        World J Orthop. Oct 18, 2015; 6(9): 705-711
Published online Oct 18, 2015. doi: 10.5312/wjo.v6.i9.705
Published online Oct 18, 2015. doi: 10.5312/wjo.v6.i9.705
            Table 1 Screw types used
        
    | Plate hole number | #1 | #2 | #3 | #4 | #5 | #6 | #7 | #8 | 
| Screw types used in composite bones | ||||||||
| L/NL/- | L | L | L | - | L | NL | L | NL | 
| Screw types used in cadaveric bones | ||||||||
| L/NL/- | L | L | L | - | L | NL | L | NL | 
            Table 2 Load to failure criterion of 2 mm gapping, displacement, stiffness and bone density per specimen
        
    | Composite bones | Human cadaver bones | |||||||||
| 1 | 2 | 3 | 4 | 5 | 1 | 2 | 3 | 4 | 5 | |
| Load to failure criteria (nm) | 11.91 | 9.40 | 10.97 | 10.67 | 9.64 | 15.10 | 12.59 | 13.05 | 13.02 | 15.38 | 
| Displacement (mm) | 1.76 | 2.00 | 1.85 | 2.00 | 1.70 | 2.00 | 2.00 | 1.71 | 2.00 | 2.00 | 
| Stiffness (nm/mm) | 6.80 | 4.69 | 5.95 | 5.33 | 5.69 | 7.55 | 6.31 | 7.62 | 6.52 | 7.72 | 
| Bone density (g/cm2) | - | - | - | - | - | 0.723 | 0.456 | 0.658 | 0.665 | 0.883 | 
| Failure mode | Distal fracture | Fracture gap | Distal fracture | Fracture gap | Distal fracture | Fracture gap | Fracture gap | Triceps failure | Fracture gap | Fracture gap | 
            Table 3 Comparison of the results from this study (composite synthetic fourth generation bones and cadaveric bones) with previous literature studies
        
    | Load to failure | Stiffness | |
| Gordon | 30-34.5 nm | 7-13 nm/mm | 
| Buijze | 4.5-27 nm | 8.8-13.3 nm/mm | 
| This study (cadaveric) | 12.59-15.38 nm | 7.55 nm/mm | 
| This study (composite) | 9.40-11.91 nm | 5.69 nm/mm | 
- Citation: Hamilton Jr DA, Reilly D, Wipf F, Kamineni S. Comminuted olecranon fracture fixation with pre-contoured plate: Comparison of composite and cadaver bones. World J Orthop 2015; 6(9): 705-711
 - URL: https://www.wjgnet.com/2218-5836/full/v6/i9/705.htm
 - DOI: https://dx.doi.org/10.5312/wjo.v6.i9.705
 