Copyright
©The Author(s) 2022.
World J Cardiol. Mar 26, 2022; 14(3): 108-138
Published online Mar 26, 2022. doi: 10.4330/wjc.v14.i3.108
Published online Mar 26, 2022. doi: 10.4330/wjc.v14.i3.108
Table 1 DEFINITION criteria for a complex coronary bifurcation lesion[40]
| Major criteria | Complex lesion definition | Minor criteria |
| For left main bifurcation (Major 1 ) | Major 1 or Major 2 + any 2 minor criteria | > Mild calcification |
| SB lesion length ≥10 mm and SB diameter stenosis ≥ 70% | Multiple lesions | |
| Bifurcation angle < 45° or > 70° | ||
| For non-left main bifurcation (Major 2) | MV-RVD < 2.5 mm | |
| SB lesion length ≥ 10 mm and SB diameter stenosis ≥ 90% | MV lesion length ≥ 25 mm | |
| Thrombus-containing lesions |
Table 2 Maximum-expansion capacity of stents
| Manufacturer | Stent type | Stent size (mm) | Balloon1 (bench test) (mm) | Maximum expansion (bench test) (mm)[72] | Maximum expansion (mm) (manufacturer recommendation) |
| Biosensors | Biomatrix A | 3 | 5 | 4.1 | - |
| 4 | 6 | 5.9 | 5.8 | ||
| Medtronic | Resolute Onyx | 2.5 | 4 | 3.3 | - |
| 3 | 5 | 4.4 | - | ||
| 4 | 6 | 5.6 | - | ||
| 5 | 6 | 6 | 6 | ||
| Biotronik | Orsiro | 3 | 5 | 4 | - |
| 4 | 6 | 5.3 | 4.4 | ||
| Abbott | Xience | 3 | 5 | 4.1 | 3.75 |
| 4 | 6 | 5.6 | 5.5 | ||
| Terumo | Ultimaster | 3 | 5 | 4.3 | - |
| 4 | 6 | 5.8 | 5.5 | ||
| Boston Scientific | Synergy | 2.75 | 5 | 3.6 | 3.5 |
| 3.5 | 5 | 4.2 | 4.25 | ||
| 4 | 6 | 5.7 | 5.75 |
- Citation: Kırat T. Fundamentals of percutaneous coronary bifurcation interventions. World J Cardiol 2022; 14(3): 108-138
- URL: https://www.wjgnet.com/1949-8462/full/v14/i3/108.htm
- DOI: https://dx.doi.org/10.4330/wjc.v14.i3.108