Table 1 Minimum technical specifications for pancreas computed tomography
| Feature | Specification | Comment |
| Scanner type | Multi-detector row scanner | |
| Detector type | Minimum of four detector rows | |
| Reconstructed slice thickness | Minimum of 5 mm | Thinner slices are preferable, especially in multiplanar reconstructions (MPR) |
| Injector | Power injector, preferably dual-chamber | Bolus tracking desirable |
| Contrast injection rate | No less than 3 mL/s of contrast, | A saline flush desirable |
| 300 mg I/mL or a higher concentration, | ||
| For a dose of 1.5 mL/kg of body weight | ||
| Mandatory dynamic phases | 1. Early arterial phase | MPR, |
| 2. Pancreatic phase | Curved MPR along the pancreatic duct, | |
| 3. Portal venous phase | Minimum intensity projections are helpful |
Table 2 Minimum technical specifications for pancreas protocol magnetic resonance imaging
| Feature | Specification | Comment |
| Scanner type | 1.5-T or greater main magnetic field | Low-field magnets not suitable |
| Coil type | Phased-array, multichannel torso coil | Unless patient-related factors preclude the use |
| Gradient type | Current-generation, high-speed gradients (providing sufficient coverage of upper abdomen) | |
| Slice thickness | 5 mm or less for dynamic series, | |
| 8 mm or less for other imaging | ||
| Breath holding and matrix | Approximately 20 s of breath hold with a minimum matrix of 128 × 256 | Breath hold instructions are very important |
| Injector | Power injector, preferably dual-chamber | Bolus tracking/MR fluoroscopy desirable |
| Contrast injection rate | 1.5-2 mL/s of gadolinium chelate | Preferably resulting in the vendor-recommended total dose |
| Minimum sequences | T1-weighted, gradient echo (3D preferable) | |
| T2-weighted, turbo spin echo (axial, coronal) | ||
| MRCP (both 2D and 3D preferable) | ||
| Post-Gd, T1-weighted gradient echo | ||
| Mandatory dynamic phases | Arterial | |
| Portal-venous phase | ||
| Equilibrium phase | ||
| Dynamic timing | Arterial: 20-40 s | |
| Portal venous: 45-65 s | ||
| Equilibrium: 3-5 min | ||
| after contrast injection |