Copyright
©The Author(s) 2017.
World J Gastroenterol. Aug 7, 2017; 23(29): 5422-5430
Published online Aug 7, 2017. doi: 10.3748/wjg.v23.i29.5422
Published online Aug 7, 2017. doi: 10.3748/wjg.v23.i29.5422
Table 1 Patient and lesion baseline data in the S and F1-10 methods
S method (n = 6) | F1-10 method (n = 6) | P3 value | |
Age, median (yr) (range) | 70 (64-87) | 71 (53-79) | 0.81 |
Sex (male/female) | 5:1 | 4:2 | 0.59 |
Lesion location, U/M/L1 | 3/3/0 | 4/2/0 | 0.64 |
Histology, well/mod/poor | 6/0/0 | 6/0/0 | 1.00 |
Tumor size, median (mm) (range) | 13.5 (8-36) | 16 (6-23) | 0.60 |
Resected size, median (mm) (range) | 50.5 (33-64) | 44.5 (31-56) | 0.63 |
Depth of tumor invasion, M/SM1/SM22 | 4/2/0 | 3/1/2 | 0.42 |
Median time of treatment (min) | 65 | 52.5 | 0.69 |
Postoperative bleeding rate | 0 | 0 | 1.00 |
En bloc resection rate | 100% | 100% | 1.00 |
Table 2 Results of the vessel diameter, frequency of the compressed vessel, and coagulation time in the vessel-sealing with endoknife compression technique during gastric endoscopic submucosal dissection
S method | F1-10 method | P1 value | |
Number of vessels processed by vessel-sealing | 49 | 62 | |
with endoknife compression | |||
Median vessel diameter (mm) (range) | 2 (1-3) | 1.5 (1-3) | 0.68 |
Median frequency of compressed vessel (times) (range) | 2 (1-8) | 2 (1-5) | 0.11 |
Median time of coagulation (s) (range) | 9 (2-48) | 10 (1-58) | 0.11 |
Table 3 Rates of bleeding after vessel-sealing with endoknife compression for all vessels and large vessels
Table 4 Electrical condition using the FlushKnife-BT in the porcine ex vivo model analyzed by the VIO DOKU
FlushKnife-BT 2.5 mm | ||
S method | F1-10 method | |
Median peak voltage (Vp) | 172 (170-173) | 560 (538-587) |
Median current elapse time (s) | 0.22 (0.20-0.30) | 1.54 (0.50-1.77) |
Median total electric energy (W) | 6.34 (3.9-8.67) | 12.5 (3.47-13.8) |
- Citation: Ishida T, Toyonaga T, Ohara Y, Nakashige T, Kitamura Y, Ariyoshi R, Takihara H, Baba S, Yoshizaki T, Kawara F, Tanaka S, Morita Y, Umegaki E, Hoshi N, Azuma T. Efficacy of forced coagulation with low high-frequency power setting during endoscopic submucosal dissection. World J Gastroenterol 2017; 23(29): 5422-5430
- URL: https://www.wjgnet.com/1007-9327/full/v23/i29/5422.htm
- DOI: https://dx.doi.org/10.3748/wjg.v23.i29.5422