World J Gastrointest Endosc. Feb 10, 2016; 8(3): 143-156
Published online Feb 10, 2016. doi: 10.4253/wjge.v8.i3.143
Published online Feb 10, 2016. doi: 10.4253/wjge.v8.i3.143
Table 1 Technical characteristics of the most commonly used biliary plastic stents
Producer | Model | Diameter (Fr) | Length (cm) | Shape | Material |
Boston Scientific | Advanix | 7, 8.5, 10 | 5-18 | Duodenal bend, centre bend, double pigtail | Polyethylene |
ConMed | Hydroduct | 7, 10, 12 | 4-15 | Straight, angled, curved, double pigtail | Polyurethane with hydrophilic hydromer coating |
Cook Endoscopy | Compass BDS | 7 | 5, 10, 15 | Double pigtail | Polyethylene |
Cook Endoscopy | Cotton-Huibregtse | 7, 8.5, 10, 11.5 | 5-18 | Angled | Polyethylene |
Cook Endoscopy | Cotton-Leung | 7, 8.5, 10, 11.5 | 5-18 | Curved | Polyethylene |
Cook Endoscopy | Cotton-Leung Sof-Flex | 7, 10 | 5-15 | Curved | Polyethylene and polyurethane blend |
Cook Endoscopy | Fusion Marathon Antireflux | 10 | 5-12 | Curved | Polyethylene with teflon sleeve |
Cook Endoscopy | Soehendra-Tannenbaum | 8.5, 10, 11.5 | 5-15 | Curved | Teflon |
Cook Endoscopy | Solus | 10 | 1-15 | Double pigtail | Polyethylene and polyurethane blend |
Cook Endoscopy | Zimmon | 5, 6, 7, 10 | 4, 7, 10 | Double pigtail | Polyethylene |
Endo-Flex | PE-Soft | 7, 8.5, 10, 11.5 | 3-15 | Bended, straight, curved, double pigtail, single pigtail | Polyethylene |
Endo-Flex | PTFE-Strong | 7, 8.5, 10, 11.5 | 5-15 | Bended, straight, curved | Polytetrafluoroethylene |
GI Supply | ViaDuct | 7, 10 | 5-15 | Winged straight | Polyurethane |
Hobbs Medical | Biliary stent | 7, 10 | 4-15 | Curved, Double pigtail | Soft polymer blend |
Indus Medical | CIBIDI | 7, 10 | 5-15 | Straight, curved, double pigtail | Polyurethane and teflon |
Olympus | Double Layer | 10 | 4-15 | Duodenal bend, centre bend | Inner layer: Perfluoro; middle layer: Stainless steel; outer layer: Polyamide elastomer |
Olympus | Biliary EVA | 7, 8.5, 10, 12 | 5-18 | Straight, proximal bend, centre bend, double pigtail | Ethylene vinyl acetate copolymers |
Olympus | Biliary FEP | 7, 8.5, 10, 12 | 3-15 | Straight, proximal bend | Fluorinated ethylene propylene |
Olympus | Biliary PE | 7, 8.5, 10, 12 | 3-15 | Straight, proximal bend, centre bend, double pigtail | Polyethylene |
Pauldrach Medical | Gallengangs | 7, 8.4, 10 | 9 | Curved | Polyethylene |
Table 2 Technical characteristics of the most commonly used uncovered self-expandable metal stents
Producer | Model | Material | Diameter (mm) | Length (cm) | Shortening | Reconstrain | Characteristics |
Boston Scientific | Wallflex® | Platinol | 8, 10 | 4, 6, 8, 10 | Yes | Yes | --------- |
ConMed | Flexxus | Nitinol | 8, 10 | 4, 6, 8, 10 | Yes | No | Pistol delivery system |
Cook Endoscopy | Zilver® | Nitinol | 6, 8, 10 | 4, 6, 8 | No | No | --------- |
Cook Endoscopy | Evolution® | Nitinol | 8, 10 | 4, 6, 8, 10 | Yes | Yes | Pistol delivery system |
Ella-CS | SX-ELLA® Nitinella Plus | Nitinol | 8, 10 | 4, 6, 8, 10 | Yes | Yes | --------- |
