Published online Mar 28, 2015. doi: 10.5320/wjr.v5.i1.17
Peer-review started: September 28, 2014
First decision: November 19, 2014
Revised: December 10, 2014
Accepted: December 29, 2014
Article in press: January 4, 2015
Published online: March 28, 2015
Processing time: 176 Days and 5.9 Hours
The prevalence and severity of obstructive sleep apnea (OSA) is higher in specific population: children, elderly, obese and patients with pulmonary and cardiovascular diseases, compared to the general population. OSA is associated with greater morbidity and mortality in these patients. Although full-night polysomnography is still the gold standard diagnostic sleep study for OSA, it is a time consuming, expensive and technically demanding exam. Over the last few years, there is growing evidence on the use of portable monitors (PM) as an alternative for the diagnosis of OSA. These devices were developed specially for sleep evaluation at home, at a familiar environment, with easy self-application of monitoring, unattended. The use of PM is stablished for populations with high pre-test probability of OSA. However, there is a lack of studies on the use of PM in age extremes and patients with comorbidities. The purpose of this review is to present the studies that evaluated the use of PM in specific population, as well as to describe the advantages, limitations and applications of these devices in this particular group of patients. Although the total loss rate of recordings is variable in different studies, the agreement with full-night polysomnography justifies the use of PM in this population.
Core tip: This is the first review that evaluated the use of Portable Monitoring as an alternative method for diagnosis of obstructive sleep apnea in specific population. Additionally, we present the physiopathological background, technical considerations and clinical implications on the use of PM in age extremes and patients with comorbidites. We also describe the advantages, limitations and applications of these devices in children, elderly, patients with cardiovascular or respiratory diseases and obese patients.
- Citation: Treptow E, Oliveira MG, Moreira GA, Togeiro S, Nery LE, Tufik S, Bittencourt L. Update on the use of portable monitoring system for the diagnosis of sleep apnea in specific population. World J Respirol 2015; 5(1): 17-27
- URL: https://www.wjgnet.com/2218-6255/full/v5/i1/17.htm
- DOI: https://dx.doi.org/10.5320/wjr.v5.i1.17
Obstructive sleep apnea (OSA) is characterized by recurrent episodes of airflow cessation or reduction despite evidence of continuing respiratory effort. Previous studies of the general population estimated a prevalence of OSA of 2% to 5% in women and 3% to 7% in adult males[1,2]. However, due to the burden of obesity worldwide and the higher life expectancy observed in the last years, those previous studies probably underestimate the actual prevalence of OSA. In an epidemiological study conducted in Brazil[3], that included 1042 inhabitants, OSA was observed in 32.8% of the participants. The prevalence of OSA was higher among males, overweight and obese patients and increased with age. Peppard et al[4] also observed the highest prevalence among older subjects. The authors estimated the current prevalence of moderate to severe sleep disordered breathing (SDB) of 17% among 50 to 70 years old man and of 9% among 50 to 70 years old women. The prevalence and impact of OSA may be even higher in specific comorbid medical conditions as cardiovascular and respiratory disorders.
Polysomnography (PSG) is the standard diagnostic test for OSA. It is a time consuming, expensive and technically demanding exam and requires that the patient sleep in an unfamiliar environment. Even if taken into consideration the initial prevalence described in the studies, the demand for PSG is higher than the capacity of sleep centers for performing the tests[5] and many patients remain undiagnosed or untreated[6]. Therefore, in the last few years, portable monitoring (PM) systems were proposed as an alternative diagnostic method. PM is cost-effective, diminishes the time on the waiting list and allows sleep studies to be done at the patient’s home.
In 1994, the American Sleep Disorders Association, a precursor to the American Academy of Sleep Medicine (AASM) classified sleep studies into four types (Table 1)[7,8]. Full-night PSG (type 1) is the gold-standard diagnostic test for OSA and must be attended by a technician able to intervene if there is an equipment malfunction or sensor disconnection during the recording time. The most evidence on the use of PM comes from studies with type 3 monitors, which requires a minimum of 4 signals: ventilation (at least two channels of respiratory movement; or respiratory movement and airflow), oxygen saturation, and ECG or heart rate. It is an unattended exam that be applied in sleep laboratories, hospitals, clinics and at home. Recently, Collop et al[9] proposed a new method for the classification of sleep studies accordingly to the sensors used to evaluate the following parameters: sleep, cardiovascular, oximetry, position, effort and respiratory (Table 2).
Type 1 full attended polysomnography | Type 2 full unattended polysomnography | Type 3 limited channels devices | Type 4 continuous single or dual bioparameter recording | |
Number of channels | Minimum 7 | Minimum 7 | Minimum 4 | Minimum of 1 |
Parameters | EEG, EOG, chin EMG, ECG, airflow, respiratory effort, oxygen saturation | EEG, EOG, chin EMG, ECG, airflow, respiratory effort, oxygen saturation | Ventilation (at least two channels of respiratory movement or respiratory movement and airflow), heart rate or ECG, oxygen saturation | Usually oxygen saturation and/or airflow |
Body position | Documented or objectively measured | May be objectively measured | May be objectively measured | Not measured |
Leg movement | EMG or motion sensor desirable | EMG or motion sensor desirable | May be recorded | Not recorded |
Personnel | In constant attendance | Not in attendance | Not in attendance | Not in attendance |
Sleep | Cardiovascular | Oximetry | Position | Effort | Respiratory |
S1-3 EEG channels1 + EOG + chin EMG | C1-More than 1 ECG lead | O1-Oximetry (ear or finger) with recommended sampling2 | P1-Video or visual position measurement | E1-Two RIP belts | R1-Nasal pressure and thermal device |
S2-Less than 3 EEG channels with/without EOG or chin EMG | C2-Peripheral arterial tonometry | O1x-Oximetry (ear or finger) without recommended sampling | P2-Nonvisual position measurement | E2-One RIP belt | R2-Nasal pressure |
S3-Sleep surrogate: e.g., actigraphy | C3-Standard ECG measure (one lead) | O2-Oximetry with alternative site | E3-Derived effort | R3-Thermal device | |
S4-Other sleep measure | C4-Derived pulse | O3-Other oximetry | E4-Other effort measures | R4-End-tidal CO2 | |
C5-other cardiac measures | R5-Other respiratory measures |
An important step towards the consolidation of the use of PM in the diagnosis of sleep apnea was achieved in 2009[10,11]. The Centers for Medicare and Medical Services approved the use of PM as a diagnostic tool and allowed continuous positive airway pressure (CPAP) coverage when clinical evaluation and unattended PM systems were used for the diagnosis of OSA. Also, a workshop sponsored by the American Thoracic Society, the American College of Chest Physicians, AASM and the European Respiratory Society was held to determine the research priorities for incorporating ambulatory management of adults with obstructive sleep apnea into healthcare systems[12]. One of the main points discussed in this workshop was the need to include specific population in the studies with PM. There is still a lack of evidence evaluating diverse ethic groups, elderly, children and comorbid conditions such as chronic obstructive pulmonary disease (COPD), asthma, heart failure, obesity and neuromuscular disorders.
