Published online Nov 26, 2026. doi: 10.4330/wjc.125622
Revised: September 6, 2026
Accepted: November 2, 2026
Published online: November 26, 2026
Processing time: 129 Days and 21.4 Hours
Congenital heart disease (CHD) is a growing health care problem, as more pe
Core Tip: Newer technologies and techniques exist, such as leadless pacing and conduction system pacing, which can be used even in the most complex congenital heart disease cohorts to address needs where traditional techniques such as epicardial pacing are not possible. Long-term data is required to determine whether these techniques will translate into superior outcomes in the years ahead.
- Citation: Papaioannou V, Jones J, Faraz F, Mayhew D, Rao A, Kahn M, Moharem-Elgamal S, Ashrafi R. Pacing in adults with complex congenital heart disease: A narrative minireview. World J Cardiol 2026; 18(11): 125622
- URL: https://www.wjgnet.com/1949-8462/full/v18/i11/125622.htm
- DOI: https://dx.doi.org/10.4330/wjc.125622
Congenital heart disease (CHD) of complex severity has previously lacked a definitive classification that is reproducible among physicians, but this was changed and simplified by the American Heart Association/American College of Cardiology 2018 guidelines for adults with CHD[1]. This guideline created the adult CHD anatomical and physiological classification system, and the conditions that make up this complex group are listed below[1]: (1) Cyanotic congenital heart defect; (2) Double-outlet ventricle; (3) Fontan procedure; Mitral atresia; (4) Single ventricle; (5) Pulmonary atresia; (6) Transposition of the great arteries; (7) Truncus arteriosus; and (8) Other abnormalities of atrioventricular and ventriculoarterial connection such as heterotaxy syndromes.
In a single review, it is difficult to encompass all conditions and pacing strategies; however, this review focuses on percutaneous pacing strategies in what may be considered by experts to be the three most difficult groups to pace in the complex classification: (1) Transposition of the great arteries-atrial switch (TGA-ASW); (2) Fontan circulation; and (3) Univentricular heart.
In this narrative minireview, we aim to address the available background evidence and newer percutaneous options in these situations and to provide a summary guide for physicians.
This expert narrative review was supported by a literature search of the available evidence. We searched PubMed from inception to June 2026 and reviewed appropriately cited articles identified during that period.
A combination of anatomical and technical search terms was used, as shown below: “Atrial switch = atrial switch, mustard or sensing, atrial redirection or D-transposition”, Fontan = Fontan Procedure or total cavopulmonary connection, atriopulmonary connection, lateral tunnel, or extracardiac conduit.
“Univentricular heart = univentricular physiology or single ventricle or single-ventricle physiology” AND “Pacing: “Pacemaker”, “cardiac pacing”, “permanent pacing”, “epicardial pacing”, “transvenous pacing”, “endocardial pacing”, “leadless pacemaker”, “leadless pacing”, “conduction system pacing”, “His-bundle pacing”, “left bundle branch area pacing”, “HBP”, “LBBAP” or “lead extraction”.
The search yielded 800 papers, of which 205 were analyzed for this article.
Inclusion criteria: Articles had to focus on pacing outcomes and techniques in the above groups in humans or contain significant data on pacing in these groups, even if the article’s focus was another area, such as arrhythmia burden.
Exclusion criteria: Articles not available, not in English, or no significant data on pacing.
In the initial data review, using published data grading systems no grade A or B-R evidence[2] was found as expected for pacing in complex CHD, in keeping with published guidelines[3]. Where possible, large-scale multi-center data has been used to inform work, although in some situations, data is limited to case reports or expert opinion. If no large-scale or long-term data were available for adults with CHD, good-quality pediatric data on CHD were used to supplement the review.
Supportive papers used to supplement the data with minimal or no evidence of CHD or to aid technical discussions were included at the author’s discretion.
This paper is designed to provide an educational summary. In many of these areas, there are limited data and no clear recommendations; therefore, data are extrapolated where useful, and small-center data are used when they are the only available data.
Estimations of CHD vary depending on the diagnostic criteria used for lesions and the severity of lesions, including a variation in incidence from 2 to 17/1000 live births[4,5]. The variation in incidence occurs when simple lesions, such as a normal functioning bicuspid valve, minor pulmonary stenosis, or a small atrial septal defect, are included, which results in studies reporting higher incidences. In addition, the above figures can only be used for the estimation of the clinical problem, as many patients with CHD may not be diagnosed at birth and sometimes not until adulthood.
For complex lesions defined using the Bethesda classification[1], the incidence in recent years at birth is suggested to have remained stable[6] with an estimated 2 per 1000/Live births reported[7].
Improvements in diagnosis, healthcare, and technologies have increased survival in several CHD analyses, increasing the population of CHD[8] and age of death[9] with survival of babies with CHD comparable to the general population[10]. In the current era, as best that can be estimated, around 60% of the complex population is now made up of adult patients with this data coming from large regional databases[11] and 20% of all adults with CHD in a large United States multicenter study had complex CHD[12] using older classification methods of complexity.
Similar to those in non-CHD cohorts, pacing requirements are likely to increase over time in an aging population, potentially because of multiple interventions or increasing fibrosis with age. The change in guidelines to upgrade recommendations for catheter ablation in CHD compared to the previous[13] may also be a factor in the coming years, as complication rates in CHD ablation may be higher than in a non-CHD cohort[14,15].
The prevalence of pacing is not fully clarified in complex CHD, often because of small groups across multiple institutions and variable data quality; for example, 7% of adults with CHD followed up in one large unit were found to have a pacemaker[16], and in a United States-based study, CHD patients as a whole represented 1% of all new implants[17]. While the numbers are significant alone, in one large single-center study, the rate of increase in pacing in CHD was three times the rate of increase in CHD surgical procedures[18] suggesting a large and increasing burden of work compared with other CHD interventions.
A summary of class 1 indications for pacing in complex CHD is shown in Table 1, based on consensus recommendations[3,19].
| Category | Indications |
| Sinus node dysfunction | Symptomatic |
| Bradycardia induced ventricular arrhythmias | |
| Refractory tachycardia-bradycardia syndrome | |
| Atrio-ventricular block | Symptomatic |
| Bradycardia induced ventricular arrhythmias | |
| Post-operative 3rd degree block persisting for more than 7 days | |
| Asymptomatic high degree atrioventricular node block with: (1) Low cardiac output; (2) Prolonged QT interval; (3) Ventricular dysfunction; (4) Complex ectopy; and (5) Wide escape rhythm |
The outcomes of pacing in complex CHD are difficult to define completely and are often obtained from case reports, single-center retrospective studies, and data included in other CHD pacing studies. In some situations, the amount of data may be limited owing to the novelty of the technology in use, such as leadless pacemakers (LPM) or conduction system pacing (CSP).
The oldest available data on pacing options are derived from pediatric epicardial pacing, which has been undertaken in complex CHD for many years. An atrial lead failure rate of 28% and a ventricular failure rate of 40% at 10 years has been reported[20] from a single-center retrospective analysis but seems broadly in line with similar single-center pediatric data[21] and single-center retrospective adult data[22]. These values can be used as a foundation for counselling by units though there will need to be an admission that rates may vary by unit experience. These figures are primarily influenced by patient background and implantation in small by weight and age children/neonates can lead to long-term poor lead survival[23].
The survival of transvenous leads seems comparatively superior than that of epicardial leads, tempered though by an increased incidence of reinterventions for reasons other than lead failure[24].
Reviewing other complications comparing pacing epicardial and endocardial modalities, single-center and retrospective data suggest the following: (1) Thromboembolic complications are higher from transvenous systems compared to epicardial systems[25,26]; (2) Long-term infection risks vary by single centre reporting but seem to be less than 5% for endocardial leads[16,27] and for epicardial systems, the risk is lower, although the numbers tend to be very small[28]; and (3) Only one small single centre study could be found to comment on the interaction of endocardial leads with atrioventricular valves in CHD, which suggested a 10% increase in regurgitation grade over 4 years[29].
Broadly, outcomes for pacing in complex CHD show a burden of complications, including high levels of lead revisions and infection[30] and at a higher rate than simple lesions[16].
LPM usage in clinical practice has increased in recent years[31] for several reasons, but there are very low complication rates in large non-CHD trials[32,33] and low infection rates when compared to traditional pacing systems in non-CHD-specific trials[34].
LPM use in complex CHD has increased, in addition to their lower infection risk, for several reasons: (1) No leads across the atrioventricular valve; (2) No long-term lead within the vasculature; (3) No baffle obstruction; and (4) Potentially reduced psychological impact because no generator is required[35].
LPM outcomes: LPMs have been commercially used since 2014 and more than a decade of research and real-world data are available, suggesting safe outcomes[36] and management of devices at the end of service[37].
Significant amounts of outcome data exist in non-CHD patients, but systematic data on LPMs in patients with complex CHD are scarce and are predominantly limited to promising case series[38,39] or pediatric series[40].
The largest summarized data in CHD with a mixed cohort of non-complex and complex patients totalled fewer than 30, with only 12 cases being complex CHD patients but showing no complications and effective pacing[41].
