Published online Jan 16, 2024. doi: 10.4253/wjge.v16.i1.37
Peer-review started: October 9, 2023
First decision: November 16, 2023
Revised: November 22, 2023
Accepted: December 5, 2023
Article in press: December 5, 2023
Published online: January 16, 2024
Processing time: 97 Days and 10.4 Hours
Gastrointestinal (GI) bleeding accounts for over half a million admissions annually and is the most common GI diagnosis requiring hospitalization in the United States. Bipolar electrocoagulation devices are used for the management of gastrointestinal bleeding. There is no data on device-related adverse events for gold probe (GP) and injection gold probe (IGP).
To analyze this using the Food and Drug Administration (FDA’s) Manufacturer and User Facility Device Experience (MAUDE) database from 2013 to 2023.
We examined post-marketing surveillance data on GP and IGP from the FDA MAUDE database to report devicerelated and patient-related adverse events between 2013-2023. The MAUDE database is a publicly available resource providing over 4 million records relating to medical device safety. Statistical analyses were performed using IBM SPSS Statistics V.27.0 (IBM Corp., Armonk, NY, United States).
Our search elicited 140 reports for GP and 202 reports for IGP, respec
GP and IGP are critical in managing gastrointestinal bleeding. This study of the FDA MAUDE database revealed the type, number, and trends of reported device-related adverse events. The endoscopist and support staff must be aware of these device-related events and be equipped to manage them if they occur.
Core Tip: Gold probe (GP) and injection gold probe (IGP) are critical in managing gastrointestinal bleeding. This study of the Food and Drug Administration Manufacturer and User Facility Device Experience database revealed the type, number, and trends of reported device-related adverse events. Our search elicited 140 reports for GP and 202 reports for IGP, respec
- Citation: Kumar VCS, Aloysius M, Aswath G. Adverse events associated with the gold probe and the injection gold probe devices used for endoscopic hemostasis: A MAUDE database analysis. World J Gastrointest Endosc 2024; 16(1): 37-43
- URL: https://www.wjgnet.com/1948-5190/full/v16/i1/37.htm
- DOI: https://dx.doi.org/10.4253/wjge.v16.i1.37
Gastrointestinal (GI) bleeding accounts for over half a million admissions annually and is the most common GI diagnosis requiring hospitalization in the United States[1]. Lesions with high-risk stigmata, which are associated with high rates of recurrent bleeding (50% to 80%) and result in significant morbidity if treated with medical therapy alone. Thus, the latest American College of Gastroenterology (ACG) guidelines recommend endoscopic therapy for ulcers with active spurting or oozing and nonbleeding visible vessels. The management of nonvariceal upper GI bleed (UGIB) has evolved tremendously with the advent of therapeutic endoscopic hemostasis devices and techniques. Studies have shown that thermal contact devices such as bipolar electrocoagulation and heater probes decrease the incidence of re-bleeding compared with no endoscopic therapy[2].
Overall, devices used to achieve hemostasis using thermal therapy were safe. The serious adverse events associated with these devices include uncontrollable bleeding and perforation[3]. Pooled data showed that the rate of bleeding that required urgent surgery was 0.3%, and perforation was 0.5%[4].
The gold probe (GP) and injection gold probe (IGP) (Boston Scientific Corp., Natick, Mass.) are two commonly used devices to achieve endoscopic hemostasis. IGP can deliver an injection as well as thermal therapy. No data on device-related adverse events for these devices used routinely to achieve endoscopic hemostasis is available.
Thus, we aimed to evaluate the events associated with using Gold Probe and Injection Gold Probe using the Food and Drug Administration (FDA’s) Manufacturer and User Facility Device Experience (MAUDE) database from 2013 to 2023.
We examined post-marketing surveillance data on GP and IGP from the FDA MAUDE database to report devicerelated and patient-related adverse events. The MAUDE database is a publicly available resource providing over 4 million re
It consists of four primary (Master Event, Device, Patient, Text) and two supplemental (Device Problems and Problem Code Descriptions) file types, which, when combined, provide a detailed account of an adverse event or product problem report. Healthcare professionals have used MAUDE to review events associated with specific products or procedures. Several articles referencing MAUDE have been published analyzing adverse events specific to a particular outcome, product, or body system. It is publicly available online and de-identified. Therefore, no institutional review board approval was required for this study.