The Portable Monitoring Task Force of AASM[13] recommends PM as an alternative to PSG for the diagnosis of OSA in patients with a high pretest probability of moderate to severe OSA and does not recommend the use in patients with comorbid medical disorders or in specific population (elderly and children). However, the need for earlier diagnosis, particularly in the presence of comorbidities and the evidences of PM as a cost-effective method[13-17] led to many studies evaluating specific population in the last years. Therefore, the aim of this review is to describe the recent findings, as well as the advantages and limitations of the use of PM in patients with comorbidities and particular population that we have chosen to call “specific population”.
The prevalence of sleep apnea increases with age[3,4] due to several mechanisms: sleep pattern, hormones, respiratory system changes and a higher prevalence of comorbidities. Total sleep time (TST) sleep efficiency, REM sleep and slow-wave sleep decrease in older subjects[18]. Sleep becomes more superficial and fragmented and this may contribute to the development of SDB[18,19]. After menopause, the prevalence of sleep apnea in women is similar to men, mainly due to the effects of lower levels of sexual hormones[18,20]. Estrogen is associated with peripheral body fat deposition and lower levels predispose an increase in central fat. It is also related to temperature regulation in the body. Hot flashes are common during menopause and increase arousals during sleep. Progesterone is a respiratory stimulant and lower levels result in impairment of the ventilatory control. Age is associated with a decrease pulmonary vital capacity, increased pharyngeal wall fat deposit and impairment of the control of pharyngeal dilator muscles that may result in diminished airway patency during sleep[18,21].
Cognitive impairment, Alzheimer’s disease and dementia contribute to difficulties when considering the use of PM at home in the elderly. Therefore, it is important a careful explanation of the exam for the patient as well as for a caregiver that may help during the application of the sensors. It is recommended verbal and written instructions and a brief practical demonstration[22]. Whenever possible, leg electromyography monitoring should be included in the sleep studies due to the higher prevalence of period leg movement in this population[22,23].
Morales et al[24], enrolled 452 participants with average age of 71.4 years, who presented sleepiness complaints, and evaluated a risk score for apnea and the use of PM ResCare Auto Set (Resmed, Sydey, Australia) at home. This PM measured airflow with the use of a nasal cannula (PM type IV). The results of PM used alone, in the absence of clinical data, had limited discriminatory power for the finding of severe cases of OSA. Therefore, the combination of sleep apnea symptoms, neck circumference, age, sex and the recording of airflow and respiratory effort were more accurate in identifying severe OSA in older adults. The agreement between PSG and the unattended sleep study was better in patients presenting more than 20.9 events/h.
PM type 3 Stardust II (Philips Respironics, Inc., PA, United States) was used to evaluate 43 patients over 65 years with high pre-test probability of OSA[22]. The receiver operator curve of PM vs PSG were greater than 0.83 in all cutoffs analyzed. However, the specificity of the use of PM was above 60% only when in patients with moderate and severe OSA. There was a loss of data of 10.5% mainly due to download failures and oximetry record loss. The authors concluded that PM had a good diagnostic agreement with attended PSG and may become an affordable and comfortable alternative method for the diagnosis of OSA in this population.
Gooneratne et al[25] evaluated, on a longitudinal cohort study, 289 older patients, followed for an average of 13.8 years and with mean age of 78 ± 6.4 years. The authors observed a mortality hazard ratio of 1.5 in older adults with excessive daytime sleepiness (EDS). The combined risk factors of EDS and SDB-defined in this study as apnea hypopnea index (AHI) over 20 events/h of sleep-increased the mortality hazard ratio to 2.3. Another study[26], evaluated healthcare costs for 2 years in elderly and middle-aged patients with OSA matched with healthy controls. The results showed that the costs were 2.02 and 1.81 times as high in elderly and middle-aged patients with OSA, respectively, in comparison to their controls. Possible reasons implicated in higher costs in elderly population are greater morbidity, medication effects, depression and anxiety. Therefore, the use of PM in elderly may provide earlier diagnosis and reduce morbidity and mortality in this population.
Multiple factors have been implicated in the development of OSA in children[27]. Although adenotonsillar hypertrophy is the leading cause of OSA in this population, other anatomical factors, increased upper airway collapsibility and obesity are responsible for the burden of OSA in children, with the peak prevalence from 2 to 8 years old. Anatomical factors such as nasal obstruction, macroglossia, micrognathia, abnormalities of the cranial base and midfacial hypoplasia result in a decrease of the upper airway patency. The increase in upper airway collapsibility is evaluated by the critical closing pressure (Pcrit)-the pressure in which the upper airway collapses. This pressure is higher in children with OSA compared to those that only present primary snoring[27-29]. Finally, obesity results in increased work of breathing and fat deposition in the upper airway.
One of the challenges on the use of PM in children is the tolerability of the sensors during the exam. The most common cause of discomfort is usually the thermistor or nasal cannula[30,31]. Also, body movements may interfere with the signals acquired[32]. A limitation of the these type of sleep studies in children is the lack of evaluation of end-tidal carbon dioxide pressure (PCO2), an important measure for detecting hypoventilation[31,33] that is an important criteria in children.
Studies have demonstrated conflicting results on the use of PM in children[30-32,34]. The type of sleep study used, the setting (home vs laboratory), different definitions for SDB and the diverse age range evaluated may account for these results.
Goodwin et al[31] evaluated the use of PM in 157 children enrolled in the cohort entitled Tucson Children’s Assessment of Sleep Apnea study. The sleep evaluation consisted of a Type II monitoring performed at home and the equipment was placed and calibrated by technicians. The failure rate was 9%, mostly due to problems with pulse oximetry. High quality data was obtained in more than 90% of the exams and the majority of patients did not experience any discomfort. The authors concluded that the use of an unattended sleep study is feasible in children ages 5 to 12 years.
In 2002, a study with only 12 participants aged 3 to 6 years[32] compared the use of PM in the laboratory setting with PSG. The PM measured the following signals: thermistor, laryngeal microphone, an ECG lead, a belt for thoracic effort, two belts for abdominal effort, body position and a pulse oximeter. PM underestimated the incidence of OSA and overestimated the number of central apnea cases. Although the sensitivity increased with the increase of respiratory disturbance index (RDI), specificity was low in most RDI cutoffs. Furthermore, these two studies failed to compare data using AASM guidelines (measures of airflow with nasal cannula, and interpretation according to pediatric criteria). Based on these results, PM should be reserved for highly suspected OSA in children when treatment is urgently needed.
Poels et al[30] used a PM that measured respiratory movement, air flow (thermistor), heart rate, oxygen saturation and body position to evaluate 24 children aged 2 to 7 years and candidates for adenotonsillectomy. This study was meant to reflect daily practice, as the placement of the sensors was done by the caregivers. The definition criteria for a successful recording was those with 390 min of artifact-free signal present in 3 traces simultaneously within the TST. Only 7 out of 24 exams fulfilled this criteria. Another limitation was that from the initial 53 eligible children, only 45% of the caregivers agreed to participate in the study.
Rosen et al[35] compared type III PM with laboratory PSG in 55 children with 8 to 11 years, and found that PM had a 88% sensibility and 98% specificity to diagnose OSA, using a criteria of AHI > 5.
Hamada et al[36] evaluated the effects of adenotonsillectomy in 48 children aged 2 to 11 years, using a home type III PM. The AHI decreases from 20.6 ± 16.6 to 4.4 ± 2.1/h. However, there were no significant improvement of oxygen saturation nadir (76.7% ± 17.1% vs 80.8% ± 14.6%). Complete cure was noted in 75.6% of the children. The author recognized that oxygen saturation values were inconsistent, probably due to an adult size of the probe, not appropriate for children.
