World J Gastroenterol. Oct 21, 2014; 20(39): 14472-14478
Published online Oct 21, 2014. doi: 10.3748/wjg.v20.i39.14472
Published online Oct 21, 2014. doi: 10.3748/wjg.v20.i39.14472
Table 1 Institutions and patients included
| Institution | n |
| Hospital de Navarra | 8 |
| Clinica Universidad de Navarra | 5 |
| Hospital Gregorio Marañon | 10 |
| Hospital Juan Canalejo | 10 |
| Hospital Puerta de Hierro | 9 |
| Hospital del Mar | 8 |
| Complejo Hospitalario de Toledo | 9 |
| Hospital La Fe | 10 |
| Hospital General de Valencia | 10 |
| Mutua de Tarrassa | 10 |
| Total | 89 |
Table 2 PillCam SB24 specifications
| Physical dimensions | 26 mm × 11 mm |
| Weight | 2.89 ± 0.15 g |
| Number of imaging heads | 1 |
| Imager type | CMOS |
| Image capture rate | 4 frames per second |
| Optical illumination | 4 white light emitting diodes with ALC |
| Field of view | 156° - working distance of 4.5 mm |
| Depth of field | 0-30 mm |
| Mucosa area images at 4.5 mm | Approximately 1100 mm2 |
| Magnification | 1.8 |
| Min. detectable object size | 0.1 mm |
| Min. operation time | 7 h and 58 min |
| Max. operation time | 9:00 h |
| Shelf life | 10 mo after manufacturing date |
| Transmitter frequency | 434.1 MHz |
| Battery type | Silver oxide, non-toxic |
| Storage temperature | 0°-30° |
| Activation | Magnetic, automatic from blister |
Table 3 Overall number of lesions and abnormal frames detected n (%)
| Lesions | Frames | |
| Total | Total | |
| Esophagus | 17 (6.6) | 132 (4) |
| Stomach | 30 (11.6) | 617 (19.1) |
| Small bowel | 193 (75) | 2227 (69) |
| Colon | 17 (6.6) | 249 (7.7) |
| Total | 257 (100) | 3225 (100) |
Table 4 Comparative analysis of landmark detection n (%)
| SB24 patients | “SB2-like” patients | P value | K value | |
| Z-line | 53 (59.5) | 45 (50.6) | NS | 0.820 |
| Papilla | 7 (7.9) | 6 (6.7) | NS | 0.917 |
Table 5 Comparative analysis of frames per landmark n (%)
| SB24 | "SB2-like" | P value | |||
| Frames | Mean ± SD | Frames | Mean ± SD | ||
| Z-line | 616 | 11.6 ± 20.4 | 391 | 8.6 ± 10.6 | NS |
| Papilla | 35 | 5.0 ± 2.0 | 15 | 2.5 ± 1.4 | < 0.05 |
Table 6 Comparative analysis of quality of images detected
| Quality1 | SB24 | “SB2-like” | P value |
| Mean ± SD | Mean ± SD | ||
| 91.94 ± 21.82 | 88.01 ± 34.87 | NS |
Table 7 Lesions detection: Per-patient analysis n (%)
Table 8 Lesions detection: Per-lesion analysis
| SB24 | “SB2-like” | P value | |||
| Lesions | Mean ± SD | Lesions | Mean ± SD | ||
| Esophagus | 17 | 0.2 ± 0.4 | 14 | 0.1 ± 0.4 | NS |
| Stomach | 30 | 0.3 ± 0.8 | 30 | 0.3 ± 0.8 | NS |
| Small bowel | 193 | 2.2 ± 3.1 | 183 | 2.1 ± 2.8 | NS |
| Colon | 17 | 0.2 ± 0.6 | 17 | 0.2 ± 0.6 | NS |
| Total | 257 | 2.9 ± 3.8 | 244 | 2.7 ± 3.5 | NS |
Table 9 Comparative analysis of frames per lesion
| SB24 | “SB2-like” | P value | |||
| Frames | Mean ± SD | Frames | Mean ± SD | ||
| Esophagus | 132 | 7.7 ± 9.0 | 91 | 6.5 ± 8.7 | NS |
| Stomach | 617 | 20.5 ± 34.2 | 403 | 13.4 ± 25.8 | NS |
| Small Bowel | 2227 | 11.5 ± 28.1 | 1490 | 8.1 ± 18.1 | NS |
| Colon | 249 | 14.6 ± 18.2 | 155 | 9.1 ± 12.4 | NS |
| Total | 3225 | 12.8 ± 15.2 | 2139 | 8.7 ± 15.6 | NS |
- Citation: Fernandez-Urien I, Carretero C, Borobio E, Borda A, Estevez E, Galter S, Gonzalez-Suarez B, Gonzalez B, Lujan M, Martinez JL, Martínez V, Menchén P, Navajas J, Pons V, Prieto C, Valle J. Capsule endoscopy capture rate: Has 4 frames-per-second any impact over 2 frames-per-second? World J Gastroenterol 2014; 20(39): 14472-14478