World J Gastroenterol. Aug 14, 2016; 22(30): 6917-6924
Published online Aug 14, 2016. doi: 10.3748/wjg.v22.i30.6917
Published online Aug 14, 2016. doi: 10.3748/wjg.v22.i30.6917
Figure 1 Resistance force to migration measurement machine.
The machine consists of a retraction robot (arrows), a silicone wall (arrowheads), and a force gauge (Model DPRS5T, Imada, Tokyo, Japan). The covered self-expandable metal stent is fixed into a round hole in the silicone wall (arrowheads) and its distal end is connected to the force gauge device. The stent is retracted from the wall at a speed of 1 mm/s. by using the retraction robot (arrows).
Figure 2 Schematic depiction of a stent.
Stent structure is expressed as outer diameter of the fully expanded stent (ODES; mm), outer diameter of the compressed stent (ODCS; mm), outer diameter of the flare (ODF; mm), height of the flare [HF = (ODF-ODES)/2; mm], length of the flare (LF; mm), and taper angle of the flare (TAF = HF/LF; degree).
Figure 3 Resistance force to migration and radial force of 6 stents.
A: The resistance force to migration (RFM) of each stent when the outer diameter of the compressed stent was 6, 8, or 10 mm. B: The radial force (RF) of each stent when the outer diameter of the compressed stent was 6, 8, or 10 mm.
Figure 4 Relationship between resistance force of migration and radial force.
A-C: The relationship between RFM and RF in all six stents when the outer diameter of the compressed stent was 6 (A), 8 (B), or 10 mm (C); D-F: The relationship between RFM and RF in just the five braided stents when the outer diameter of the compressed stent was 6 (D), 8 (E), or 10 mm (F). A simple linear regression model was used. RFM: Resistance force of migration; RF: Radial force.
Figure 5 Relationship between resistance force of migration and taper of the flare.
A-C: The relationship between RFM and TAF in all six stents when the outer diameter of the compressed stent was 6 (A), 8 (B), or 10 mm (C); D-F: The relationship between RFM and TAF in just the five braided stents when the outer diameter of the compressed stent was 6 (D), 8 (E), or 10 mm (F). A simple linear regression model was used. RFM: Resistance force of migration; TAF: Taper of the flare.
- Citation: Minaga K, Kitano M, Imai H, Harwani Y, Yamao K, Kamata K, Miyata T, Omoto S, Kadosaka K, Sakurai T, Nishida N, Kudo M. Evaluation of anti-migration properties of biliary covered self-expandable metal stents. World J Gastroenterol 2016; 22(30): 6917-6924