small bowel | 4 | No | 3.727 | 30.3 | 25.44 | 11.4 |
| MiroCam 2000 | Small bowel | 2 | No | 3.634 | 30.0 | 20.23 | 11.2 |
| PillCam Colon2 | Colon | 2 | No | 2.944 | 32.2 | 21.54 | 11.6 |
| OMOM CC | Colon | 2 | No | 2.956 | 33.3 | 23.42 | 11.7 |
| NaviCam Colon | Colon | 2 | No | 2.821 | 31.0 | 24.17 | 11.5 |
| NaviCam | Stomach, small bowel | 1 | Yes | 4.653 | 27.0 | 18.15 | 11.5 |
| OMOM RC | Stomach | 1 | Yes | 5.84 | 30.5 | 23.73 | 11.5 |
| MiroCam Navi | Stomach | 1 | Yes | 4.756 | 25.6 | 20.45 | 11.2 |
Table 2 Calculated gravitational force and magnetically induced displacement force on the tested video capsule endoscopes
| Video capsule endoscope | Gravitational force (mN) | Magnetically induced displacement force (mN) | F mag < F grav |
| PillCam SB3 | 29.14 | 288.27 | Not fulfilled |
| OMOM HD | 31.27 | 47.91 | Not fulfilled |
| MiroCam 1200 | 32.87 | 571.96 | Not fulfilled |
| CapsoCam Plus | 36.56 | 672.66 | Not fulfilled |
| MiroCam 2000 | 35.65 | 575.92 | Not fulfilled |
| PillCam Colon2 | 28.88 | 299.57 | Not fulfilled |
| OMOM CC | 29.00 | 127.5 | Not fulfilled |
| NaviCam Colon | 27.67 | 21.96 | Fulfilled |
| NaviCam | 45.65 | 2232 | Not fulfilled |
| OMOM RC | 57.29 | 2366 | Not fulfilled |
| MiroCam Navi | 46.66 | 2415 | Not fulfilled |
Table 3 Magnetically induced torque on the tested video capsule endoscopes
| Video capsule endoscope | Max magnetically induced torque (mNm) | Calculated upper bound of magnetically induced torque (mNm) | Worst-case torque due to gravity (mNm) | T mag < T grav |
| PillCam SB3 | 19.2 | Not determined | Approximately 0.76 | Not fulfilled |
| OMOM HD | 5.1 | Not determined | Approximately 0.82 | Not fulfilled |
| MiroCam 1200 | 41.7 | Not determined | Approximately 0.80 | Not fulfilled |
| CapsoCam Plus | 51.0 | Not determined | Approximately 1.11 | Not fulfilled |
| MiroCam 2000 | 54.0 | Not determined | Approximately 1.07 | Not fulfilled |
| PillCam Colon2 | 15.6 | Not determined | Approximately 0.93 | Not fulfilled |
| OMOM CC | 5.1 | Not determined | Approximately 0.97 | Not fulfilled |
| NaviCam Colon | Below measuring limit | Not determined | Approximately 0.86 | Fulfilled |
| NaviCam | Not determined | 271.95 | Approximately 1.23 | Not fulfilled |
| OMOM RC | Not determined | 289.02 | Approximately 1.75 | Not fulfilled |
| MiroCam Navi | Not determined | 293.90 | Approximately 1.19 | Not fulfilled |
- Citation: Beck MJ, Keuchel M, Paschen J, Laser RM, Roling J, Bandorski D, Windhorst AC, Kurniawan N, Baltes P. Magnetic resonance imaging-induced forces on video capsule endoscopes: First systematic in vitro assessment. World J Gastroenterol 2027; 33(1): 122550
- URL: https://www.wjgnet.com/1007-9327/full/v33/i1/122550.htm
- DOI: https://dx.doi.org/10.3748/wjg.122550