A more recent study[34], evaluated, for the first time simultaneously, the use of PM and PSG in 53 children aged 2 to 14 years with clinical suspicious of sleep apnea-hypopnea syndrome (SAHS). SAHS was defined by an obstructive apnea-hypopnea index (OAHI) of 3 or more in PSG and a RDI of 3 or more in PM. The rate of diagnostic agreement between the two methods was 84.9%. The authors concluded that PM is a valid method for diagnosis of SAHS in children.
Sárdon Prado et al[37] compared type III PM in two different setting: home and laboratory, in 44 (8.3 years) and 88 (7.3 years) children. Both groups had 96.5% valid studies. The diagnosis were similar in both groups, however the study design did not allowed sensibility/specificity calculations, and there were no comparison with a gold standard (Type I sleep study).
Amorim et al[38] performed type III PM in 33 children, aged 10.6 ± 3.4 years, with different risk factors (adenotonsillar hypertrophy, obesity, craniofacial malformation and neuromuscular disease). Authors found 94% of valid results, adequate nasal airway in 67.7%, and adequate oxygen saturation in 96.8%. There was no comparison with a gold standard.
Adenotonsillar hypertrophy, the most common cause of OSA in children, is a treatable condition and adenotonsillectomy is associated with improvements in quality of life, behavior and sleep parameters[27,39]. The intermittent hypoxia and arousals during the night are associated with EDS, poor academic performance, hyperactivity attention deficits and may result in behavioral and neurocognitive morbidity. Furthermore, OSA affects the metabolic and cardiovascular systems: Obesity, pulmonary hypertension and systemic hypertension are more prevalent in children with SDB[27,40]. Therefore, an earlier diagnosis of OSA is mandatory in order to stablish the appropriate treatment and diminish the impact of SDB during children’s development.
SDB are highly prevalent in patients with heart failure (HF). Whilst obstructive sleep apnea affects 53% of HF patients[41]; central apnea, in particularly Cheyne-Stokes respiration, is present in 15% to 50% of these population[41-43]. Over the last few years, studies have demonstrated the paramount role of nocturnal rostral fluid shift in the pathogenesis of SDB[44-46]. Fluid shift from the legs to the neck increase neck circumference and airway collapsibility resulting in obstructive sleep apnea. On the other hand, fluid shift to the lungs accentuate the hyperventilation already present in HF patients and reduces PCO2 below the threshold require to stimulate breathing, which results in central sleep apnea.
Furthermore, SDB affects patients with cerebrovascular disease. A meta-analysis of 2343 patients ischemic or hemorrhagic stroke and transient ischemic attack demonstrated a prevalence of SDB with AHI > 5 of 72%[47] which may be even higher in patients with acute stroke.
Both central and obstructive apneas may be present at the same patient with cerebro-cardiovascular disease and exhibit a variation from night to night[48,49]. The overnight shift from one type of apnea to another is related to the PCO2 and the cycle length of periodic breathing which reflect cardiac function. Therefore, it is important to differentiate the type of apnea in order to offer adequate treatment for patients.
Quintana-Gallego et al[50] conducted the first study that compared PSG and unattended PM for the diagnosis of OSA in HF. A total of 75 patients, with left ventricular ejection fraction ≤ 45%, were randomly assigned for two sleep studies: PM at home and PSG. The failure rate was 9%. The majority of patients 59/75 were correctly classified with PM, with a diagnostic accuracy of 78.6%. The diagnostic accuracy of PM varied between 78% to 84%, with high sensitivities (68% to 94%) and high specificities (88% to 97%) considering AHI thresholds of 5, 10 and 15. An important finding was that the PM was able to detect different patterns of apnea (central and obstructive) in 100% of cases.
The Sleep Events, Arrhytmias, and Respiratory analysis in congestive heart failure study[51] evaluated the use of a cardiorespiratory system that records 2 leads of ECG, pulse oximetry and respiratory impedance [Clear Path System (CPS), Nexan Inc., Alpharetta, GA] in 50 patients with HF and LVEF ≤ 35%. The study was performed in two nights: (1) simultaneous monitoring of PSG and CPS and (2) CPS at home. Failures of recording were due to technical errors with CPS, severe pacemaker interference and premature patient withdrawal. Predictive accuracy for RDI thresholds of 5, 10 and 15 ranged from 73% to 75% in the first night. When comparing CPS between nights, the same RDI event rates were observed in 69% to 80% of sessions for the different cutoffs. The majority of patients (74%) preferred home testing to PSG in the sleep laboratory.
Smith et al[52] compared simultaneous monitoring of PM and PSG to PM at home in 20 patients with HF and LVEF < 45%. Although there was a good agreement in outcomes between PM at the same night as PSG (kappa coefficient 0.63, P < 0.001), the agreement between PM and PSG measured in different nights was poor (kappa coefficient 0.27, P = 0.06). A negative result of PM of home was not able to exclude the diagnosis of SDB. The authors reported several possible explanations for the different results: diverse calculation for RDI (PSG measures RDI during TST, whilst PM measures RDI during total time in bed); night-to-night variation in respiratory events[53]; the small sample size; the amount of time spent supine and poor sleep efficiency in HF patients which may underestimate the event rate by PM.
OSA is implicated in the progression of coronary artery disease (CAD)[54-56]. Oxidative stress, endothelial disfunction, inflammation, impairment of coagulation factors and increased levels of soluble adhesion molecules are features of OSA and predispose to atherosclerosis. Therefore, an early diagnosis in this population is essential to prevent the consequents of CAD: angina, acute coronary syndrome and cerebrovascular accident. Takama et al[57] evaluated PSG and PM performed at the same night. The PM (Morpheus; Compumedics, Melbourne, Australia) measured airflow (thermistor), oxygen saturation, heart rate and respiratory movements (chest and abdominal bands). The specificity and sensitivity of diagnosing SDB using PM were 865 and 81% respectively. However, there was a tendency for higher AHI values obtained by PSG in comparison to PM.
Danzi-Soares et al[58] evaluated 79 patients with severe CAD referred for coronary artery bypass grafting. PSG and PM were performed on separate nights before surgery. The PM (Stardust II, Respironics Inc., Murrysville, Pennsylvania, United States) was performed at the hospital, as an unattended sleep study. The failure rate was 11% due to computer malfunction, problems with pulse oximetry, nasal cannula and data corrupted during download process. The agreement between PM and PSG was 73%, with PM underestimating AHI in 23% and overestimating in 4% of the cases.
The impact of OSA on primary and secondary prevention of cardiovascular disease is under evaluation on two large prospective trials that aim to provide better understanding of the use of CPAP in OSA[59,60]. The sleep apnea cardiovascular endpoints (SAVE) study[59] was designed to determine whether CPAP treatment can reduce the incidence of serious cardiovascular events in patients with established cardiovascular disease. Meanwhile, the impact of CPAP on patients with acute coronary syndrome (ACS) and Nonsleepy OSA: The ISAACC Trial[60], aims to determine if CPAP treatment reduces the incidence of cardiovascular events (cardiovascular death, nonfatal myocardial infarction, nonfatal stroke, hospitalization for heart failure, and hospitalization for unstable angina or transient ischemic attack) in patients with ACS.