Compared with transvenous pacing, the differences in complications reported in the structurally normal heart cohort are: (1) Major tamponade rates were higher[42]; (2) Device embolization, which does not exist as a risk in the same way; and (3) Access-site complications[43].
Complications of LPMs in CHD are similar to those in patients with structurally normal hearts, with a reported complication rate of 0.5%[44] from a single registry. A difference that may exist between the normal heart and CHD groups may be a possible reduced rate of tamponade because patients with complex CHD often have had multiple sternotomies with adhesions that may preclude significant bleeding with tamponade.
LPM-current technology: LPM options now exist for several clinical indications and are available from various manufacturers. Summarized in Table 2[41,45-52] is the technical data for the options currently commercially available or close to final regulatory approval by the United States Drug and Food Administration for routine use as of August 2026.
| Micra VR | Micra AV | Aveir AR | Aveir VR | Empower | EBR wise | |
| Width (mm) | 6.7 | 6.7 | 6.5 | 6.5 | 6 | 2.7 |
| Length (mm) | 25.9 | 25.9 | 32 | 38 | 31.9 | 9.1 |
| Fixation | Passive tine | Passive tine | Active helix | Active helix | Passive tine | Active spike |
| Sensor | Accelerometer | Accelerometer | Temperature | Temperature | Accelerometer | N/A |
| Delivery catheter French size | 23 | 23 | 25 | 25 | 21 | 12 |
| Maximum mode | VVIR | VDD | AAIR | VVIR | VVIR | BIVP |
| Co-implant | No | No | Can be DR | Can be DR | With S-ICD | Mandatory |
| Designed to retrieve | No | No | Yes | Yes | Yes | No |
| Battery projected maximum | 16 years | 16 years | 10 years | 20 years | 10 years | 4.5 years |
| FDA Approved | Yes | Yes | Yes | Yes | No | Yes |
| Commercial use | Yes | Yes | Yes | Yes | No | Yes |
| Indication | Bradycardia pacing | Bradycardia pacing | Bradycardia pacing | Bradycardia pacing | Bradycardia pacing and anti-tachycardia ventricular pacing | Resynchronisation therapy |
| Paediatric use | Yes | Yes | Yes | Yes | No | No |
The Medtronic Micra™ system, which is available as a ventricular-only VVI system or AV system that senses mechanical atrial activity to provide VDD pacing, has the longest continuous data record.
The first LPM system was Nanostim™ from Abbott, but this was withdrawn and superseded with the Aveir™ system. Aveir™ is available as a ventricular or atrial system, or as a dual-chamber system with wireless communication between the two devices.
The Empower™ system by Boston Scientific provides leadless VVIR pacing which communicates wirelessly to a co-implanted subcutaneous defibrillator system. As a tandem system, this can then allow both defibrillation and anti-tachycardia pacing[53].
Lastly, the EBR Wise™ system consists of an electrode directly in the systemic ventricle linked to a co-implanted subpulmonic ventricular device, which provides a pacing stimulus sensed from the co-implanted device to offer biven
CSP has been available for some time, evolving initially from His-bundle pacing (HBP) to predominantly left bundle branch area pacing (LBBAP), whose uptake has increased significantly in recent years[54].
There are no large-scale studies on HBP in congenital cardiology, with most data being case reports and small projects[55]. The two largest published studies contained 11 HBP patients[56] and 29 patients[57] respectively, with good success rates but a tendency for worse lead performance in HBP vs LBBAP.
LBBAP, which has developed as a major CSP option for many implanters, has very good data in simple CHD and complex CHD with excellent safety and lead outcomes[57] including in many difficult anatomical challenges[58].
In the largest CSP series on CHD published, there was a statistically significant improvement in ventricular function in patients with a systemic left ventricle and no deterioration in systemic right ventricular function[57]. While there are case series and reports of CSP in many CHD groups and moderate- to complex lesions, in some situations, such as the Fontan cohort, there is only a single case report that highlights the lack of data and technical difficulties[59].
Long-term data are still not available for the extraction of HBP leads and LBBAP leads both in structurally normal hearts and particularly in the CHD cohort, and this remains a legitimate concern, particularly for complex CHD.
There are two predominant options for CSP: Lumenless leads or traditional stylet-driven leads, both of which require guide catheters to deploy the leads to the correct position. Operator preference tends to inform the choice as to which lead to use, but the sole lumenless option in routine use, the Selectsecure™ (Medtronic), is smaller at 4fr than other leads, which may be important in certain conditions. In the largest study of lead integrity for CSP leads (no comment on the inclusion of CHD patients), the failure rate of lumenless leads was 2 of 5609 patients and 10 of 2646 for stylet-driven leads, although all failures in the stylet group came from Biotronik Solia™ leads and none from other manufacturers[60].
Lead extraction in complex CHD is often confined to specialist centers and in small numbers overall, compared to non-CHD extractions. The reported data for transvenous extraction outcomes in CHD is comparable to non-CHD extraction in the largest single-center 20-year study showing no procedural deaths and a high success rate, though with a higher-than-expected need for valve surgical intervention[61]. The outcomes of other studies, including patients with a Fontan operation undergoing transvenous lead extraction, show a low rate of complications[62,63]. The use of mechanical tools and laser extraction seem to have increased the success rates of extraction in CHD[63,64].
Overall, the multiple single-center data cited above suggest potentially comparable outcomes of CHD to non-CHD extraction with the caveat that data is small and may bias to larger expert centers.
An example of the challenges faced in the extraction is shown in Figure 1 below from a patient with a complex CHD.
Epicardial lead removal is at best confined to a small number of case reports with no large-scale high-quality data. In a small number of cases, including CHD and non-CHD cases, surgical removal can be performed with difficulty and at some risk to the patient, whereas all other options have failed[65,66].
Extraction has not been a focus in LPMs, but situations are likely to occur when extraction is wanted or needed for clinical reasons. In increasingly complex anatomy, extraction may be more difficult and to date, data on long-term extraction in CHD is lacking. Currently, most information is available from Nanostim™ data out to 9 years with reasonable success rates in a predominantly non-CHD mixed patient cohort[67].
It is recommended that extraction in complex CHD is only undertaken in expert centres where there are experienced extracting physicians, critical care facilities and surgical support immediately available.
In structural CHD, there are no major case series or significant series of case reports on LPM extraction; therefore, recommendations are not published and expert opinion is required in difficult situations.
In structural CHD, operators will need to review and consider the particular challenges involved such as: (1) Vascular anatomical issues- e.g. anomalous venous return; (2) Abandoned leads with obstruction/reduced vascular luminal size; (3) Valve replacements- tissue or stented; (4) Vascular calcification with higher risks of increased bleeding risk or reduced compliance making sheath progress difficult; (5) Anatomical challenges such as dextrocardia, Ebstein’s anomaly and small ventricular size; and (6) Tine entanglement with the Micra™/Empower™[68].
Figure 2 shows an example of chronic extraction of an LPM in a patient with repaired tetralogy of Fallot and multiple abandoned leads to create more long-term space within the right ventricle. The device had been in place for two years, and thresholds had risen, so the device was extracted to make more room for a new device. A dedicated trilumen LPM retrieval sheath was used to remove the device, using a protective sleeve to cover the Micra™ tines to prevent entanglement and valve injury.
Adequate understanding of the underlying anatomy, surgical repair, and current pacing is key when pacing in complex CHD and commonly up-to-date cross-sectional imaging may be undertaken before beginning to help with: (1) Vascular access-vascular stenoses or occlusions due to multiple childhood procedures[69] or anatomical vascular changes such as an interrupted inferior vena cava (IVC)[70]; (2) Stent patency-this may be of particular importance in the atrial switch cohort[71]; and (3) Confirmation or exclusion of thrombus or calcification in the heart or vascular system[72].
In most cases, expert opinion would suggest computed tomography (CT) as the modality of choice for periprocedural planning due to its excellent spatial resolution, calcium and thrombus diagnosis[73] and the fact that there will not be artifacts from devices or stents in the same way as magnetic resonance imaging[74].
Modern CT is performed extremely quickly with low radiation doses and can cover a wide area, enabling a lot of information to be obtained to help plan procedures, including airway and lung function, if anesthetic support is needed. Figure 3, shown below, highlights an example of good-quality CT used to plan a transbaffle puncture in complex CHD.
Most LPMs require large-bore access of at least 12 French and the use of vascular ultrasound is suggested to reduce access site complications[75].
Many patients with complex CHD are anticoagulated, and although there are no CHD-specific data for anticoagulated patients undergoing pacing procedures, most operators and centers would take a tailored approach while considering positive data from trials[76] not specifically focused on CHD. Interruption of anticoagulation for procedures remains a choice for the operator individually for each procedure.
The level of evidence for these newer techniques in the complex CHD groups discussed previously is summarized in Table 3[39,57,59,77-83].
| Technique | TGA-ASW | Fontan | Univentricular |
| CSP | Case reports-multiple | Case report-single | None |
| LPMs | Case reports-multiple | Case reports-multiple Case series | Case reports-multiple Case series |
Although standard transvenous pacing in TGA-ASW has been in routine use for many years[84] with changes in technology, there are differing options available and newer techniques that can be utilized.
