Copyright
©The Author(s) 2023.
World J Gastrointest Endosc. Dec 16, 2023; 15(12): 705-714
Published online Dec 16, 2023. doi: 10.4253/wjge.v15.i12.705
Published online Dec 16, 2023. doi: 10.4253/wjge.v15.i12.705
Test method | Endoscopes | Postuse | Postprecleaning | Postmanual cleaning |
ATP test (RLU) | Gastroscopes (n = 60) | 5313645 ± 3919731 | 120035 ± 214287a | 1567 ± 1152b |
Colonoscopes (n = 48) | 1235303 ± 1182027 | 31293 ± 78392a | 1450 ± 1945c | |
Residual protein test (μg) | Gastroscopes (n = 88) | 102.39 ± 58.09 | 39.08 ± 19.60a | 32.04 ± 15.63c |
Colonoscopes (n = 32) | 121.87 ± 68.66 | 42.80 ± 18.53a | 30.34 ± 10.42c |
Groups | Samples | ATP test (RLU) | Samples | Residual protein test (μg) | |
T1 | 24 | 2141 ± 1913 | 30 | 31.27 ± 12.81 | |
Ts | T2 | 10 | 2362 ± 1981 | 10 | 31.44 ± 14.19 |
T3 | 10 | 2340 ± 2017 | 10 | 29.28 ± 8.46 | |
T4 | 10 | 1795 ± 1406 | 10 | 33.67 ± 19.75 | |
Total | 30 | 2166 ± 1779 | 30 | 31.46 ± 14.46 | |
D1 | 24 | 756 ± 834 | 30 | 30.51 ± 17.95 | |
Ds | D2 | 10 | 1083 ± 854 | 10 | 28.80 ± 10.64 |
D3 | 10 | 913 ± 730 | 10 | 34.33 ± 12.96 | |
D4 | 10 | 918 ± 956 | 10 | 36.18 ± 13.28 | |
Total | 30 | 971 ± 826 | 30 | 34.66 ± 15.50 |
- Citation: Wang YF, Wu Y, Liu XW, Li JG, Zhan YQ, Liu B, Fan WL, Peng ZH, Xiao JT, Li BB, He J, Yi J, Lu ZX. Effect of a disposable endoscope precleaning kit in the cleaning procedure of gastrointestinal endoscope: A multi-center observational study. World J Gastrointest Endosc 2023; 15(12): 705-714
- URL: https://www.wjgnet.com/1948-5190/full/v15/i12/705.htm
- DOI: https://dx.doi.org/10.4253/wjge.v15.i12.705