The coexistence of OSA and chronic pulmonary diseases, especially COPD is called the overlap syndrome[61-65]. Several physiopathological features of COPD may predispose OSA development: smoking, which results in local upper airway inflammation, chronic use of corticosteroids that may promote truncal obesity and upper airway edema and poor sleep quality. Moreover, the worsening of hypoxemia during the night, especially during rapid eye movement (REM) sleep. As described by Douglas et al[64]: “The major cause of REM hypoxemia is hypoventilation, with additional contributions from alteration in ventilation/perfusion matching and functional residual capacity (FRC) reduction”. Intermittent hypoxemia during the night is associated with an increase in pulmonary hypertension and cardiovascular events in this population.
The basis of noninvasive SpO2 measurements relies on pulsatile blood flow and is affected by poor peripheral arterial blood flow[66]. Furthermore, COPD patients often present anemia and changes in hemoglobin structure such as carboxyhemoglobinemia that may affect oxygen saturation acquisition. During PSG, the monitoring equipment may cause sleep disturbances such as arousals and body movement artifacts which interfere on the pulse signal[67]. Although the limited number of signals measured in PM could minimize this effect, the accuracy of oximetry signal in these devices is usually lower than in PSG and the possible incorrect placement of the sensor at home environment might result in sensor disconnection during the recording time.
The only study that evaluated the use of PM in COPD patients presented negative results[68]. Oliveira et al[68] evaluated 72 patients with COPD in Global Initiative for Chronic Obstructive Lung Disease (GOLD)’s stages II and III. Patients were randomly assigned for PSG and PM at the same night at the sleep laboratory and for PM at home. Only 26 patients could be included in the final analysis due to several reasons for recording losses: oximetry and air flow measurements, problems with download and patient exacerbation. The total loss rate was 61%. The authors concluded that there is insufficient data to support the use of PM in this population.
Many studies have demonstrated a higher survival in overlap syndrome in patients treated with CPAP[69-73]. Machado et al[69] compared the surveillance of patients with overlap syndrome treated with CPAP to patients that did not accept the treatment or were not adherent. Kaplan-Meier survival curves indicated significantly higher survival in CPAP-treated patients compared to the nontreated group (P < 0.0001), with 5 year survival estimates of 71% and 26%, respectively. The benefits of CPAP are associated with improvement in respiratory mechanics and a reduction in cardiovascular morbidity and mortality.
Obesity remains the main risk factor for the development of OSA and is the most important reversible risk factor. Obesity results in increases in neck circumference and fat deposition around the upper airways, tongue, soft palate and uvula. Moreover, a combination of increased abdominal fat mass and the recumbent position during sleep determines a reduction in FRC[74,75]. In morbidly obese patients the FRC may become so marked that the FRC presents values close to residual volume. The decrease in FRC also affects the ventilatory control system[76]. Hormonal changes are also implicated in the development of OSA in obesity. Leptin is a hormone responsible for the control of appetite and has inhibitory effects in respiratory drive. Its levels are elevated in obesity, suggesting a leptin resistance.
The pattern of ventilation in obese individuals is characterized by increased work of breathing, higher respiratory rate compared to lean subjects and lower respiratory volumes. Taken together, all these factors can have an impact on the accuracy of measurements from airflow, respiratory and abdominal movements. Although the adipose tissue and poor tissue perfusion may prevent adequate penetration of light, Dumitrache-Rujinski et al[77] compared a group of obese patients to morbid obese patients (BMI > 40 kg/m2) and concluded that the use of oximetry is accurate in both groups. It is also important to use belts with the right size, in order to minimize any discomfort during the exam.
Lesser et al[78] evaluated the utility of PM for screening OSA in 29 obese (BMI ≥ 95th percentile for age/gender) healthy subjects between 9 and 18 years old with high pre-test probability of SDB. The study consisted of a simultaneous performance of PSG and PM. The Apnea Link Plus (PesMed Corporation, Poway, CA) is a type III of PM that records nasal airflow, respiratory effort, pulse rate, and pulse oximetry. The sensitivity of PM in diagnosing OSA was 100% at OAHI > 1.5; 85.7% at OAHI > 5; and 100% at OAHI > 10. The specificity of the PM increased at higher cutoffs, with the best results when an OAHI cutoff of 10 was used.
Fredheim et al[79] compared the simultaneous use of Apnea Link (ResMed Corporation, Poway, California, United States) to Embletta. A total of 105 subjects, with a mean BMI of 43.6 were included in the study. The sensitivity of the PM was 93% for an AHI cutoff 5 events/h and 94% for an AHI cutoff 15 events/h. The specificity was 71% for an AHI cutoff 5 events/h and 94% for an AHI cutoff 15 events/h. The authors concluded that the PM has a high diagnostic accuracy for diagnosing OSA in obese patients.
Oliveira et al[80] enrolled 58 obese patients candidates for bariatric surgery in a study that performed simultaneous sleep studies with PM (Stardust II, Philips Respironics, Inc. United States) and PSG and PM at home. Although the results demonstrated good agreement between methods analysed by Bland-Altman graphics, intraclass correlation coefficients and kappa coefficients, the failure rate was 45%. Inadequate flow recordings, oximetric recordings, problems with downloading data and the failure of a patient to start the device were responsible for these losses, mostly when PM was performed at home.
In patients with obesity candidates for bariatric surgery, the treatment of OSA with CPAP before and following the surgery is safe and minimizes the risk of respiratory complications such as pulmonary atelectasis and pneumonia[81,82]. After surgery, weight loss determines changes in CPAP pressure requirements and these patients should be closely followed especially in the first years after the procedure.
Full-night PSG remains the gold standard for the diagnosis of OSA. However, the use of unattended PM is an adequate alternative for earlier recognition of sleep disturbances. There is growing evidence of its importance particularly in patients with comorbidities (Table 3). Specific population, as described in this review, may present even higher prevalence of OSA than the general population; difficulties to attend to a sleep laboratory and lower tolerability to use the equipment required for PSG. Furthermore, the presence of OSA is associated with higher morbidity and mortality in these patients. Therefore, although there is no standardization of PM devices, the majority of studies have shown that the use of PM is safe, has good agreement with PSG and may be used not only for the diagnosis of OSA, but also for evaluation of autotitration at home and response after treatment.
