Copyright: ©Author(s) 2026.
World J Nephrol. Sep 25, 2026; 15(3): 119841
Published online Sep 25, 2026. doi: 10.5527/wjn.119841
Published online Sep 25, 2026. doi: 10.5527/wjn.119841
Table 1 Comparative summary of shear-wave elastography cutoff values for renal fibrosis across different devices and study populations
| Ref. | Population | Device/technique | Diagnostic target | Cutoff (kPa) | AUC | Sensitivity | Specificity |
| Prasad et al[7], 2026 | Chronic glomerulonephritis | 2D-SWE (Supersonic Mach 30, Hologic) | Any fibrosis vs none (IF/TA ≥ I) | 13.98 | 0.986 | 98% | 73% |
| Maralescu et al[45], 2023 | Chronic glomerulonephritis | 2D-SWE PLUS | Fibrosis vs no fibrosis | < 20.77 | 0.860 | 88.89% | 75% |
| Maralescu et al[45], 2023 | Chronic glomerulonephritis | 2D-SWE PLUS | > 40% fibrosis vs < 40% | < 19.75 | 0.789 | 100% | 74.29% |
| Leong et al[3], 2021 | CKD (biopsy-proven) | Philips EPIQ 7 (ElastPQ) | Moderate vs mild histologic impairment | 5.81 | 0.702 | 67.3% | 61.5% |
| Wang et al[43], 2025 | CKD | ACUSON Sequoia (Siemens) | Mild vs moderate-severe fibrosis | 6.70 | 0.824 | 91.5% | 59.3% |
- Citation: Mora-Loján JL, García-Trujillo OF, Vélez-Hurtado G, Bermúdez-Mercado E, Pacheco-Salazar F, Balderas-Juarez J, Martínez-Sánchez FD. Renal elastography in kidney disease: Current evidence, limitations, and future directions. World J Nephrol 2026; 15(3): 119841
- URL: https://www.wjgnet.com/2220-6124/full/v15/i3/119841.htm
- DOI: https://dx.doi.org/10.5527/wjn.119841