Table 1 Class 1 indications for pacing in congenital heart disease
Category
Indications
Sinus node dysfunctionSymptomatic
Bradycardia induced ventricular arrhythmias
Refractory tachycardia-bradycardia syndrome
Atrio-ventricular blockSymptomatic
Bradycardia induced ventricular arrhythmias
Post-operative 3rd degree block persisting for more than 7 days
Asymptomatic high degree atrioventricular node block with: (1) Low cardiac output; (2) Prolonged QT interval; (3) Ventricular dysfunction; (4) Complex ectopy; and (5) Wide escape rhythm
Table 2 A summary of the quoted technical specifications of current leadless pacing systems as of September 2026 based on regulatory submissions and supplier technical submissions

Micra VR
Micra AV
Aveir AR
Aveir VR
Empower
EBR wise
Width (mm)6.76.76.56.562.7
Length (mm)25.925.9323831.99.1
FixationPassive tinePassive tineActive helixActive helixPassive tineActive spike
SensorAccelerometerAccelerometerTemperatureTemperature AccelerometerN/A
Delivery catheter French size232325252112
Maximum mode VVIRVDDAAIRVVIRVVIRBIVP
Co-implantNoNoCan be DRCan be DRWith S-ICDMandatory
Designed to retrieveNoNoYesYesYesNo
Battery projected maximum16 years16 years10 years20 years10 years4.5 years
FDA ApprovedYesYesYesYesNoYes
Commercial useYesYesYesYesNoYes
IndicationBradycardia pacingBradycardia pacingBradycardia pacingBradycardia pacingBradycardia pacing and anti-tachycardia ventricular pacingResynchronisation therapy
Paediatric useYesYesYesYesNoNo
Table 3 A summary of the evidence for conduction system pacing and leadless pacemakers in the complex congenital heart disease cohorts-transposition of the great arteries-atrial switch, Fontan circulation and Univentricular anatomy
Technique
TGA-ASW
Fontan
Univentricular
CSPCase reports-multipleCase report-singleNone
LPMsCase reports-multipleCase reports-multiple
Case series
Case reports-multiple
Case series
Table 4 A proposed summary of the potential advantages and disadvantages of conduction system pacing and leadless pacemakers in the specific complex congenital heart disease groups reviewed in this paper based on expert opinion
Lesion
CSP-advantages
CSP-disadvantages
LPM-advantages
LPM- disadvantages
TGA-ASWPhysiological pacing/treatment of systemic ventricular dysfunctionNo long-term data/Leads through the baffle/sheath reshaping and delivery difficulty Lower infection risk/no baffle obstructionDifficulty negotiating baffles/difficult to place passive LPM in a smooth walled sub pulmonary left ventricle
APFAs aboveTechnical difficulties delivering to conduction system areaSuperior catheter support in a dilated atrium to deliver atrial LPMsLarge bore vascular access/potentially higher embolisation risk for atrial devices/transeptal delivery for ventricular devices/risk of embolisation in a systemic ventricle
LTAs aboveThrombotic risk with systemic ventricular lead/fenestration creation difficulty if neededAs above/could be lower thrombotic risk than a systemic endocardial leadAs above
ECCAs aboveAs above- same risk with atrial lead if in pulmonary venous atriumCould be lower thrombotic risk than a systemic endocardial leadAs above
UniventricularAs aboveThrombotic risk increased with endocardial leadsAs aboveLarge bore vascular access/embolisation risk to the systemic circulation
Table 5 A suggested patient, indication and device choice summary algorithm based on available evidence and expert opinion
Indication
TGA-ASW
APF
LT
ECC
Univentricular
SND onlyEndocardial lead system or atrial LPMEndocardial lead system or atrial LPMEndocardial lead system or atrial LPMPreferentially epicardial system first choice and transbaffle endocardial lead system or LPM if surgical turndownEndocardial lead system or atrial LPM
SND and AVBEndocardial lead system or dual chamber LPMEndocardial lead system- epicardial as back up if surgically suitable and endocardial access failed. Ventricle LPM only if surgical turndownEndocardial lead system- epicardial as back up if surgically suitable and endocardial access failed.
Ventricle LPM only if surgical turndown
Preferentially epicardial system first choice and transbaffle endocardial lead system Endocardial lead system
AVB-low burden of pacing predictedVentricular LPM- active fix device to the sub pulmonary left ventricleAs aboveAs abovePreferentially epicardial system first choice and transbaffle LPM if surgical turndownVentricular LPM
AVB- high burden of pacing predictedEndocardial lead system- CSP reasonable to offer upfront or VDD or dual chamber LPMAs aboveAs abovePreferentially epicardial system first choice and transbaffle endocardial lead systemEndocardial lead system
Any of the above and baffle stenosisAtrial, ventricular or dual chamber LPMPreferentially epicardial system first choice and LPM if surgical turndownPreferentially epicardial system first choice and transbaffle LPM if surgical turndown
Heart failure with wide QRSCSP with endocardial LBBAPEndocardial lead system- epicardial as back up if surgically suitable and endocardial not feasibleDependent on surgical risk- transbaffle endocardial lead system if high risk vs epicardial if lower riskDependent on surgical risk- transbaffle endocardial lead system if high risk vs epicardial if lower riskTraditional resynchronisation pacing or CSP with endocardial LBBAP if no coronary sinus