Ref. | Year | Population | Sample size | Device | Protocol |
Morales et al[24] | 2012 | Elderly mean age 71.4 yr | 452 | Rescare AutosetTM (ResMed, Sydney, Australia) | PM at home PSG at sleep lab |
Polese et al[22] | 2013 | Elderly > 65 yr | 43 | Stardust II® (Phillips Respironics, Inc., Murrysville, PA, United States) | PM home PM + PSG at sleep lab |
Goodwin et al[31] | 2001 | Children 5 to 12 yr | 157 | Compumedics PS-2 System (Abbotsford, Victoria, Australia) | PM at home |
Zucconi et al[32] | 2003 | Children 3 to 6 yr | 12 | POLY-MESAM; MAP; Martinsried, Germany | PSG at sleep lab PM at sleep lab |
Poels et al[30] | 2003 | Children 2 to 7 yr | 24 | Embletta PDS; Flaga hf Medical Devices, Reykjavik, Iceland) | PM at home |
Rosen et al[35] | 2003 | Children 8 to 11 yr | 55 | PT-2 System SensorMedics, Yorba Linda, California, United States | PM home PSG lab |
Alvarez et al[34] | 2008 | Children 2 to 14 yr | 53 | Edentec Monitoring System (Edentrace II, Model 3711; Edentec Corporation, Nellcor Puritan Bennett, Eden Prairie, Minnesota, United States) | PSG + PM at sleep lab |
Sardón Prado et al[37] (abstract) | 2006 | Children | 132 | Not informed | PM at hospital PM at home |
Amorim et al[38] (abstract) | 2004 | Children Mean age 10.6 yr | 31 | Not informed | PM at home |
Quintana-Gallego et al[50] | 2004 | Heart failure LVEF ≤ 45% | 75 | Apnoescreen II; Erich Jaeger Gmbh and CoKg, Wuerzburg, Germany | PSG at sleep lab PM at home |
Abraham et al[51] | 2006 | Heart failure LVEF ≤ 35% | 50 | ClearPath System Nx-301; Nexan Inc, Alpharetta, GA | PSG + PM at sleep lab PM at home |
Smith et al[52] | 2007 | Heart failure LVEF ≤ 45% | 20 | Embletta; Flaga, Iceland | PSG + PM at sleep lab PM at home |
Takama et al[57] | 2010 | CAD patients | 83 | Morpheus; Compumedics, Melbourne, Australia | PSG + PM at sleep lab |
Danzi-Soares et al[58] | 2011 | CAD patients | 79 | Stardust II, Respironics Inc., Murrysville, Pennsylvania, United States | PSG + PM at sleep lab |
Oliveira et al[68] | 2012 | COPD patients | 72 | Stardust II® (Phillips Respironics, Inc., Murrysville, PA, United States) | PSG + PM at sleep lab PM at home |
Lesser et al[78] | 2012 | Obese pediatric patients (9-18 yr ) | 25 | ApneaLink Plus (PesMed Corporation, Poway, CA) | PSG + PM at sleep lab |
Fredheim et al[79] (abstract) | 2014 | Morbidly obese patients | 105 | ApneaLink | PM PM + Embletta |
Oliveira et al[80] | 2014 | Obese patients | 58 | Stardust II® (Phillips Respironics, Inc., Murrysville, PA, United States) | PSG + PM at sleep lab PM at home |
P- Reviewer: Hattori N S- Editor: Ji FF L- Editor: A E- Editor: Lu YJ
1. | Young T, Palta M, Dempsey J, Skatrud J, Weber S, Badr S. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med. 1993;328:1230-1235. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 6643] [Cited by in F6Publishing: 6251] [Article Influence: 201.6] [Reference Citation Analysis (0)] |
2. | Punjabi NM. The epidemiology of adult obstructive sleep apnea. Proc Am Thorac Soc. 2008;5:136-143. [PubMed] [Cited in This Article: ] |
3. | Tufik S, Santos-Silva R, Taddei JA, Bittencourt LR. Obstructive sleep apnea syndrome in the Sao Paulo Epidemiologic Sleep Study. Sleep Med. 2010;11:441-446. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 528] [Cited by in F6Publishing: 604] [Article Influence: 43.1] [Reference Citation Analysis (0)] |
4. | Peppard PE, Young T, Barnet JH, Palta M, Hagen EW, Hla KM. Increased prevalence of sleep-disordered breathing in adults. Am J Epidemiol. 2013;177:1006-1014. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 2961] [Cited by in F6Publishing: 2874] [Article Influence: 261.3] [Reference Citation Analysis (0)] |
5. | Flemons WW, Douglas NJ, Kuna ST, Rodenstein DO, Wheatley J. Access to diagnosis and treatment of patients with suspected sleep apnea. Am J Respir Crit Care Med. 2004;169:668-672. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 364] [Cited by in F6Publishing: 378] [Article Influence: 18.9] [Reference Citation Analysis (0)] |
6. | Young T, Evans L, Finn L, Palta M. Estimation of the clinically diagnosed proportion of sleep apnea syndrome in middle-aged men and women. Sleep. 1997;20:705-706. [PubMed] [Cited in This Article: ] |
7. | Practice parameters for the use of portable recording in the assessment of obstructive sleep apnea. Standards of Practice Committee of the American Sleep Disorders Association. Sleep. 1994;17:372-377. [PubMed] [Cited in This Article: ] |
8. | Ferber R, Millman R, Coppola M, Fleetham J, Murray CF, Iber C, McCall V, Nino-Murcia G, Pressman M, Sanders M. Portable recording in the assessment of obstructive sleep apnea. ASDA standards of practice. Sleep. 1994;17:378-392. [PubMed] [Cited in This Article: ] |
9. | Collop NA, Tracy SL, Kapur V, Mehra R, Kuhlmann D, Fleishman SA, Ojile JM. Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation. J Clin Sleep Med. 2011;7:531-548. [PubMed] [Cited in This Article: ] |
10. | Centers for Medicare & Medicaid Services. Decision Memo for Sleep Testing for Obstructive 455 Sleep Apnea (OSA) (CAG-00405N) 2009 - highwire.ashx [Internet]. Available from: http: //www.cms.gov/medicare-coverage-database/shared/handlers/highwire.ashx. [Cited in This Article: ] |
11. | Centers for Medicare & Medicaid Services. Decision Memo for Continuous Positive Airway 458 Pressure (CPAP) Therapy for Obstructive Sleep Apnea (OSA) (CAG-00093N) 2008 [Internet]. Available from: http: //www.cms.gov/medicare-coverage-database/details/nca-decision-460 memo.aspx. [Cited in This Article: ] |
12. | Kuna ST, Badr MS, Kimoff RJ, Kushida C, Lee-Chiong T, Levy P, McNicholas WT, Strollo PJ. An official ATS/AASM/ACCP/ERS workshop report: Research priorities in ambulatory management of adults with obstructive sleep apnea. Proc Am Thorac Soc. 2011;8:1-16. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 45] [Cited by in F6Publishing: 52] [Article Influence: 4.0] [Reference Citation Analysis (0)] |
13. | Collop NA, Anderson WM, Boehlecke B, Claman D, Goldberg R, Gottlieb DJ, Hudgel D, Sateia M, Schwab R. Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med. 2007;3:737-747. [PubMed] [Cited in This Article: ] |
14. | Polese JF, Santos-Silva R, Kobayashi RF, Pinto IN, Tufik S, Bittencourt LR. Portable monitoring devices in the diagnosis of obstructive sleep apnea: current status, advantages, and limitations. J Bras Pneumol. 2010;36:498-505. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 11] [Cited by in F6Publishing: 14] [Article Influence: 1.1] [Reference Citation Analysis (0)] |
15. | Oliveira MG, Garbuio S, Treptow EC, Polese JF, Tufik S, Nery LE, Bittencourt L. The use of portable monitoring for sleep apnea diagnosis in adults. Expert Rev Respir Med. 2014;8:123-132. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 4] [Cited by in F6Publishing: 5] [Article Influence: 0.5] [Reference Citation Analysis (0)] |