Endochoice | Bonastent® | Nitinol | 8, 10 | 4, 5, 6, 8, 10, 12 | Yes | Yes | --------- |
Endo-Flex | BIL-stent | Nitinol | 10 | 6, 8, 10 | Yes | No | --------- |
Endo-Technik | NIT-BIL-1010® | Nitinol | 10 | 4, 6, 8, 10 | Yes | Yes | --------- |
Leufen Medical | Aixstent® Gallengang | Nitinol | 8, 10 | 4, 6, 8 | Yes | No | --------- |
Leufen Medical | Aixstent® Gallengang BDL - BDH | Nitinol | 8, 10 | 4, 6, 8, 10, 12 | Yes | No | The open cell design allows for Y stenting at the hilar region |
Merit Endotek | Alimaxx-B® | Nitinol | 8, 10 | 4, 6, 8 | Yes | No | The open cell design allows for Y stenting at the hilar region |
M.I. Tech | Hanarostent® | Nitinol | 8, 10 | 4, 5, 6, 7, 8, 9, 10, 12 | Yes | Yes | --------- |
M.I. Tech | Hanarostent® Hilar | Nitinol | 10 | 8 | Yes | No | The large cell design allows for Y stenting at the hilar region |
Micro-Tech | BD stents Classic or Platinum-Line | Nitinol | 10 | 4, 6, 8, 10 | Yes | No | --------- |
Olympus | NIRflex | Nitinol | 8, 10 | 4, 6, 8, 10 | Yes | No | --------- |
S and G Biotech | EGIS® Biliary DC Stent | Nitinol | 8, 10, 12 | 4, 5, 6, 7, 8, 9, 10, 12 | Yes | No | Single or double bare |
TaeWoong Medical | LCD® | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8, 9, 10, 12 | Yes | No | The large cell design allows for Y stenting at the hilar region |
TaeWoong Medical | Niti-S® D-type | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8, 9, 10, 12 | Yes | No | --------- |
TaeWoong Medical | Niti-S® S-type | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8, 9, 10, 12 | Yes | No | --------- |
Table 3 Technical characteristics of the most commonly used partially and fully-covered self-expandable metal stents
Producer | Model | Material | Diameter (mm) | Length (cm) | Shortening | Reconstrain | Shape | Covering |
Allium Medical | Allium BIS® | Nitinol | 8, 10 | 6, 8, 10, 12 | No | No | Straight with anchoring segment | FC in polyurethane |
Boston Scientific | Wallflex® | Platinol | 8, 10 | 4, 6, 8, 10 | Yes | Yes | Two flanges | PC and FC in permalume |
Cook Endoscopy | Evolution® | Nitinol | 8, 10 | 4, 6, 8, 10 | Yes | Yes | Two flanges | PC and FC in silicone |
Ella-CS | SX-ELLA® Nitinella Plus | Nitinol | 8, 10 | 4, 6, 8, 10 | Yes | Yes | Two flanges | PC and FC in silicone |
Endochoice | Bonastent® | Nitinol | 8, 10 | 4, 5, 6, 8, 10, 12 | Yes | Yes | Two flanges | FC in silicone |
Endo-Flex | BIL-stent | Nitinol | 10 | 6, 8 | Yes | No | Straight | FC in silicone |
Endo-Technik | NIT-BIL-1010® | Nitinol | 10 | 4, 6, 8, 10 | Yes | No | Straight | PC in silicone |
Gore Medical | Viabil® | Nitinol | 8, 10 | 4, 6, 8, 10 | No | No | Straightwith anchoring fins | FC in PTFE with/without drainage holes |
Leufen Medical | Aixstent® Gallengang | Nitinol | 8, 10 | 4, 6, 8 | Yes | No | Two flanges | PC and FC in polyurethane |
M.I. Tech | Hanarostent® BCT | Nitinol | 10 | 4, 6, 8, 10 | Yes | Yes | One flange with flaps and lasso | FC in silicone |
