Table 1 Wavelengths in nm (Fujinon intelligent chromo endoscopy)[13]
| Red | Green | Blue | |
| FICE channel 1 | 595 | 540 | 535 |
| FICE channel 2 | 420 | 520 | 530 |
| FICE channel 3 | 595 | 570 | 415 |
Table 2 Number of lesions (P0, P1, P2) detected by the two observers and the senior consultant, respectively
| Observer 1 | Observer 2 | Consultant (reference findings) | |
| Total | 117 | 154 | 131 |
| P0 | 20 | 27 | 15 |
| P1 | 37 | 55 | 41 |
| P2 | 60 | 72 | 75 |
Table 3 Number of small bowel findings (patients) described by the two observers when using either white light or Fujinon intelligent chromo endoscopy compared to the senior consultant (white light)
| White light(two observers) | FICE(two observers) | P value(FICE vs WL for observers 1 and 2) | White light(senior consultant) | |
| P0 (no patients) | 8 (8) | 39 (24) | < 0.01 | 15 (10) |
| P1 (no patients) | 43 (33) | 49 (32) | 0.30 | 41 (29) |
| P2 (no patients) | 67 (38) | 65 (37) | 1.00 | 75 (40) |
- Citation: Gupta T, Ibrahim M, Deviere J, Gossum AV. Evaluation of Fujinon intelligent chromo endoscopy-assisted capsule endoscopy in patients with obscure gastroenterology bleeding. World J Gastroenterol 2011; 17(41): 4590-4595