The first of the newer techniques we will look at is LBBAP in the TGA-ASW group, which has gained momentum as a method of resynchronization for a failing systemic RV[82]. In this situation, due to the redirection of the baffle, the standard ventricular lead trajectory will normally be to the left ventricular apex or free wall; an example is shown in Figure 4.
Using a reshaped guide catheter as described previously, with the primary curve reversed[82] and aimed anteriorly rather than posteriorly, the ventricular lead can be turned toward the septum; an example is shown in Figure 5.
Occasionally, the coronary sinus (CS) can be cannulated in patients with TGA-ASW in the systemic venous baffle zone and operators can attempt resynchronization by placing a CS lead to a middle cardiac vein branch or some other vein traversing towards the right ventricle[85]. This approach is heavily dependent on the available anatomy and may not always be possible as shown in Figure 6, highlighting a CS draining only towards the left ventricle.
Prior to LBBAP for the TGA-ASW cohort, a small number of studies had used the placement of a lead into the systemic right ventricle (Figure 7) via a baffle puncture[86]. The data from this small number of cases are good[87] though the technique has been superseded by LBBAP in most centers.
A particular weakness of the TGA-ASW is baffle leak or stenosis, and in this instance, leadless devices have a particular attraction due to the lack of long-term vascular obstruction. This has progressed rapidly from initial back up ventricular pacing[83] to leadless atrial pacing[88] with an example shown in Figure 8, and now dual-chamber leadless pacing[89] with acceptable safety and pacing data.
Percutaneous pacing in the Fontan cohort is often a source of significant technical challenges for operators, and we have summarized some techniques available for each type of Fontan operation. The three types of Fontan operations are shown (Figure 9)[90].
Atriopulmonary Fontan (APF), in which the right atrial appendage is connected to the pulmonary artery to provide pulmonary blood flow in conditions such as tricuspid atresia.
Lateral tunnel Fontan (LT), where the superior vena cava (SVC) is anastomosed to the pulmonary artery and an intracardiac tunnel is formed using a patch in the atrium to divert IVC blood flow into the tunnel and up to the pulmonary artery. This has been used in many conditions where there is no subpulmonary ventricle.
The extracardiac conduit Fontan (ECC) is the most recent development, where like the LT the SVC connects to the pulmonary artery, but now the IVC flow diverts to the pulmonary artery via an extracardiac conduit. This results in no systemic venous return within the heart.
In this cohort, there is no consensus on percutaneous techniques, and data comes mostly from case series or reports[91]. Most of these techniques are therefore only used in expert centers, often where traditional surgical epicardial leads have failed, and when the risk of surgical intervention is thought to be unacceptable. Due to the limited number of cases, it cannot be suggested that one technique is superior to another, and the examples shown are offered mainly as potential solutions to clinical challenges.
In the APF group, atrial pacing was largely performed in a standard manner; the main challenges were the lead length required in massive atrial dilatation and the need to find healthy tissue despite progressive fibrosis over time.
The major challenge for operators is ventricular pacing, which is dependent on the severity of the lesion, as there may be no access to a subpulmonary ventricle in a normal manner, such as in tricuspid atresia. The options that can be used to overcome this dependence on the exact available anatomy are summarized below.
Pacing- to a residual atrio-ventricular-valve ring or pocket of tissue: Here, sometimes even with tricuspid atresia, there is still a small area of annular tissue with ventricular capture that can be utilized with an active lead to screw and ventricular pacing. Sometimes, there can be a small valvular orifice and hypoplastic ventricle with tissue pockets.
CS only pacing: In some patients with APF, the CS may still drain into the main atrial mass and so can be accessed for pacing. Due to atrial dilatation, this can be difficult, and the CS anatomy may be abnormal[92].
Pacing- to a residual sub-pulmonary outflow tract or pulmonary stump via the atrio-pulmonary connection: via the right atrial appendage to the pulmonary artery connection, it may be possible to wire up into the pulmonary arteries and, in some patients, progress retrograde to a stump of the pulmonary valve or sometimes through this to the ventricular mass. This often requires a shaped guide catheter to make turns before a lead can be maneuvered to this area, and it is not always possible to obtain electrical capture in the stump depending on the position and level of hypoplasia or where the surgical team may have ligated tissue (Figure 10).
In the LT group, there was much less atrial dilatation than in the APF group, and there was natural atrial tissue available to pace, making things a little less difficult than in other Fontan situations, but the same difficulties exist in accessing ventricular tissue as faced with the APF group.
Similarly, it is possible to retrogradely wire from the pulmonary arteries, as shown below, where the stump was wired and dilated sufficiently for a guide catheter to pass and deliver a lead into the ventricle (Figure 11).
Where this option does not exist, access may need to be created to the ventricular mass; sometimes, this may already exist via a fenestration made at the time of Fontan surgery or may need to be created percutaneously.
The easiest option is to cross the fenestration superiorly, and we would recommend treating the fenestration similarly to a CS and using a wire and guide catheter to cross the wire support before trying to deliver a lead. Sometimes, it may be necessary to increase the size of the fenestration, and this is best done before starting superiorly using a non-compliant balloon, often at approximately 6-8 mm in size from the leg (Figure 12).
In the ECC group, there will not be any normal access to electrically active tissue, and almost always some form of fenestration or baffle puncture will be needed to secure paceable tissue.
The small number of exceptions to this is that, occasionally with a deep enough screw of the lead, atrial tissue on the other side of the conduit may be capturable, and this has been performed with leadless devices as well[79]. The other exception is that very occasionally paceable tissue for the atrium may be found via the pulmonary arterial system[93] or, as we have shown above, via the pulmonary outflow stump for ventricular capture.
Transbaffle pacing has become an option for patients in whom epicardial pacing may not be desired owing to surgical risk, patient preference, or previous epicardial lead failure. Several techniques have been described, including transpulmonary puncture, to allow access to the epicardial surface of the atrium for atrial pacing[81].
In addition to the option of a transvenous lead via existing fenestration, as described in the LT group above, or via created fenestrations, an option to pace leadless in the ECC Fontan group is possible.
The major difficulty in pacing using an LPM in the ECC group is creation of access and manipulation of the large bore sheath to the pulmonary venous atrium before delivering the LPM, with an example of the technique (Figure 13).
In situations where the femoral approach may not be feasible, it is possible to puncture superiorly from the internal jugular approach via the pulmonary artery bed (Figure 14).
After puncture, the angle from the superior approach is often simpler and more direct, leading to a relatively shorter procedure time.
The only other option that can be considered is a retrograde aortic approach using an LPM (Figure 15)[77] and using a carotid artery cutdown, as has been described in transcatheter aortic valve implantation[94].
Carotid access provides a straightforward angle and route for ventricular devices, but the risk of vascular thrombotic complications is always a major concern. Therefore, it remains a final percutaneous bailout option and always requires surgical support for access and closure.
The evidence for systemic leads in the Fontan group, including from a thrombotic perspective, is as reassuring as possible given these patients’ higher thrombotic risk, and the risks seem manageable if patients are anticoagulated long term[26]. The evidence for LPMs in this group is very limited, as this is an evolving technique and largely case series/report-based, though seems positive so far[78,80] but will require further evidence in the long term.
This is an extremely heterogeneous group, and as a rule, it is difficult to cover all the anatomical complexities that implanters may see and how to overcome them.
For some univentricular systems, ventricular lead placement may be difficult because of large ventricular septal defects, systemic pressure levels, and clot concerns on the leads in the systemic ventricle. In this setting, it may be reasonable to consider CS only lead placement to try and mitigate some of these risks, as in the non-CHD patients, the rate of thrombotic complications with a systemic ventricular lead is quoted at 7%[95]. Currently, there is no high-quality evidence regarding the rate of thrombotic events in patients with univentricular endocardial leads.
LPMs can be used both in the atrium and ventricle of univentricular patients and given the ongoing risk of infection for cyanotic patients, they have theoretical benefits over lead-based systems, with the parylene coating being resistant to infection and thrombotic adhesion[34].
A patient with a univentricular heart and a previous Glen anastomosis, with no superior access, initially received an atrial LPM as part of arrhythmia suppression before a ventricular device was inserted transeptally via a large atrial septal defect and the atrioventricular node was ablated to achieve rate control (Figure 16).
For complex transbaffle procedures, LPM implantation, and extraction in complex CHD, substantial experience and structural support are needed. Based on expert opinion, we suggest the following requirements: (1) Electrophysiologist and device doctor with expertise in CHD; (2) Structural CHD interventionalist availability; (3) CHD surgical support available 24/7; (4) Vascular surgical support available 24/7; (5) Imaging cardiology support with expertise in CHD; (6) Anaesthetic and critical care doctor with expertise in CHD; (7) Modern biplane catheter lab; (8) Cross sectional imaging with high-quality CT imaging; (9) Critical care ward with ability to care for the patient with CHD; and (10) This list is not definitive or prescriptive but is our suggestion based on expert opinion.
