Copyright
©2005 Baishideng Publishing Group Inc.
World J Gastroenterol. Jan 28, 2005; 11(4): 521-524
Published online Jan 28, 2005. doi: 10.3748/wjg.v11.i4.521
Published online Jan 28, 2005. doi: 10.3748/wjg.v11.i4.521
Table 1 Range and precision of each sensor.
| Sensor | Range | Precision |
| Pressure | -60 to +200 mmHg | 1% FS |
| Temperature | 34 to 42 °C | ±0.2 °C |
| pH | 1 to 13 | ±0.2 |
Table 2 Weight and dimension of each capsule.
| Parameter | PT capsule | pH capsule |
| Diameter (mm) | 10 | 10 |
| Length (mm) | 21.1 | 24 |
| Weight (gram) | 2.9 | 5.2 |
Table 3 Volunteers and CIB time.
| Volunteer no. | Sex | Age (yr) | CIB time (h) |
| 1 | M | 42 | 25.78 |
| 2 | M | 52 | 33.83 |
| 3 | M | 34 | 28.22 |
| 4 | F | 50 | 11.72 |
| 5 | M | 55 | 19.07 |
| 6 | M | 54 | 33.12 |
| 7 | M | 49 | 18.77 |
Table 4 Pressure data.
| Pressure | Gauge data | ||||
| (kPa) | (mmHg) | Up | Down | Up | Down |
| -8 | -60 | -59.6 | -60 | -60.4 | -60.7 |
| -4 | -30 | -29.3 | -28.9 | -30.4 | -30.6 |
| 0 | 0 | 0 | 1.2 | -0.4 | -0.4 |
| 5.33 | 40 | 38 | 40.4 | 38 | 39.3 |
| 10.67 | 80 | 79 | 80.7 | 79 | 79.8 |
| 16 | 120 | 121.2 | 121.8 | 120.8 | 121.2 |
| 21.33 | 160 | 161.7 | 162.5 | 159.9 | 160.9 |
| 26 | 200 | 200.8 | 199.8 | 199.8 | 199.8 |
Table 5 Temperature data.
| Temperature (°C) | Gauge data |
| 34 | 34.02 |
| 36 | 35.91 |
| 38 | 38.03 |
| 40 | 40.06 |
| 42 | 42.11 |
Table 6 pH data.
| pH | Gauge data | ||||
| 1 | 2 | 3 | 4 | 5 | |
| 4 | 4.4 | 4.3 | 4.3 | 4.3 | 4.4 |
| 6.86 | 6.7 | 6.7 | 6.8 | 6.7 | 6.7 |
| 9.18 | 9.2 | 9.3 | 9.3 | 9.2 | 9.3 |
- Citation: Wang WX, Yan GZ, Sun F, Jiang PP, Zhang WQ, Zhang GF. A non-invasive method for gastrointestinal parameter monitoring. World J Gastroenterol 2005; 11(4): 521-524
- URL: https://www.wjgnet.com/1007-9327/full/v11/i4/521.htm
- DOI: https://dx.doi.org/10.3748/wjg.v11.i4.521