We queried the MAUDE database from January 2013 to August 2023. The MAUDE web search feature is limited to adverse event reports within the past ten years. The data was analyzed for device issues and patient adverse events. The primary outcome measure of this study was the failure modes of the endoscopic diathermy Gold ProbeTM (Ò Boston Scientific) and injection diathermy Injection Gold ProbeTM (Ò Boston Scientific). Secondary outcomes included significant complications associated with device failure. The MAUDE database cannot capture the utilization of IGP in the United States; therefore, the actual incidence rate of each failure or complication type cannot be assessed. Categorical variables were presented as numbers; all statistical analyses were performed using IBM SPSS Statistics V.27.0 (IBM Corp., Armonk, NY, United States).
Our search elicited 140 reports for GP and 202 reports for IGP, respectively, during the study period from January 2013 to August 2023. The procedure type for GP use was esophagogastroduodenoscopy (47) followed by colonoscopy (25), bronchoscopy (7), endoscopic retrograde cholangiopancreatography (ERCP) (6), enteroscopy (3), missing procedure information (52), Table 1. The procedure types for IGP were esophagogastroduodenoscopy (174) followed by colonoscopy (16), ERCP (11), and enteroscopy (1), Table 2.
Procedure | Number |
Bronchoscopy | 7 |
Colonoscopy | 25 |
Esophagogastroduodenoscopy | 47 |
Enteroscopy | 3 |
Endoscopic retrograde cholangiopancreatography | 6 |
Missing | 52 |
Total | 140 |
Procedure | Number |
EGD | 174 |
Colonoscopy | 16 |
ERCP | 11 |
Enteroscopy | 1 |
Total | 202 |
Primary outcomes outlining failure modes for the GP and IGP is outlined in Tables 3 and 4. GP failure modes were failure to deliver energy (107), followed by material separation or fracture of the probe tip (28), arcing (1), missing component (1), bent tip (1), and detachment of device (2). IGP failure modes were failure to deliver energy (115), followed by material separation or fracture of the probe tip (34), crack (9), device detachment (27), material puncture (5), and mechanical problems (12).
Failure mode | Number | Percentage |
Arcing | 1 | 0.7 |
Component missing | 1 | 0.7 |
Detachment of device component | 1 | 0.7 |
Electrical connector broke during use | 1 | 0.7 |
Failure to deliver energy | 107 | 76.4 |
Material separation | 28 | 20.0 |
Tip bent (from packaging) | 1 | 0.7 |
Total | 140 | 100.0 |
Failure mode | Number | Percentage |
Crack | 9 | 4.5 |
Material separation | 34 | 16.8 |
Device detachment | 27 | 13.4 |
Failure to deliver energy | 115 | 56.9 |
Material puncture/hole | 5 | 2.5 |
Mechanical problem | 12 | 5.9 |
Total | 202 | 100.0 |
Malfunctions reportedly occurred in 130 cases for GP, and actual patient injury or event occurred in 10 patients. In assessing secondary outcomes, no deaths were reported, although two patients experienced prolonged hemorrhage and two fiberoptic endoscopes were damaged by the device; 7 patients required a secondary procedure to retrieve the detached probe. Most patients with a reported GP event suffered no significant consequences due to the device failure (93%), but 7% required a second procedure or experienced prolonged stay or discomfort, Table 5. Most patients reported with IGP events (74%) suffered no significant consequences due to the device failure, but 26% of patients were affected by an event (prolonged hemorrhage, need for a secondary procedure due to a detached probe), Table 6. Reports by year decreased significantly after 2017 for both GP and IGP, Table 7.
Events | Number |
Bowel burn | 1 |
Hemorrhage/Bleeding | 1 |
Prolonged hospital stay | 1 |
Removal requiring a second procedure | 7 |
Total | 10 |
Events | Number |
Prolonged hemorrhage | 2 |
The secondary procedure to retrieve the detached probe | 48 |
Probe damaged scope | 3 |
Total | 53 |
Year | MAUDE reports for gold probe | MAUDE reports for injection gold probe |
2013 | 32 | 38 |
2014 | 32 | 36 |
2015 | 17 | 33 |
2016 | 26 | 30 |
2017 | 14 | 23 |
2018 | 3 | 6 |
2019 | 3 | 9 |
2020 | 2 | 8 |
2021 | 4 | 7 |
2022 | 1 | 5 |
2023 | 6 | 7 |
Total | 140 | 202 |
Our study comprehensively analyzes events reported with GP and IGP from 2013 to 2023. For both GP and IGP, the most reported problem is the "failure to deliver energy." Investigating the root cause of this recurrent issue with these devices is imperative. If user error is identified as a significant factor, offering additional training to the healthcare professionals using these devices and refining the user guidelines would be beneficial.
