Minireviews
Copyright ©The Author(s) 2017.
World J Cardiol. Feb 26, 2017; 9(2): 139-146
Published online Feb 26, 2017. doi: 10.4330/wjc.v9.i2.139
Table 1 Different endocardial left atrial appendage occlusion devices[31]
Device nameCompanyDesign
PLAATOAppriva Medical Inc.Single-lobe occluder; nitinol cage; ePTFE membrane hooks
WATCHMANBoston ScientificSingle-lobe occluder;nitinol frame; PET membrane; hooks
ACPSt. Jude MedicalLobe and disc (polyester mesh); nitinol mesh structure; stabilizing wires
AmuletSt. Jude MedicalLobe and disk (polyester mesh in both); nitinol mesh structure; stabilizing wires
WaveCrestCoherex MedicalSingle-lobe occluder; nitinol frame, polyurethane foam and ePTFE membrane; retractable anchors
Occlutech LAAOcclutechSingle-lobe occluder; nitinol wire mesh; stabilizing loops; nanomaterial covering
Sideris PatchCustom Medical DevicesFrameless detachable latex balloon covered with polyurethane
LambreLifetechLobe and disk; nitinol; PET membrane; distal barbs anchors
PfmPFM MedicalDual disk (distal anchor, variable middle connector, proximal disk); nitinol frame
UltraseptCardiaLobe and disk; nitinol frame; Ivalon covering; distal anchors
Table 2 Summary of data for left atrial appendage occlusion
PROTECT-AF[46]CAP[43]ASAP[42]EWOLUTION[49]ACP[50]
Patients (n)46346015010211047
Follow-up4 yr16 mo14 mo30 d13 mo
CHADS-score2.22.42.82.8n.a.
CHA2DS2-Vasc scoren.a.n.a.4.44.54.5
Procedural success88.00%95.00%94.70%98.50%97.30%
Procedural stroke1.30%00.70%n.a.0.90%
Pericardial effusion4.80%2.20%1.30%0.50%1.20%
Device embolization0.60%01.30%0.20%n.a.
Major bleeding3.50%n.a.n.a.1.60%1.50%
Long-term stroke2.30%1.50%0.70%0.30%2.30%
Table 3 Different aspects of left atrial appendage occlusio
ProContra
Non-inferiority to oral anticoagulationEvaluation of other atherothrombotic sources
Alternative in patients with contraindication for anticoagulationUnknown hemodynamic impact
Cost-effectivePostprocedural medical treatment not well defined
Reduced cumulative bleeding events during follow-upNo comparison between different devices
Good results in real-world registriesUndefined impact of residual leaks