16. | Corral-Peñafiel J, Pepin JL, Barbe F. Ambulatory monitoring in the diagnosis and management of obstructive sleep apnoea syndrome. Eur Respir Rev. 2013;22:312-324. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 51] [Cited by in F6Publishing: 50] [Article Influence: 5.0] [Reference Citation Analysis (0)] |
17. | El Shayeb M, Topfer LA, Stafinski T, Pawluk L, Menon D. Diagnostic accuracy of level 3 portable sleep tests versus level 1 polysomnography for sleep-disordered breathing: a systematic review and meta-analysis. CMAJ. 2014;186:E25-E51. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 144] [Cited by in F6Publishing: 144] [Article Influence: 14.4] [Reference Citation Analysis (0)] |
18. | Launois SH, Pépin JL, Lévy P. Sleep apnea in the elderly: a specific entity? Sleep Med Rev. 2007;11:87-97. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 88] [Cited by in F6Publishing: 83] [Article Influence: 4.9] [Reference Citation Analysis (0)] |
19. | Ohayon MM, Carskadon MA, Guilleminault C, Vitiello MV. Meta-analysis of quantitative sleep parameters from childhood to old age in healthy individuals: developing normative sleep values across the human lifespan. Sleep. 2004;27:1255-1273. [PubMed] [Cited in This Article: ] |
20. | Eichling P, Sahni J. JCSM - Menopause Related Sleep Disorders [Internet]. Available from: http: //www.aasmnet.org/jcsm/ViewAbstract.aspx?pid=26347. [Cited in This Article: ] |
21. | Cooke JR, Ancoli-Israel S. Normal and abnormal sleep in the elderly. Handb Clin Neurol. 2011;98:653-665. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 105] [Cited by in F6Publishing: 100] [Article Influence: 7.7] [Reference Citation Analysis (0)] |
22. | Polese JF, Santos-Silva R, de Oliveira Ferrari PM, Sartori DE, Tufik S, Bittencourt L. Is portable monitoring for diagnosing obstructive sleep apnea syndrome suitable in elderly population? Sleep Breath. 2013;17:679-686. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 20] [Cited by in F6Publishing: 25] [Article Influence: 2.1] [Reference Citation Analysis (0)] |
23. | Roepke SK, Ancoli-Israel S. Sleep disorders in the elderly. Indian J Med Res. 2010;131:302-310. [PubMed] [Cited in This Article: ] |
24. | Morales CR, Hurley S, Wick LC, Staley B, Pack FM, Gooneratne NS, Maislin G, Pack A, Gurubhagavatula I. In-home, self-assembled sleep studies are useful in diagnosing sleep apnea in the elderly. Sleep. 2012;35:1491-1501. [PubMed] [Cited in This Article: ] |
25. | Gooneratne NS, Richards KC, Joffe M, Lam RW, Pack F, Staley B, Dinges DF, Pack AI. Sleep disordered breathing with excessive daytime sleepiness is a risk factor for mortality in older adults. Sleep. 2011;34:435-442. [PubMed] [Cited in This Article: ] |
26. | Tarasiuk A, Greenberg-Dotan S, Simon-Tuval T, Oksenberg A, Reuveni H. The effect of obstructive sleep apnea on morbidity and health care utilization of middle-aged and older adults. J Am Geriatr Soc. 2008;56:247-254. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 69] [Cited by in F6Publishing: 69] [Article Influence: 4.3] [Reference Citation Analysis (0)] |
27. | Tan HL, Gozal D, Kheirandish-Gozal L. Obstructive sleep apnea in children: a critical update. Nat Sci Sleep. 2013;5:109-123. [PubMed] [Cited in This Article: ] |
28. | Marcus CL, McColley SA, Carroll JL, Loughlin GM, Smith PL, Schwartz AR. Upper airway collapsibility in children with obstructive sleep apnea syndrome. J Appl Physiol (1985). 1994;77:918-924. [PubMed] [Cited in This Article: ] |
29. | Gozal D, Burnside MM. Increased upper airway collapsibility in children with obstructive sleep apnea during wakefulness. Am J Respir Crit Care Med. 2004;169:163-167. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 79] [Cited by in F6Publishing: 83] [Article Influence: 4.0] [Reference Citation Analysis (0)] |
30. | Poels PJ, Schilder AG, van den Berg S, Hoes AW, Joosten KF. Evaluation of a new device for home cardiorespiratory recording in children. Arch Otolaryngol Head Neck Surg. 2003;129:1281-1284. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 45] [Cited by in F6Publishing: 51] [Article Influence: 2.6] [Reference Citation Analysis (0)] |
31. | Goodwin JL, Enright PL, Kaemingk KL, Rosen GM, Morgan WJ, Fregosi RF, Quan SF. Feasibility of using unattended polysomnography in children for research--report of the Tucson Children’s Assessment of Sleep Apnea study (TuCASA). Sleep. 2001;24:937-944. [PubMed] [Cited in This Article: ] |
32. | Zucconi M, Calori G, Castronovo V, Ferini-Strambi L. Respiratory monitoring by means of an unattended device in children with suspected uncomplicated obstructive sleep apnea: a validation study. Chest. 2003;124:602-607. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 42] [Cited by in F6Publishing: 45] [Article Influence: 2.1] [Reference Citation Analysis (0)] |
33. | Morielli A, Desjardins D, Brouillette RT. Transcutaneous and end-tidal carbon dioxide pressures should be measured during pediatric polysomnography. Am Rev Respir Dis. 1993;148:1599-1604. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 47] [Cited by in F6Publishing: 48] [Article Influence: 1.5] [Reference Citation Analysis (0)] |
34. | Alonso Alvarez ML, Terán Santos J, Cordero Guevara JA, Navazo Egüia AI, Ordax Carbajo E, Masa Jiménez JF, Pelayo R. [Reliability of respiratory polygraphy for the diagnosis of sleep apnea-hypopnea syndrome in children]. Arch Bronconeumol. 2008;44:318-323. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 27] [Cited by in F6Publishing: 28] [Article Influence: 1.8] [Reference Citation Analysis (0)] |
35. | Rosen CL, Larkin EK, Kirchner HL, Emancipator JL, Bivins SF, Surovec SA, Martin RJ, Redline S. Prevalence and risk factors for sleep-disordered breathing in 8- to 11-year-old children: association with race and prematurity. J Pediatr. 2003;142:383-389. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 463] [Cited by in F6Publishing: 462] [Article Influence: 22.0] [Reference Citation Analysis (0)] |
36. | Hamada M, Iida M. Home monitoring using portable polygraphy for perioperative assessment of pediatric obstructive sleep apnea syndrome. Tokai J Exp Clin Med. 2012;37:66-70. [PubMed] [Cited in This Article: ] |
37. | Sardón Prado O, González Pérez-Yarza E, Aldasoro Ruiz A, Estévez Domingo M, Mintegui Aranburu J, Korta Murua J, Emparanza Knörr JI. [Diagnostic utility of nocturnal in-home respiratory polygraphy]. An Pediatr (Barc). 2006;65:310-315. [PubMed] [Cited in This Article: ] |
38. | Amorim A, Sucena M, Winck JC, Almeida J. [Home cardiorespiratory sleep study in children. Will it be feasible?]. Rev Port Pneumol. 2004;10:463-474. [PubMed] [Cited in This Article: ] |