Table 4 summarizes the potential advantages and disadvantages of the newer pacing techniques in complex CHD discussed already by lesion category on an expert opinion basis as there are no formal comparison studies. Currently, the summarized potential below is based on the available evidence, which in many situations is not long-term and is in very small numbers, as discussed earlier.
| Lesion | CSP-advantages | CSP-disadvantages | LPM-advantages | LPM- disadvantages |
| TGA-ASW | Physiological pacing/treatment of systemic ventricular dysfunction | No long-term data/Leads through the baffle/sheath reshaping and delivery difficulty | Lower infection risk/no baffle obstruction | Difficulty negotiating baffles/difficult to place passive LPM in a smooth walled sub pulmonary left ventricle |
| APF | As above | Technical difficulties delivering to conduction system area | Superior catheter support in a dilated atrium to deliver atrial LPMs | Large bore vascular access/potentially higher embolisation risk for atrial devices/transeptal delivery for ventricular devices/risk of embolisation in a systemic ventricle |
| LT | As above | Thrombotic risk with systemic ventricular lead/fenestration creation difficulty if needed | As above/could be lower thrombotic risk than a systemic endocardial lead | As above |
| ECC | As above | As above- same risk with atrial lead if in pulmonary venous atrium | Could be lower thrombotic risk than a systemic endocardial lead | As above |
| Univentricular | As above | Thrombotic risk increased with endocardial leads | As above | Large bore vascular access/embolisation risk to the systemic circulation |
In complex situations, it is difficult to offer a prescriptive guide to patient and device selection because of the limited, short-term data discussed above. Table 5 summarizes an expert opinion-based pathway (level of evidence-C-EO), not a prescriptive recommendation.
| Indication | TGA-ASW | APF | LT | ECC | Univentricular |
| SND only | Endocardial lead system or atrial LPM | Endocardial lead system or atrial LPM | Endocardial lead system or atrial LPM | Preferentially epicardial system first choice and transbaffle endocardial lead system or LPM if surgical turndown | Endocardial lead system or atrial LPM |
| SND and AVB | Endocardial lead system or dual chamber LPM | Endocardial lead system- epicardial as back up if surgically suitable and endocardial access failed. Ventricle LPM only if surgical turndown | Endocardial lead system- epicardial as back up if surgically suitable and endocardial access failed. Ventricle LPM only if surgical turndown | Preferentially epicardial system first choice and transbaffle endocardial lead system | Endocardial lead system |
| AVB-low burden of pacing predicted | Ventricular LPM- active fix device to the sub pulmonary left ventricle | As above | As above | Preferentially epicardial system first choice and transbaffle LPM if surgical turndown | Ventricular LPM |
| AVB- high burden of pacing predicted | Endocardial lead system- CSP reasonable to offer upfront or VDD or dual chamber LPM | As above | As above | Preferentially epicardial system first choice and transbaffle endocardial lead system | Endocardial lead system |
| Any of the above and baffle stenosis | Atrial, ventricular or dual chamber LPM | Preferentially epicardial system first choice and LPM if surgical turndown | Preferentially epicardial system first choice and transbaffle LPM if surgical turndown | ||
| Heart failure with wide QRS | CSP with endocardial LBBAP | Endocardial lead system- epicardial as back up if surgically suitable and endocardial not feasible | Dependent on surgical risk- transbaffle endocardial lead system if high risk vs epicardial if lower risk | Dependent on surgical risk- transbaffle endocardial lead system if high risk vs epicardial if lower risk | Traditional resynchronisation pacing or CSP with endocardial LBBAP if no coronary sinus |
We have reviewed the evidence for pacing in complex CHD and some of the more novel techniques now available. Although an increasing choice of techniques may improve outcomes for this cohort, this remains to be shown.
Currently, there is a lack of long-term data, high-quality comparative data, and certainly no randomized data for physicians to work with, although it is hoped that this will change as larger registries are built between centers.
This article, like many in this area, is hindered due to selection and referral bias in that many of these cases will be referred to and managed by expert centres who are then the most likely to publish the data with potentially higher success rate than may be achieved in other locations. The paucity of wider data does mean that the publication bias leads to papers like this relying on a small cohort of centres, small numbers of patients and sometimes decision-making pathways that may not be applicable in other centres. We accept these limitations and hope that the educational value alerts centres that some other options exist even if they are not available locally or routinely recommended.
Several important research questions remain to be addressed in the future: (1) Management of LPMs at the end of service life in complex CHD; (2) Long-term retrieval of LPMs if needed (e.g., for infection); (3) Is CSP a suitable long-term treatment for a failing ventricle in TGA-ASW? (4) Can CSP be used in the failing Fontan patient? (5) Anticoagulation strategies in LPM procedures in complex CHD; and (6) Long-term performance of LPMs in the systemic circulation in Fontan patients and associated thrombotic risk.
At this point, it is not possible to know how much of this data is likely to be of high quality, although it would seem possible that comparative data on CSP in the TGA-ASW cohort would be possible, as would data on LPMs in this cohort in the coming years.
The key findings of this review are as follows: (1) The number of adults with complex CHD is increasing; (2) Much of the evidence and guidance in complex CHD pacing are expert opinion-based or based on a very small number of patients; (3) LPM use in complex CHD seems safe and effective in the small number of cases described; (4) CSP in complex CHD shows early promise in very small numbers; (5) Lead extraction in complex CHD in expert centers seems safe; and (6) Newer percutaneous techniques can be used even in the most challenging CHD anatomies, where traditional options may not be available.
It is acknowledged that many of the options discussed here are not based on large-scale evidence and cannot be considered guideline recommendations; instead, they offer options where none currently exist or alternatives in prohibitively high-risk situations.
In the years to come, as technology, case numbers, and operator experience change, the threshold for some of the above techniques may differ, and we await larger datasets to decide on the optimal treatment for these complex CHD cohorts.
| 1. | Stout KK, Daniels CJ, Aboulhosn JA, Bozkurt B, Broberg CS, Colman JM, Crumb SR, Dearani JA, Fuller S, Gurvitz M, Khairy P, Landzberg MJ, Saidi A, Valente AM, Van Hare GF. 2018 AHA/ACC Guideline for the Management of Adults With Congenital Heart Disease: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. J Am Coll Cardiol. 2019;73:e81-e192. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 297] [Cited by in RCA: 734] [Article Influence: 91.8] [Reference Citation Analysis (0)] |
| 2. | Brennan SO, Gordon B, Anderson CS, Coutts SB, Dawson J, Hill MD, Lemmens R, Menon BK, Price CI, Aguiar de Sousa D, Thomalla G, Tsivgoulis G, Turc G, Kelly PJ, McCabe JJ. Quality of Evidence in European Stroke Organisation and American Heart Association Stroke Guidelines. Stroke. 2026;57:381-390. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1] [Cited by in RCA: 2] [Article Influence: 2.0] [Reference Citation Analysis (0)] |
| 3. | Kusumoto FM, Schoenfeld MH, Barrett C, Edgerton JR, Ellenbogen KA, Gold MR, Goldschlager NF, Hamilton RM, Joglar JA, Kim RJ, Lee R, Marine JE, McLeod CJ, Oken KR, Patton KK, Pellegrini CN, Selzman KA, Thompson A, Varosy PD. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2019;140:e382-e482. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 51] [Cited by in RCA: 226] [Article Influence: 32.3] [Reference Citation Analysis (0)] |