The significantly higher number of reported events with IGP devices than with GP devices is noteworthy. While a higher usage frequency might contribute to the increased reporting, the pronounced rate of patient-related adverse events stemming from IGP failures cannot be dismissed lightly. Especially concerning are instances requiring repeat procedures, as they amplify the risk profile for patients and accentuate the resource burdens on healthcare institutions.
The manufacturer for the GP and IGP reports patient-related adverse events, including perforation, bleeding, aspiration pneumonia, and septicemia/infection, and reports a potential electrical hazard to the patient and operator with possible adverse including fulguration, burns, stimulation, and cardiac arrhythmia[5]. However, there have been no studies so far that have looked at the device-related events that could occur with GP and IGP. Our study is the first to analyze the device-related events reported. It sheds light on device-related complications, thus enhancing the existing knowledge pool crucial for daily clinical applications. Data regarding other bipolar devices was sparse and thus a comparative analysis could not be done.
The 2021 ACG guidelines for managing UGIB strongly recommend endoscopic hemostatic therapy with bipolar electrocoagulation, heater probe, or injection of absolute ethanol for patients with UGIB due to ulcers. Several studies have proven the efficacy and overall safety of GP and IGP to manage gastrointestinal hemorrhage[2,6,7]. The safety and efficacy of bipolar devices have been also established while managing lower GI bleeding[8,9]. GP and IGP are Bipolar devices used to manage GI bleeding during endoscopy. Given the ubiquity of these bipolar devices in clinical scenarios, endoscopists, and auxiliary staff must be apprised of potential device-related pitfalls.
Interestingly, the findings of this study also suggest that there was a decline in the events for both IGP and GP from 2017. Endoscopists familiarity with the device and adequate training in its usage, and manufacturer’s improvement of the quality of the device could have led to fewer events. Usage of other hemostatic devices could have also contributed to this. Over-the-Scope Clips (OTSC) has been shown to be as effective as standard therapy in non-variceal upper gastrointestinal bleeding since 2017[10]. OTSC has also proven effective in large ulcers up to 5 cm[11], with a high success rate of hemostasis (80%) even in recurrent bleeding and has also competed with GP and IGP as first line hemostatic method since 2017[12].
At around the same time, hemostatic aerosolized powders such as TC 325 (Hemospray) have become part of the hemostatic armamentarium available to the endoscopist, especially effective in the setting of diffuse mucosal bleeding[13,15].
These newer hemostatic technologies may have contributed to a decline in use of IGP and GP since 2017. It’s also conceivable that the manufacturing process may have effectively addressed the prior device failure reports to redesign and improve quality control hence leading to a decline in the device malfunction/failure reports since 2017.
Guidelines for non-variceal upper gastrointestinal bleeding have emphasized that epinephrine injection needs to be combined with a secondary hemostatic modality and hence IGP use may have increased over GP use. IGP conveniently uses both injection and thermocoagulation sequentially without interruption to introduce another hemostatic method endoscopically. This may have contributed to increase in IGP use over GP and consequently higher device malfunction reports[15].
This study has limitations. The MAUDE web search feature is limited to adverse event reports within the past ten years. This passive surveillance system has its limitations. There is a potential for submission of incomplete, inaccurate, untimely, unverified, or biased data. In addition, the incidence or prevalence of an event cannot be determined from this reporting system alone due to under-reporting of events, inaccuracies in reports, lack of verification that the device caused the reported event, and lack of information about the frequency of the device use.
GP and IGP are critical in managing gastrointestinal bleeding. This study of the FDA MAUDE database revealed the type, number, and trends of reported device-related adverse events. The endoscopist and support staff must be aware of these device-related events and be equipped to manage them if they occur.
Gold probe (GP) and gold probe (GP) are vital in managing gastrointestinal bleeding, yet they present notable risks. Awareness of these risks is essential for endoscopists and support staff. The study highlights the need for improved device safety and better management strategies in case of device failure.