M.I. Tech | Hanarostent® BCS | Nitinol | 10 | 4, 6, 8, 10, 12 | Yes | No | One flange and with flaps | FC in silicone |
M.I. Tech | Hanarostent® BPE | Nitinol | 8, 10 | 8, 10 | Yes | No | One flange and with flaps | PC in silicone |
Micro-Tech | BD stents | Nitinol | 10 | 4, 6, 8, 10 | Yes | No | Two flanges | PC and FC in silicone |
S and G Biotech | EGIS® Biliary DC Stent | Nitinol | 8, 10, 12 | 4, 5, 6, 7, 8, 9, 10, 12 | Yes | No | Two flanges | PC in PTFE |
TaeWoong Medical | Niti-S® S-type covered | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8, 10, 12 | Yes | No | Two flanges | FC in silicone |
TaeWoong Medical | Niti-S® Kaffes | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8 | Yes | No | Tapered with long lasso | FC in silicone |
TaeWoong Medical | Niti-S® Bumpy | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8, 10, 12 | Yes | No | Two flanges | FC in silicone and PTFE |
TaeWoong Medical | Niti-S® Giobor | Nitinol | 8, 10 | 8, 10 | Yes | No | One flange | PC in silicone |
TaeWoong Medical | Niti-S® ComVi | Nitinol | 6, 8, 10 | 4, 5, 6, 7, 8, 10, 12 | Yes | No | Straight | FC in PTFE |
Table 4 Technical characteristics of the most commonly used pancreatic plastic stents
Producer | Model | Diameter (Fr) | Length (cm) | Shape | Material |
Boston scientific | Advanix | 3, 4, 5, 7, 10 | 2-18 | Straight or single pigtail with or without internal flap | Polyethylene |
Cook endoscopy | Geenan Sof-Flex | 5 | 3-12 | Curved with or without internal flap | Polyethylene and polyurethane blend |
Cook endoscopy | Geenan | 3, 5, 7 | 3-15 | Curved | Polyethylene |
Cook endoscopy | Johlin Wedge | 8.5, 10 | 8-22 | Wedge | Polyethylene and polyurethane blend |
Cook endoscopy | Zimmon | 3, 5, 7 | 2-12 | Single pigtail with or without internal flap | Polyethylene |
Endo-Flex | PTFE-Strong | 5, 7 | 3-9 | Curved | Polytetrafluoroethylene |
GI supply | ViaDuct | 5, 7 | 3-12 | Winged straight or single pigtail with or without internal flap | Polyurethane |
Hobbs medical | Freeman Flexi-Stents | 3, 4, 5, 7 | 2-18 | Straight or single pigtail with or without internal flap | Soft polymer |
Olympus | Pancreatic PE | 7, 8.5, 10 | 3-15 | Straight, S-shaped | Polyethylene |
Table 5 Technical characteristics of the lumen apposing metal stents
Producer | Model | Internal diameter (mm) | Length (mm) | Flange diameter (mm) |
Boston Scientific | Axios | 10, 15 | 10 | 21, 24 |
Leufen Medical | Aix | 10, 14 | 20 | 14/16, 18/20 |
M.I. Tech | Hanarostent BCF | 10, 12 | 30, 40 | 25 |
TaeWoong Medical | Spaxus | 8, 10, 16 | 20 | 25 |
TaeWoong Medical | Nagi | 10, 12, 14, 16 | 10, 20, 30 | 22, 24, 26, 28 |
- Citation: Mangiavillano B, Pagano N, Baron TH, Arena M, Iabichino G, Consolo P, Opocher E, Luigiano C. Biliary and pancreatic stenting: Devices and insertion techniques in therapeutic endoscopic retrograde cholangiopancreatography and endoscopic ultrasonography. World J Gastrointest Endosc 2016; 8(3): 143-156