| 4. | Hoffman JI, Kaplan S. The incidence of congenital heart disease. J Am Coll Cardiol. 2002;39:1890-1900. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 4784] [Cited by in RCA: 3829] [Article Influence: 159.5] [Reference Citation Analysis (7)] |
| 5. | van der Linde D, Konings EE, Slager MA, Witsenburg M, Helbing WA, Takkenberg JJ, Roos-Hesselink JW. Birth prevalence of congenital heart disease worldwide: a systematic review and meta-analysis. J Am Coll Cardiol. 2011;58:2241-2247. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 2852] [Cited by in RCA: 2474] [Article Influence: 164.9] [Reference Citation Analysis (12)] |
| 6. | Best KE, Draper E, Kurinczuk J, Stoianova S, Tucker D, Wellesley D, Rankin J. OP29 The prevalence of congenital heart disease in the UK: a population-based register study. J Epidemiol Community Health. 2014;68:A17-A17. [RCA] [DOI] [Full Text] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 7. | Parker DM, Stabler ME, MacKenzie TA, Zimmerman MS, Shi X, Everett AD, Bucholz EM, Brown JR. Population-Based Estimates of the Prevalence of Children With Congenital Heart Disease and Associated Comorbidities in the United States. Circ Cardiovasc Qual Outcomes. 2024;17:e010657. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 24] [Reference Citation Analysis (0)] |
| 8. | Dellborg M, Giang KW, Eriksson P, Liden H, Fedchenko M, Ahnfelt A, Rosengren A, Mandalenakis Z. Adults With Congenital Heart Disease: Trends in Event-Free Survival Past Middle Age. Circulation. 2023;147:930-938. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1] [Cited by in RCA: 106] [Article Influence: 35.3] [Reference Citation Analysis (0)] |
| 9. | Kholeif Z, Abozied O, Abdelhalim AT, ElZalabany S, Moustafa A, Ali A, Egbe AC. Temporal change in the age at time of death in adults with congenital heart disease. Int J Cardiol Congenit Heart Dis. 2025;20:100578. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 2] [Reference Citation Analysis (0)] |
| 10. | Goldstein SA, D'Ottavio A, Spears T, Chiswell K, Hartman RJ, Krasuski RA, Kemper AR, Meyer RE, Hoffman TM, Walsh MJ, Sang CJ, Paolillo J, Li JS. Causes of Death and Cardiovascular Comorbidities in Adults With Congenital Heart Disease. J Am Heart Assoc. 2020;9:e016400. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 7] [Cited by in RCA: 29] [Article Influence: 4.8] [Reference Citation Analysis (0)] |
| 11. | Marelli AJ, Ionescu-Ittu R, Mackie AS, Guo L, Dendukuri N, Kaouache M. Lifetime prevalence of congenital heart disease in the general population from 2000 to 2010. Circulation. 2014;130:749-756. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 678] [Cited by in RCA: 945] [Article Influence: 78.8] [Reference Citation Analysis (0)] |
| 12. | Gurvitz M, Dunn JE, Bhatt A, Book WM, Glidewell J, Hogue C, Lin AE, Lui G, McGarry C, Raskind-Hood C, Van Zutphen A, Zaidi A, Jenkins K, Riehle-Colarusso T. Characteristics of Adults With Congenital Heart Defects in the United States. J Am Coll Cardiol. 2020;76:175-182. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 17] [Cited by in RCA: 57] [Article Influence: 9.5] [Reference Citation Analysis (0)] |
| 13. | Philip Saul J, Kanter RJ; WRITING COMMITTEE, Abrams D, Asirvatham S, Bar-Cohen Y, Blaufox AD, Cannon B, Clark J, Dick M, Freter A, Kertesz NJ, Kirsh JA, Kugler J, LaPage M, McGowan FX, Miyake CY, Nathan A, Papagiannis J, Paul T, Pflaumer A, Skanes AC, Stevenson WG, Von Bergen N, Zimmerman F. PACES/HRS expert consensus statement on the use of catheter ablation in children and patients with congenital heart disease: Developed in partnership with the Pediatric and Congenital Electrophysiology Society (PACES) and the Heart Rhythm Society (HRS). Endorsed by the governing bodies of PACES, HRS, the American Academy of Pediatrics (AAP), the American Heart Association (AHA), and the Association for European Pediatric and Congenital Cardiology (AEPC). Heart Rhythm. 2016;13:e251-e289. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 207] [Cited by in RCA: 179] [Article Influence: 17.9] [Reference Citation Analysis (0)] |
| 14. | Waldmann V, Duthoit G, Pasquié JL, Champ-Rigot L, Albertini M, Anselme F, Bartoletti S, Bonnet D, Bredy C, Bun SS, Clerici G, Da Costa A, De Chillou C, Defaye P, de Guillebon M, Davril C, Delinière A, Derval N, Ditac G, Gardey K, Ghanimé C, Gourraud JB, Hascoet S, Hammache N, Henaine R, Iserin L, Jacon P, Jourda F, Karsenty C, Koutbi L, Laurent G, Maille B, Maltret A, Mansourati J, Marimpouy N, Martins R, Maury P, Milhem A, Moceri P, Ollitrault P, Pinon P, Piot O, Richard-Vitton R, Sacher F, Sebag F, Tortigue M, Venier S, Wilkin M, Winum P, Marijon E, Combes N, Bessière F. Catheter ablation in congenital heart diseases: a French nationwide study. Eur Heart J. 2026;47:816-827. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 6] [Cited by in RCA: 14] [Article Influence: 14.0] [Reference Citation Analysis (0)] |
| 15. | Mori H, Nishii N, Kato Y, Kinjo T, Suzuki T, Yoshimoto J, Tateno S, Miyazaki A, Aoki H, Higuchi S, Nakagawa K, Toyohara K, Takeuchi D, Sakai T, Takefuta K, Henmi R, Mori M, Yagishita D, Hasegawa S, Matsui Y, Watanabe S, Kikuchi K, Kani K, Hiranuma Y, Matsuyama T, Matsumoto K, Narita M, Sasaki W, Nabeshima T, Nagaoka K, Kojima T, Kato R, Shoda M, Sumitomo N. REgistry Of Catheter AbLation After Congenital Heart Disease Surgery(REAL-CHD Registry). J Arrhythm. 2026;42:e70347. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 16. | Chami J, Moore BM, Nicholson C, Cordina R, Baker D, Celermajer DS. Outcomes of permanent pacemakers and implantable cardioverter-defibrillators in an adult congenital heart disease population. Int J Cardiol Congenit Heart Dis. 2024;15:100490. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 17. | Hayward RM, Dewland TA, Moyers B, Vittinghoff E, Tanel RE, Marcus GM, Tseng ZH. Device complications in adult congenital heart disease. Heart Rhythm. 2015;12:338-344. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 14] [Cited by in RCA: 22] [Article Influence: 1.8] [Reference Citation Analysis (0)] |
| 18. | Mondal A, Yoo M, Tuttle S, Mah D, Nelson R, Sachse FB, Hitchcock R, Kaza AK. Cost of Pacing in Pediatric Patients With Postoperative Heart Block After Congenital Heart Surgery. JAMA Netw Open. 2023;6:e2341174. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 7] [Cited by in RCA: 8] [Article Influence: 2.7] [Reference Citation Analysis (0)] |
| 19. | Hernández-Madrid A, Paul T, Abrams D, Aziz PF, Blom NA, Chen J, Chessa M, Combes N, Dagres N, Diller G, Ernst S, Giamberti A, Hebe J, Janousek J, Kriebel T, Moltedo J, Moreno J, Peinado R, Pison L, Rosenthal E, Skinner JR, Zeppenfeld K; ESC Scientific Document Group. Arrhythmias in congenital heart disease: a position paper of the European Heart Rhythm Association (EHRA), Association for European Paediatric and Congenital Cardiology (AEPC), and the European Society of Cardiology (ESC) Working Group on Grown-up Congenital heart disease, endorsed by HRS, PACES, APHRS, and SOLAECE. Europace. 2018;20:1719-1753. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 253] [Cited by in RCA: 228] [Article Influence: 28.5] [Reference Citation Analysis (1)] |
| 20. | Lau KC, William Gaynor J, Fuller SM, Karen A Smoots, Shah MJ. Long-term atrial and ventricular epicardial pacemaker lead survival after cardiac operations in pediatric patients with congenital heart disease. Heart Rhythm. 2015;12:566-573. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 31] [Cited by in RCA: 43] [Article Influence: 3.9] [Reference Citation Analysis (0)] |
| 21. | Silvetti MS, Drago F, Di Carlo D, Placidi S, Brancaccio G, Carotti A. Cardiac pacing in paediatric patients with congenital heart defects: transvenous or epicardial? Europace. 2013;15:1280-1286. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 35] [Cited by in RCA: 54] [Article Influence: 4.2] [Reference Citation Analysis (0)] |
| 22. | Walker F, Siu SC, Woods S, Cameron DA, Webb GD, Harris L. Long-term outcomes of cardiac pacing in adults with congenital heart disease. J Am Coll Cardiol. 2004;43:1894-1901. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 64] [Cited by in RCA: 56] [Article Influence: 2.5] [Reference Citation Analysis (0)] |
| 23. | Zeriouh S, Koutbi L, Mancini J, Maille B, Hourdain J, Franceschi F, Deharo JC. Permanent epicardial pacing: a first line strategy in children. EP Europace. 2024;26:Suppl 1. [RCA] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1] [Cited by in RCA: 2] [Article Influence: 1.0] [Reference Citation Analysis (2)] |