39. | Marcus CL, Moore RH, Rosen CL, Giordani B, Garetz SL, Taylor HG, Mitchell RB, Amin R, Katz ES, Arens R. A randomized trial of adenotonsillectomy for childhood sleep apnea. N Engl J Med. 2013;368:2366-2376. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 852] [Cited by in F6Publishing: 891] [Article Influence: 81.0] [Reference Citation Analysis (0)] |
40. | Alexander NS, Schroeder JW. Pediatric obstructive sleep apnea syndrome. Pediatr Clin North Am. 2013;60:827-840. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 28] [Cited by in F6Publishing: 29] [Article Influence: 2.6] [Reference Citation Analysis (0)] |
41. | Ferrier K, Campbell A, Yee B, Richards M, O’Meeghan T, Weatherall M, Neill A. Sleep-disordered breathing occurs frequently in stable outpatients with congestive heart failure. Chest. 2005;128:2116-2122. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 146] [Cited by in F6Publishing: 153] [Article Influence: 8.1] [Reference Citation Analysis (0)] |
42. | Sin DD, Fitzgerald F, Parker JD, Newton G, Floras JS, Bradley TD. Risk factors for central and obstructive sleep apnea in 450 men and women with congestive heart failure. Am J Respir Crit Care Med. 1999;160:1101-1106. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 769] [Cited by in F6Publishing: 728] [Article Influence: 29.1] [Reference Citation Analysis (0)] |
43. | Silva RS, Figueiredo AC, Mady C, Lorenzi-Filho G. Breathing disorders in congestive heart failure: gender, etiology and mortality. Braz J Med Biol Res. 2008;41:215-222. [PubMed] [Cited in This Article: ] |
44. | White LH, Bradley TD. Role of nocturnal rostral fluid shift in the pathogenesis of obstructive and central sleep apnoea. J Physiol. 2013;591:1179-1193. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 225] [Cited by in F6Publishing: 202] [Article Influence: 18.4] [Reference Citation Analysis (0)] |
45. | Shiota S, Ryan CM, Chiu KL, Ruttanaumpawan P, Haight J, Arzt M, Floras JS, Chan C, Bradley TD. Alterations in upper airway cross-sectional area in response to lower body positive pressure in healthy subjects. Thorax. 2007;62:868-872. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 128] [Cited by in F6Publishing: 140] [Article Influence: 8.2] [Reference Citation Analysis (0)] |
46. | Yumino D, Redolfi S, Ruttanaumpawan P, Su MC, Smith S, Newton GE, Mak S, Bradley TD. Nocturnal rostral fluid shift: a unifying concept for the pathogenesis of obstructive and central sleep apnea in men with heart failure. Circulation. 2010;121:1598-1605. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 318] [Cited by in F6Publishing: 323] [Article Influence: 23.1] [Reference Citation Analysis (0)] |
47. | Johnson KG, Johnson DC. Frequency of sleep apnea in stroke and TIA patients: a meta-analysis. J Clin Sleep Med. 2010;6:131-137. [PubMed] [Cited in This Article: ] |
48. | Tkacova R, Niroumand M, Lorenzi-Filho G, Bradley TD. Overnight shift from obstructive to central apneas in patients with heart failure: role of PCO2 and circulatory delay. Circulation. 2001;103:238-243. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 180] [Cited by in F6Publishing: 177] [Article Influence: 7.7] [Reference Citation Analysis (0)] |
49. | Tkacova R, Wang H, Bradley TD. Night-to-night alterations in sleep apnea type in patients with heart failure. J Sleep Res. 2006;15:321-328. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 47] [Cited by in F6Publishing: 50] [Article Influence: 2.9] [Reference Citation Analysis (0)] |
50. | Quintana-Gallego E, Villa-Gil M, Carmona-Bernal C, Botebol-Benhamou G, Martínez-Martínez A, Sánchez-Armengol A, Polo-Padillo J, Capote F. Home respiratory polygraphy for diagnosis of sleep-disordered breathing in heart failure. Eur Respir J. 2004;24:443-448. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 57] [Cited by in F6Publishing: 66] [Article Influence: 3.3] [Reference Citation Analysis (0)] |
51. | Abraham WT, Trupp RJ, Phillilps B, Bourge RC, Bailey B, Harding SM, Schofield P, Pilsworth S, Shneerson JM, Di Salvo T. Validation and clinical utility of a simple in-home testing tool for sleep-disordered breathing and arrhythmias in heart failure: results of the Sleep Events, Arrhythmias, and Respiratory Analysis in Congestive Heart Failure (SEARCH) study. Congest Heart Fail. 2006;12:241-247; quiz 248-249. [PubMed] [Cited in This Article: ] |
52. | Smith LA, Chong DW, Vennelle M, Denvir MA, Newby DE, Douglas NJ. Diagnosis of sleep-disordered breathing in patients with chronic heart failure: evaluation of a portable limited sleep study system. J Sleep Res. 2007;16:428-435. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 16] [Cited by in F6Publishing: 23] [Article Influence: 1.4] [Reference Citation Analysis (0)] |
53. | Bittencourt LR, Suchecki D, Tufik S, Peres C, Togeiro SM, Bagnato MC, Nery LE. The variability of the apnoea-hypopnoea index. J Sleep Res. 2001;10:245-251. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 96] [Cited by in F6Publishing: 110] [Article Influence: 4.8] [Reference Citation Analysis (0)] |
54. | Prinz C, Bitter T, Piper C, Horstkotte D, Faber L, Oldenburg O. Sleep apnea is common in patients with coronary artery disease. Wien Med Wochenschr. 2010;160:349-355. [PubMed] [Cited in This Article: ] |
55. | De Torres-Alba F, Gemma D, Armada-Romero E, Rey-Blas JR, López-de-Sá E, López-Sendon JL. Obstructive sleep apnea and coronary artery disease: from pathophysiology to clinical implications. Pulm Med. 2013;2013:768064. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 24] [Cited by in F6Publishing: 29] [Article Influence: 2.6] [Reference Citation Analysis (0)] |
56. | Drager LF, Bortolotto LA, Lorenzi MC, Figueiredo AC, Krieger EM, Lorenzi-Filho G. Early signs of atherosclerosis in obstructive sleep apnea. Am J Respir Crit Care Med. 2005;172:613-618. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 368] [Cited by in F6Publishing: 377] [Article Influence: 19.8] [Reference Citation Analysis (0)] |
57. | Takama N, Kurabayashi M. Effectiveness of a portable device and the need for treatment of mild-to-moderate obstructive sleep-disordered breathing in patients with cardiovascular disease. J Cardiol. 2010;56:73-78. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 8] [Cited by in F6Publishing: 12] [Article Influence: 0.9] [Reference Citation Analysis (0)] |
58. | Danzi-Soares NJ, Genta PR, Nerbass FB, Pedrosa RP, Soares FS, César LA, Drager LF, Skomro R, Lorenzi-Filho G. Obstructive sleep apnea is common among patients referred for coronary artery bypass grafting and can be diagnosed by portable monitoring. Coron Artery Dis. 2012;23:31-38. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 48] [Cited by in F6Publishing: 63] [Article Influence: 5.3] [Reference Citation Analysis (0)] |
59. | McEvoy RD, Anderson CS, Antic NA, Chen B, He Q, Heeley E, Huang S, Huang Y, Wang J, Zhong N. The sleep apnea cardiovascular endpoints (SAVE) trial: Rationale and start-up phase. J Thorac Dis. 2010;2:138-143. [PubMed] [Cited in This Article: ] |
