Effectsetting | Power output (W) | Shot duration (s) | Shot location | Depth of tissue injury (μm) | Observations |
4 | 20 | 90 | Bile duct | 2700 | Transmural necrosis |
6 | 10 | 90 | Bile duct | 2600 | Transmural necrosis |
8 | 10 | 90 | Ampulla | 2810 | Extensive necrosis of the papilla, involving pancreas and adipose tissue |
8 | 10 | 30 | Ampulla | 1530 | Deep coagulation, necrosis of the duodenal wall. No significant pancreatic injury |
8 | 6 | 60 | Bile duct | 1420 | Deep transmural coagulation, involving peribiliary adipose tissue |
8 | 8 | 30 | Bile duct | 1075 ± 4171 | Coagulation involving almost the entire duct wall |
8 | 6 | 30 | Ampulla | 350 | Superficial coagulation, moderate duodenal lesion |
- Citation: Barret M, Leblanc S, Vienne A, Rouquette A, Beuvon F, Chaussade S, Prat F. Optimization of the generator settings for endobiliary radiofrequency ablation. World J Gastrointest Endosc 2015; 7(16): 1222-1229