| 24. | Bowman HC, Shannon KM, Biniwale R, Moore JP. Cardiac implantable device outcomes and lead survival in adult congenital heart disease. Int J Cardiol. 2021;324:52-59. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 4] [Cited by in RCA: 14] [Article Influence: 2.3] [Reference Citation Analysis (0)] |
| 25. | Khairy P, Landzberg MJ, Gatzoulis MA, Mercier LA, Fernandes SM, Côté JM, Lavoie JP, Fournier A, Guerra PG, Frogoudaki A, Walsh EP, Dore A; Epicardial Versus ENdocardial pacing and Thromboembolic events Investigators. Transvenous pacing leads and systemic thromboemboli in patients with intracardiac shunts: a multicenter study. Circulation. 2006;113:2391-2397. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 208] [Cited by in RCA: 215] [Article Influence: 10.8] [Reference Citation Analysis (3)] |
| 26. | Assaad IE, Pastor T, O'Leary E, Gauvreau K, Rathod RH, Gurvitz M, Wu F, Fynn-Thompson F, DeWitt ES, Mah DY. Atrial pacing in Fontan patients: The effect of transvenous lead on clot burden. Heart Rhythm. 2021;18:1860-1867. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 17] [Cited by in RCA: 19] [Article Influence: 3.8] [Reference Citation Analysis (0)] |
| 27. | Takeuchi D, Toyohara K, Yagishita D, Yazaki K, Higuchi S, Ejima K, Shoda M, Hagiwara N. Acute and Long-Term Outcomes of Transvenous Cardiac Pacing Device Implantation in Patients With Congenital Heart Disease. Circ Rep. 2019;1:445-455. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 2] [Reference Citation Analysis (0)] |
| 28. | McLeod CJ, Attenhofer Jost CH, Warnes CA, Hodge D 2nd, Hyberger L, Connolly HM, Asirvatham SJ, Dearani JA, Hayes DL, Ammash NM. Epicardial versus endocardial permanent pacing in adults with congenital heart disease. J Interv Card Electrophysiol. 2010;28:235-243. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 48] [Cited by in RCA: 62] [Article Influence: 3.9] [Reference Citation Analysis (0)] |
| 29. | Webster G, Margossian R, Alexander ME, Cecchin F, Triedman JK, Walsh EP, Berul CI. Impact of transvenous ventricular pacing leads on tricuspid regurgitation in pediatric and congenital heart disease patients. J Interv Card Electrophysiol. 2008;21:65-68. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 38] [Cited by in RCA: 39] [Article Influence: 2.1] [Reference Citation Analysis (0)] |
| 30. | Klug D, Vaksmann G, Jarwé M, Wallet F, Francart C, Kacet S, Rey C. Pacemaker lead infection in young patients. Pacing Clin Electrophysiol. 2003;26:1489-1493. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 55] [Cited by in RCA: 48] [Article Influence: 2.1] [Reference Citation Analysis (0)] |
| 31. | Khan MZ, Nassar S, Nguyen A, Khan MU, Sattar Y, Alruwaili W, Gonuguntla K, Mazek H, Asad ZUA, Agarwal S, Raina S, Balla S, Nguyen B, Fan D, Darden D, Munir MB. Contemporary trends of leadless pacemaker implantation in the United States. J Cardiovasc Electrophysiol. 2024;35:1351-1359. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 11] [Cited by in RCA: 13] [Article Influence: 6.5] [Reference Citation Analysis (0)] |
| 32. | Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, Friedman PA, Estes NA 3rd, Ip J, Niazi I, Plunkitt K, Banker R, Porterfield J, Ip JE, Dukkipati SR; LEADLESS II Study Investigators. Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker. N Engl J Med. 2015;373:1125-1135. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 327] [Cited by in RCA: 396] [Article Influence: 36.0] [Reference Citation Analysis (0)] |
| 33. | Reddy VY, Knops RE, Sperzel J, Miller MA, Petru J, Simon J, Sediva L, de Groot JR, Tjong FV, Jacobson P, Ostrosff A, Dukkipati SR, Koruth JS, Wilde AA, Kautzner J, Neuzil P. Permanent leadless cardiac pacing: results of the LEADLESS trial. Circulation. 2014;129:1466-1471. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 198] [Cited by in RCA: 221] [Article Influence: 18.4] [Reference Citation Analysis (0)] |
| 34. | El-Chami MF, Bonner M, Holbrook R, Stromberg K, Mayotte J, Molan A, Sohail MR, Epstein LM. Leadless pacemakers reduce risk of device-related infection: Review of the potential mechanisms. Heart Rhythm. 2020;17:1393-1397. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 36] [Cited by in RCA: 126] [Article Influence: 21.0] [Reference Citation Analysis (0)] |
| 35. | Sears SF, Jordan EW, Hashmath Z, Sargeant MM, Catanzaro J, Nekkanti R, Shantha G. Leadless Pacemakers: The "Leading Edge" of Quality of Life in Cardic Electrophysiology. Curr Cardiol Rep. 2025;27:77. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1] [Cited by in RCA: 3] [Article Influence: 3.0] [Reference Citation Analysis (1)] |
| 36. | El-Chami MF, Garweg C, Clementy N, Al-Samadi F, Iacopino S, Martinez-Sande JL, Roberts PR, Tondo C, Johansen JB, Vinolas-Prat X, Cha YM, Grubman E, Bordachar P, Stromberg K, Fagan DH, Piccini JP. Leadless pacemakers at 5-year follow-up: the Micra transcatheter pacing system post-approval registry. Eur Heart J. 2024;45:1241-1251. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 23] [Cited by in RCA: 101] [Article Influence: 50.5] [Reference Citation Analysis (0)] |
| 37. | Andreae A, Breitenstein A, Piccini JP. Long-term management of leadless pacemakers. Eur Heart J Suppl. 2025;27:ii26-ii38. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 11] [Reference Citation Analysis (0)] |
| 38. | Russell MR, Galloti R, Moore JP. Initial experience with transcatheter pacemaker implantation for adults with congenital heart disease. J Cardiovasc Electrophysiol. 2019;30:1362-1366. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 20] [Cited by in RCA: 18] [Article Influence: 2.6] [Reference Citation Analysis (0)] |
| 39. | Bassareo PP, Walsh KP. Micra pacemaker in adult congenital heart disease patients: A case series. J Cardiovasc Electrophysiol. 2022;33:2335-2343. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 12] [Reference Citation Analysis (0)] |
| 40. | Shah MJ, Borquez AA, Cortez D, McCanta AC, De Filippo P, Whitehill RD, Imundo JR, Moore JP, Sherwin ED, Howard TS, Rosenthal E, Kertesz NJ, Chang PM, Madan N, Kutalek SP, Hammond BH, Janson CM, Ramesh Iyer V, Williams MR. Transcatheter Leadless Pacing in Children: A PACES Collaborative Study in the Real-World Setting. Circ Arrhythm Electrophysiol. 2023;16:e011447. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 29] [Reference Citation Analysis (0)] |
| 41. | Sherwin ED, Shah MJ. Leadless Pacemakers in Patients with Congenital Heart Disease. Card Electrophysiol Clin. 2023;15:421-432. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 11] [Reference Citation Analysis (0)] |
| 42. | Hauser RG, Gornick CC, Abdelhadi RH, Tang CY, Casey SA, Sengupta JD. Major adverse clinical events associated with implantation of a leadless intracardiac pacemaker. Heart Rhythm. 2021;18:1132-1139. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 20] [Cited by in RCA: 62] [Article Influence: 12.4] [Reference Citation Analysis (0)] |
| 43. | Sattar Y, Ullah W, Roomi S, Rauf H, Mukhtar M, Ahmad A, Ali Z, Abedin MS, Alraies MC. Complications of leadless vs conventional (lead) artificial pacemakers - a retrospective review. J Community Hosp Intern Med Perspect. 2020;10:328-333. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 2] [Cited by in RCA: 19] [Article Influence: 3.2] [Reference Citation Analysis (0)] |
| 44. | Palmisano P, Facchin D, Ziacchi M, Nigro G, Nicosia A, Bongiorni MG, Tomasi L, Rossi A, De Filippo P, Sgarito G, Verlato R, Di Silvestro M, Iacopino S. Rate and nature of complications with leadless transcatheter pacemakers compared with transvenous pacemakers: results from an Italian multicentre large population analysis. Europace. 2023;25:112-120. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 26] [Cited by in RCA: 37] [Article Influence: 12.3] [Reference Citation Analysis (0)] |
| 45. | Carretta DM, Tomasi L, Tondo C, De Filippo P, Nigro G, Zucchelli G, Nicosia A, Coppola G, Bontempi L, Sgarito G, Viscusi M, Giannola G, Morani G, Strangio A, Iacopino S. Leadless Micra pacemakers: Estimating long-term longevity. A real word data analysis. Int J Cardiol. 2025;426:133062. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 7] [Cited by in RCA: 8] [Article Influence: 8.0] [Reference Citation Analysis (0)] |