60. | Esquinas C, Sánchez-de-la Torre M, Aldomá A, Florés M, Martínez M, Barceló A, Barbé F. Rationale and methodology of the impact of continuous positive airway pressure on patients with ACS and nonsleepy OSA: the ISAACC Trial. Clin Cardiol. 2013;36:495-501. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 46] [Cited by in F6Publishing: 58] [Article Influence: 5.3] [Reference Citation Analysis (0)] |
61. | Mieczkowski B, Ezzie ME. Update on obstructive sleep apnea and its relation to COPD. Int J Chron Obstruct Pulmon Dis. 2014;9:349-362. [PubMed] [Cited in This Article: ] |
62. | Flenley DC. Sleep in chronic obstructive lung disease. Clin Chest Med. 1985;6:651-661. [PubMed] [Cited in This Article: ] |
63. | Weitzenblum E, Chaouat A. Sleep and chronic obstructive pulmonary disease. Sleep Med Rev. 2004;8:281-294. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 81] [Cited by in F6Publishing: 85] [Article Influence: 4.3] [Reference Citation Analysis (0)] |
64. | Douglas NJ. Sleep in patients with chronic obstructive pulmonary disease. Clin Chest Med. 1998;19:115-125. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 32] [Cited by in F6Publishing: 33] [Article Influence: 1.3] [Reference Citation Analysis (0)] |
65. | Lee R, McNicholas WT. Obstructive sleep apnea in chronic obstructive pulmonary disease patients. Curr Opin Pulm Med. 2011;17:79-83. [PubMed] [Cited in This Article: ] |
66. | Netzer N, Eliasson AH, Netzer C, Kristo DA. Overnight pulse oximetry for sleep-disordered breathing in adults: a review. Chest. 2001;120:625-633. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 222] [Cited by in F6Publishing: 239] [Article Influence: 10.4] [Reference Citation Analysis (0)] |
67. | Brijker F, van den Elshout FJ, Heijdra YF, Folgering HT. Underestimation of nocturnal hypoxemia due to monitoring conditions in patients with COPD. Chest. 2001;119:1820-1826. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 9] [Cited by in F6Publishing: 10] [Article Influence: 0.4] [Reference Citation Analysis (0)] |
68. | Oliveira MG, Nery LE, Santos-Silva R, Sartori DE, Alonso FF, Togeiro SM, Tufik S, Bittencourt L. Is portable monitoring accurate in the diagnosis of obstructive sleep apnea syndrome in chronic pulmonary obstructive disease? Sleep Med. 2012;13:1033-1038. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 19] [Cited by in F6Publishing: 32] [Article Influence: 2.7] [Reference Citation Analysis (0)] |
69. | Machado MC, Vollmer WM, Togeiro SM, Bilderback AL, Oliveira MV, Leitão FS, Queiroga F, Lorenzi-Filho G, Krishnan JA. CPAP and survival in moderate-to-severe obstructive sleep apnoea syndrome and hypoxaemic COPD. Eur Respir J. 2010;35:132-137. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 110] [Cited by in F6Publishing: 117] [Article Influence: 7.8] [Reference Citation Analysis (0)] |
70. | Stanchina ML, Welicky LM, Donat W, Lee D, Corrao W, Malhotra A. Impact of CPAP use and age on mortality in patients with combined COPD and obstructive sleep apnea: the overlap syndrome. J Clin Sleep Med. 2013;9:767-772. [PubMed] [Cited in This Article: ] |
71. | Marin JM, Soriano JB, Carrizo SJ, Boldova A, Celli BR. Outcomes in patients with chronic obstructive pulmonary disease and obstructive sleep apnea: the overlap syndrome. Am J Respir Crit Care Med. 2010;182:325-331. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 401] [Cited by in F6Publishing: 461] [Article Influence: 32.9] [Reference Citation Analysis (0)] |
72. | Esquinas AM, Marrero FL, Esteva JN. Effect of CPAP on long-term mortality in overlap syndrome: is hypercapnic the best appropriate determinant? Lung. 2014;192:631-632. [PubMed] [DOI] [Cited in This Article: ] [Cited by in F6Publishing: 1] [Reference Citation Analysis (0)] |
73. | Jaoude P, Kufel T, El-Solh AA. Survival benefit of CPAP favors hypercapnic patients with the overlap syndrome. Lung. 2014;192:251-258. [PubMed] [Cited in This Article: ] |
74. | Drager LF, Togeiro SM, Polotsky VY, Lorenzi-Filho G. Obstructive sleep apnea: a cardiometabolic risk in obesity and the metabolic syndrome. J Am Coll Cardiol. 2013;62:569-576. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 436] [Cited by in F6Publishing: 515] [Article Influence: 46.8] [Reference Citation Analysis (0)] |
75. | Ong CW, O’Driscoll DM, Truby H, Naughton MT, Hamilton GS. The reciprocal interaction between obesity and obstructive sleep apnoea. Sleep Med Rev. 2013;17:123-131. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 82] [Cited by in F6Publishing: 90] [Article Influence: 7.5] [Reference Citation Analysis (0)] |
76. | Khoo MC. Determinants of ventilatory instability and variability. Respir Physiol. 2000;122:167-182. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 215] [Cited by in F6Publishing: 200] [Article Influence: 8.3] [Reference Citation Analysis (0)] |
77. | Dumitrache-Rujinski S, Calcaianu G, Zaharia D, Toma CL, Bogdan M. The role of overnight pulse-oximetry in recognition of obstructive sleep apnea syndrome in morbidly obese and non obese patients. Maedica (Buchar). 2013;8:237-242. [PubMed] [Cited in This Article: ] |
78. | Lesser DJ, Haddad GG, Bush RA, Pian MS. The utility of a portable recording device for screening of obstructive sleep apnea in obese adolescents. J Clin Sleep Med. 2012;8:271-277. [PubMed] [Cited in This Article: ] |
79. | Fredheim JM, Røislien J, Hjelmesæth J. Validation of a portable monitor for the diagnosis of obstructive sleep apnea in morbidly obese patients. J Clin Sleep Med. 2014;10:751-757, 757A. [PubMed] [Cited in This Article: ] |
80. | Oliveira MG, Treptow EC, Fukuda C, Nery LE, Valadares RM, Tufik S, Bittencourt L, Togeiro SM. Diagnostic Accuracy of Home-Based Monitoring System in Morbidly Obese Patients with High Risk for Sleep Apnea. Obes Surg. 2014;Oct 25; Epub ahead of print. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 9] [Cited by in F6Publishing: 11] [Article Influence: 1.2] [Reference Citation Analysis (0)] |
81. | Ramirez A, Lalor PF, Szomstein S, Rosenthal RJ. Continuous positive airway pressure in immediate postoperative period after laparoscopic Roux-en-Y gastric bypass: is it safe? Surg Obes Relat Dis. 2009;5:544-546. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 48] [Cited by in F6Publishing: 50] [Article Influence: 3.3] [Reference Citation Analysis (0)] |
82. | Huerta S, DeShields S, Shpiner R, Li Z, Liu C, Sawicki M, Arteaga J, Livingston EH. Safety and efficacy of postoperative continuous positive airway pressure to prevent pulmonary complications after Roux-en-Y gastric bypass. J Gastrointest Surg. 2002;6:354-358. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 116] [Cited by in F6Publishing: 80] [Article Influence: 3.6] [Reference Citation Analysis (0)] |