| 46. | Reddy VY, Exner DV, Doshi R, Tomassoni G, Bunch TJ, Friedman P, Estes NAM, Neužil P, de la Concha JF, Cantillon DJ; LEADLESS II Investigators. 1-Year Outcomes of a Leadless Ventricular Pacemaker: The LEADLESS II (Phase 2) Trial. JACC Clin Electrophysiol. 2023;9:1187-1189. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 34] [Reference Citation Analysis (0)] |
| 47. | von Felten E, Breitenstein A, Müller A, Wiederkehr C, Brix T, Jeger R, Hofer D. First European experience with a leadless atrial pacemaker. Heart Rhythm O2. 2025;6:1062-1069. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 5] [Cited by in RCA: 4] [Article Influence: 4.0] [Reference Citation Analysis (0)] |
| 48. | Tjong FVY, Brouwer TF, Koop B, Soltis B, Shuros A, Schmidt B, Swackhamer B, Quast AEB, Wilde AAM, Burke MC, Knops RE. Acute and 3-Month Performance of a Communicating Leadless Antitachycardia Pacemaker and Subcutaneous Implantable Defibrillator. JACC Clin Electrophysiol. 2017;3:1487-1498. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 37] [Cited by in RCA: 57] [Article Influence: 6.3] [Reference Citation Analysis (0)] |
| 49. | Wijesuriya N, Mehta V, Vere F, Howell S, Behar JM, Shute A, Lee M, Bosco P, Niederer SA, Rinaldi CA. Cost-effectiveness analysis of leadless cardiac resynchronization therapy. J Cardiovasc Electrophysiol. 2023;34:2590-2598. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1] [Cited by in RCA: 5] [Article Influence: 1.7] [Reference Citation Analysis (0)] |
| 50. | EBR. WiSE® Cardiac Resynchronization Therapy (CRT) System. 2025. https://www.accessdata.fda.gov/cdrh_docs/pdf24/P240028D.pdf. |
| 51. | Administration UFaD. Premarket Approval Summary: Micra VR. 2016. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=p150033. |
| 52. | Administration UFaD. Premarket Approval Summary: Aveir DR. 2023. https://www.accessdata.fda.gov/cdrh_docs/pdf15/P150035S003B.pdf. |
| 53. | Lloyd MS, Brisben AJ, Reddy VY, Blomström-Lundqvist C, Boersma LVA, Bongiorni MG, Burke MC, Cantillon DJ, Doshi R, Friedman PA, Gras D, Kutalek SP, Neuzil P, Roberts PR, Wright DJ, Appl U, West J, Carter N, Stein KM, Mont L, Knops RE. Design and rationale of the MODULAR ATP global clinical trial: A novel intercommunicative leadless pacing system and the subcutaneous implantable cardioverter-defibrillator. Heart Rhythm O2. 2023;4:448-456. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 26] [Reference Citation Analysis (0)] |
| 54. | Kron J, Bernabei M, Kaiser D, Nanda S, Subzposh FA, Zimmerman P, Rose R, Butler K, Ellenbogen KA. Real-World Performance of Conduction System Pacing Compared With Traditional Pacing. Circ Arrhythm Electrophysiol. 2023;16:411-412. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 14] [Reference Citation Analysis (0)] |
| 55. | Jimenez E, Zaban N, Sharma N, Narasimhan S, Martin-Chafee C, Lohr JL, Cortez D. His Bundle and Left Bundle Pacing in Pediatrics and Congenital Heart Disease: A Single Center Experience. Pediatr Cardiol. 2020;41:1425-1431. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 11] [Cited by in RCA: 22] [Article Influence: 3.7] [Reference Citation Analysis (0)] |
| 56. | Harris TD, Rhee EK, Balcius AD, Papez AL. Comparison of His Bundle and Left Bundle Area Pacing in Pediatric and Adult Congenital Heart Disease Patients: A Single-Center Experience. Pediatr Cardiol. 2026;47:2029-2035. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1] [Cited by in RCA: 2] [Article Influence: 2.0] [Reference Citation Analysis (0)] |
| 57. | Cano Ó, Rosés F, Molina-Lerma M, de Groot NMS, Vijayaraman P, Zemrak F, Beltrán S, BenDavid E, Rivas-Gándara N, Misier NR, Moore JP. Long-term performance of conduction system pacing in patients with congenital heart disease. Heart Rhythm. 2026;23:e2532-e2542. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1] [Cited by in RCA: 1] [Article Influence: 1.0] [Reference Citation Analysis (0)] |
| 58. | O'Connor M, Riad O, Shi R, Hunnybun D, Li W, Jarman JWE, Foran J, Rinaldi CA, Markides V, Gatzoulis MA, Wong T. Left bundle branch area pacing in congenital heart disease. Europace. 2023;25:561-570. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 21] [Reference Citation Analysis (0)] |
| 59. | Hale BW, Devanagondi R, Aldoss O, Olson MD, Law IH. Transvenous endocardial conduction system pacing in patients with a Fontan operation. HeartRhythm Case Rep. 2025;11:1131-1136. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 60. | De Pooter J, Breitenstein A, Özpak E, Haeberlin A, Hofer D, Le Polain de Waroux JB, Wauters A, Kim TH, Lee SR, Park YJ, Gobitz M, Kiełbasa G, Zalavadia D, Tolppanen H, Žižek D, Zanon F, Marcantoni L, Sundaram Ponnusamy S, Karvonen J, Cano O, Jastrzebski M, Vijayaraman P, Burri H. Lead Integrity and Failure Evaluation in Left Bundle Branch Area Pacing: The LIFE-LBBAP Study. JACC Clin Electrophysiol. 2025;11:158-170. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 17] [Cited by in RCA: 13] [Article Influence: 13.0] [Reference Citation Analysis (0)] |
| 61. | Gourraud JB, Chaix MA, Shohoudi A, Pagé P, Dubuc M, Thibault B, Poirier NC, Dore A, Marcotte F, Mongeon FP, Asgar AW, Ibrahim R, Khairy P, Mondésert B. Transvenous Lead Extraction in Adults With Congenital Heart Disease: Insights From a 20-Year Single-Center Experience. Circ Arrhythm Electrophysiol. 2018;11:e005409. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 18] [Cited by in RCA: 23] [Article Influence: 2.9] [Reference Citation Analysis (0)] |
| 62. | Csillik A, Gagyi RB, Kardos A, Földesi C, Som Z, Vamos M, Szili-Torok T. Transvenous Lead Extraction in Patients with Congenital Heart Disease. J Clin Med. 2025;14:4178. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 63. | Fender EA, Killu AM, Cannon BC, Friedman PA, Mcleod CJ, Hodge DO, Broberg CS, Henrikson CA, Cha YM. Lead extraction outcomes in patients with congenital heart disease. Europace. 2017;19:441-446. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 1] [Cited by in RCA: 5] [Article Influence: 0.6] [Reference Citation Analysis (0)] |
| 64. | Pham TDN, Cecchin F, O'Leary E, Fynn-Thompson F, Triedman JK, Gauvreau K, Mah DY. Lead Extraction at a Pediatric/Congenital Heart Disease Center: The Importance of Patient Age at Implant. JACC Clin Electrophysiol. 2022;8:343-353. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 22] [Reference Citation Analysis (0)] |
| 65. | Hachiro Y, Kikuchi S, Ito M, Takahashi K, Abe T. Infection of a retained permanent epicardial pacemaker lead. Ann Thorac Surg. 2001;71:2038-2039. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 5] [Cited by in RCA: 6] [Article Influence: 0.2] [Reference Citation Analysis (0)] |
| 66. | Dulam V, Cooper JM, Gangireddy C, Kashem MA, Toyoda Y, Keshavamurthy S. From leads to leadless: A convoluted journey. HeartRhythm Case Rep. 2020;6:757-760. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1] [Cited by in RCA: 3] [Article Influence: 0.5] [Reference Citation Analysis (0)] |
| 67. | Neuzil P, Exner DV, Knops RE, Cantillon DJ, Defaye P, Banker R, Friedman P, Hubbard C, Delgado SM, Bulusu A, Reddy VY. Worldwide Chronic Retrieval Experience of Helix-Fixation Leadless Cardiac Pacemakers. J Am Coll Cardiol. 2025;85:1111-1120. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 2] [Cited by in RCA: 21] [Article Influence: 21.0] [Reference Citation Analysis (0)] |
| 68. | McNamara GPJ, Haber ZM, Lee EW, Levi DS, Bender A. Successful removal of a leadless pacemaker from the pulmonary artery via a novel basket retrieval system. HeartRhythm Case Rep. 2023;9:215-218. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 5] [Reference Citation Analysis (0)] |
| 69. | Frazer JR, Ing FF. Stenting of stenotic or occluded iliofemoral veins, superior and inferior vena cavae in children with congenital heart disease: acute results and intermediate follow up. Catheter Cardiovasc Interv. 2009;73:181-188. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 29] [Cited by in RCA: 25] [Article Influence: 1.5] [Reference Citation Analysis (0)] |
| 70. | Kamel DW, Abdelhameed AM, Mohammad SA, Abbas SN. CT of cardiac and extracardiac vascular anomalies: embryological implications. Egypt J Radiol Nucl Med. 2021;52:232. [RCA] [DOI] [Full Text] [Cited by in RCA: 3] [Reference Citation Analysis (0)] |
| 71. | Alanís-Naranjo JM, Guerrero-Villalpando AM, De la Cruz-Pelayo J, Basso-Barba G, De Ávila-Gómez MM, Rosas-Vázquez AM, Jiménez-Santos M, Lara-Ameca VJ, García-Montes JA, de la Mora-Cervantes R. Successful Percutaneous Treatment of Stenosed Mustard Atrial Switch Associated With Superior Vena Cava Baffle Thrombi. JACC Case Rep. 2025;30:105413. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 72. | Kumar P, Bhatia M. Computed Tomography in the Evaluation of Fontan Circulation. J Cardiovasc Imaging. 2021;29:108-122. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 9] [Reference Citation Analysis (3)] |
| 73. | Han BK, Rigsby CK, Hlavacek A, Leipsic J, Nicol ED, Siegel MJ, Bardo D, Abbara S, Ghoshhajra B, Lesser JR, Raman S, Crean AM; Society of Cardiovascular Computed Tomography; Society of Pediatric Radiology; North American Society of Cardiac Imaging. Computed Tomography Imaging in Patients with Congenital Heart Disease Part I: Rationale and Utility. An Expert Consensus Document of the Society of Cardiovascular Computed Tomography (SCCT): Endorsed by the Society of Pediatric Radiology (SPR) and the North American Society of Cardiac Imaging (NASCI). J Cardiovasc Comput Tomogr. 2015;9:475-492. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 158] [Cited by in RCA: 133] [Article Influence: 12.1] [Reference Citation Analysis (0)] |
| 74. | Pushparajah K, Duong P, Mathur S, Babu-Narayan S. EDUCATIONAL SERIES IN CONGENITAL HEART DISEASE: Cardiovascular MRI and CT in congenital heart disease. Echo Res Pract. 2019;6:R121-R138. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 19] [Cited by in RCA: 26] [Article Influence: 3.7] [Reference Citation Analysis (0)] |
| 75. | Boersma LV, El-Chami M, Steinwender C, Lambiase P, Murgatroyd F, Mela T, Theuns DAMJ, Khelae SK, Kalil C, Zabala F, Stuehlinger M, Lenarczyk R, Clementy N, Tamirisa KP, Rinaldi CA, Knops R, Lau CP, Crozier I, Boveda S, Defaye P, Deharo JC, Botto GL, Vassilikos V, Oliveira MM, Tse HF, Figueroa J, Stambler BS, Guerra JM, Stiles M, Marques M. Practical considerations, indications, and future perspectives for leadless and extravascular cardiac implantable electronic devices: a position paper by EHRA/HRS/LAHRS/APHRS. Europace. 2022;24:1691-1708. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 72] [Cited by in RCA: 56] [Article Influence: 14.0] [Reference Citation Analysis (0)] |
| 76. | Birnie DH, Healey JS, Wells GA, Ayala-Paredes F, Coutu B, Sumner GL, Becker G, Verma A, Philippon F, Kalfon E, Eikelboom J, Sandhu RK, Nery PB, Lellouche N, Connolly SJ, Sapp J, Essebag V. Continued vs. interrupted direct oral anticoagulants at the time of device surgery, in patients with moderate to high risk of arterial thrombo-embolic events (BRUISE CONTROL-2). Eur Heart J. 2018;39:3973-3979. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 98] [Cited by in RCA: 155] [Article Influence: 22.1] [Reference Citation Analysis (0)] |
| 77. | Calvert P, Yeo C, Rao A, Neequaye S, Mayhew D, Ashrafi R. Transcarotid implantation of a leadless pacemaker in a patient with Fontan circulation. HeartRhythm Case Rep. 2023;9:53-58. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 6] [Reference Citation Analysis (0)] |
| 78. | Crossley GH. PO-03-043 Trans-Baffle Insertion of a Leadless Pacemaker in a Patient with Extracardiac Fontan. Heart Rhythm. 2024;21:S359. [DOI] [Full Text] |
| 79. | Cortez D. First Report on Atrial Leadless Pacing in an Extra-Cardiac Fontan Patient. Pacing Clin Electrophysiol. 2026;49:466-470. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 80. | Goulden CJ, Khanra D, Llewellyn J, Rao A, Evans A, Ashrafi R. Novel approaches for leadless pacemaker implantation in the extra-cardiac Fontan cohort: Options to avoid leaded systems or epicardial pacing. J Cardiovasc Electrophysiol. 2023;34:2386-2392. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 11] [Reference Citation Analysis (0)] |
| 81. | Moore JP, Shannon KM. Transpulmonary atrial pacing: an approach to transvenous pacemaker implantation after extracardiac conduit Fontan surgery. J Cardiovasc Electrophysiol. 2014;25:1028-1031. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 22] [Cited by in RCA: 33] [Article Influence: 2.8] [Reference Citation Analysis (0)] |
| 82. | O'Connor M, Ho SY, McCarthy KP, Gatzoulis M, Wong T. Left bundle pacing in transposition of the great arteries with previous atrial redirection operation. HeartRhythm Case Rep. 2022;8:176-179. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 1] [Cited by in RCA: 16] [Article Influence: 4.0] [Reference Citation Analysis (0)] |
| 83. | Alasti M, Kotschet E, Alison J. Micra Transcatheter Pacing System implantation in a Patient with a History of Mustard Repair and Transposition of Great Arteries. Heart Lung Circ. 2018;27 Suppl 2:S178. [RCA] [DOI] [Full Text] [Cited by in Crossref: 3] [Cited by in RCA: 4] [Article Influence: 0.5] [Reference Citation Analysis (0)] |
| 84. | Singh HR, Batra AS, Balaji S. Pacing in children. Ann Pediatr Cardiol. 2013;6:46-51. [RCA] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 20] [Cited by in RCA: 24] [Article Influence: 1.8] [Reference Citation Analysis (0)] |
| 85. | Reyes JL, Feuerborn M, Shabbir A, Prisco A, Kalra R, Von Wald L, Benditt DG, Roukoz H. Transvenous Coronary Sinus Lead Placement in Systemic Right Ventricle With D-TGA and Atrial Baffle. JACC Case Rep. 2026;31:106229. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 86. | Chakrabarti S, Szantho G, Turner MS, Stuart G, Martin RP. Use of radiofrequency perforation for lead placement in biventricular or conventional endocardial pacing after mustard or senning operations for D-transposition of the great arteries. Pacing Clin Electrophysiol. 2009;32:1123-1129. [RCA] [PubMed] [DOI] [Full Text] [Cited by in Crossref: 6] [Cited by in RCA: 9] [Article Influence: 0.5] [Reference Citation Analysis (0)] |
| 87. | Smith NR, Parekh DR, Garcia-Sayan E, Dwarakanath K, Lam WW. Transbaffle, Endocardial Systemic Right Ventricle Cardiac Resynchronization Therapy in D-Transposition of the Great Arteries. JACC Case Rep. 2025;30:104936. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 1] [Reference Citation Analysis (0)] |
| 88. | Mitchell AJ, Murphy N, Walsh KP. Fist in man implantation of a leadless pacemaker in the left atrial appendage following Mustard repair. Europace. 2022;24:795. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 9] [Reference Citation Analysis (0)] |
| 89. | Tanawuttiwat T, Vasilottos N, Borman WA, Giro P, Das MK. A dual-chamber leadless pacemaker in d-TGA patient after senning procedure. HeartRhythm Case Rep. 2025;11:61-65. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in RCA: 3] [Reference Citation Analysis (0)] |
| 90. | Sathananthan G, Lowe BS, Hornung T. Editor’s Pick: The Fontan Circulation. EMJ Cardiol. 2016;4:76-83. [DOI] [Full Text] |
| 91. | Umamaheshwar KL, Singh AS, Sivakumar K. Endocardial transvenous pacing in patients with surgically palliated univentricular hearts: A review on different techniques, problems and management. Indian Pacing Electrophysiol J. 2019;19:15-22. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 5] [Cited by in RCA: 14] [Article Influence: 1.8] [Reference Citation Analysis (0)] |
| 92. | Pinsker BL, Moore JP, Bashore TM, Krasuski RA. Permanent Cardiac Pacing in the Fontan Population: A Contemporary Review of Indications, Approaches, and Outcomes. JACC Adv. 2025;4:101667. [RCA] [PubMed] [DOI] [Full Text] [Cited by in RCA: 6] [Reference Citation Analysis (0)] |
| 93. | Misra A, Puri K, Hope K, Broda C, Lam WW. Single Atrial Lead Pacing from Left Pulmonary Artery in Lateral Tunnel Fontan. J Am Coll Cardiol. 2025;85 Suppl 12:3068. [DOI] [Full Text] |
| 94. | Hudziak D, Hajder A, Gocol R, Malinowski M, Kazmierski M, Morkisz L, Ciosek J, Wanha W, Jarosinski G, Parma R, Darocha T, Wojakowski W. Long-Term Clinical Outcomes and Carotid Ultrasound Follow-Up of Transcarotid TAVI. Prospective Single-Center Registry. J Clin Med. 2021;10:1499. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 2] [Cited by in RCA: 7] [Article Influence: 1.4] [Reference Citation Analysis (0)] |
| 95. | Gellér L, Salló Z, Molnár L, Tahin T, Özcan EE, Kutyifa V, Osztheimer I, Szilágyi S, Szegedi N, Ábrahám P, Apor A, Nagy KV, Kosztin A, Becker D, Herczeg S, Zima E, Merkely B. Long-term single-centre large volume experience with transseptal endocardial left ventricular lead implantation. Europace. 2019;21:1237-1245. [RCA] [PubMed] [DOI] [Full Text] [Full Text (PDF)] [Cited by in Crossref: 5] [Cited by in RCA: 15] [Article Influence: 2.5] [Reference Citation Analysis (0)] |