The analysis revealed 140 reports for GP and 202 reports for IGP, with the majority of device failures being attributed to the failure to deliver energy. While most events did not lead to significant patient consequences, a notable proportion (26% for IGP) resulted in adverse outcomes like prolonged hemorrhage or the need for secondary procedures.
The study utilized post-marketing surveillance data from the Food and Drug Administration (FDA’s) Manufacturer and User Facility Device Experience (MAUDE) database, analyzing reports for GP and IGP from January 2013 to August 2023. Statistical analyses were performed using IBM SPSS Statistics V.27.0 to identify primary and secondary outcome measures.
The primary objective is to evaluate the events associated with the use of GP and IGP, specifically focusing on the types and frequencies of device failures and their impact on patient outcomes.
The motivation for this research stems from the lack of comprehensive data on device-related adverse events for GP and IGP, devices commonly used in managing gastrointestinal bleeding, despite their widespread clinical use.
This study investigates the device-related adverse events associated with the use of GP and IGP in endoscopic hemostasis, leveraging data from the FDA's MAUDE database over a decade (2013-2023).
The findings underscore the need for ongoing surveillance, device improvement, and consideration of emerging hemostatic technologies. Further research into device design and usage guidelines could enhance safety and efficacy in clinical practice.
Provenance and peer review: Unsolicited article; Externally peer reviewed.
Peer-review model: Single blind
Corresponding Author's Membership in Professional Societies: American College of Gastroenterology; American Society for Gastrointestinal Endoscopy; American Gastroenterological Association.
Specialty type: Gastroenterology and hepatology
Country/Territory of origin: United States
Peer-review report’s scientific quality classification
Grade A (Excellent): 0
Grade B (Very good): 0
Grade C (Good): C
Grade D (Fair): D
Grade E (Poor): 0
P-Reviewer: Qi XS, China; Wang S, China S-Editor: Liu JH L-Editor: A P-Editor: Cai YX
1. | Peery AF, Crockett SD, Murphy CC, Lund JL, Dellon ES, Williams JL, Jensen ET, Shaheen NJ, Barritt AS, Lieber SR, Kochar B, Barnes EL, Fan YC, Pate V, Galanko J, Baron TH, Sandler RS. Burden and Cost of Gastrointestinal, Liver, and Pancreatic Diseases in the United States: Update 2018. Gastroenterology. 2019;156:254-272.e11. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 776] [Cited by in F6Publishing: 973] [Article Influence: 194.6] [Reference Citation Analysis (0)] |
2. | Laine L, McQuaid KR. Endoscopic therapy for bleeding ulcers: an evidence-based approach based on meta-analyses of randomized controlled trials. Clin Gastroenterol Hepatol. 2009;7:33-47; quiz 1. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 235] [Cited by in F6Publishing: 216] [Article Influence: 14.4] [Reference Citation Analysis (38)] |
3. | Waterman M, Gralnek IM. Preventing complications of endoscopic hemostasis in acute upper gastrointestinal hemorrhage. Gastrointest Endosc Clin N Am. 2007;17:157-167, viii. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 4] [Cited by in F6Publishing: 4] [Article Influence: 0.2] [Reference Citation Analysis (0)] |
4. | Laine L, Peterson WL. Bleeding peptic ulcer. N Engl J Med. 1994;331:717-727. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 498] [Cited by in F6Publishing: 430] [Article Influence: 14.3] [Reference Citation Analysis (0)] |
5. | Bipolar Hemostasis Catheters. Injection Gold Probe™ and Gold Probe™. Available from: https://www.bostonscientific.com/en-US/products/probes/injection-gold-probe.html. [Cited in This Article: ] |
6. | Cook DJ, Guyatt GH, Salena BJ, Laine LA. Endoscopic therapy for acute nonvariceal upper gastrointestinal hemorrhage: a meta-analysis. Gastroenterology. 1992;102:139-148. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 501] [Cited by in F6Publishing: 491] [Article Influence: 15.3] [Reference Citation Analysis (0)] |
7. | Chung SC, Leung JW, Sung JY, Lo KK, Li AK. Injection or heat probe for bleeding ulcer. Gastroenterology. 1991;100:33-37. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 119] [Cited by in F6Publishing: 104] [Article Influence: 3.2] [Reference Citation Analysis (0)] |
8. | Jensen DM. Diagnosis and treatment of definitive diverticular hemorrhage (DDH). Am J Gastroenterol. 2018;113:1570-1573. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 21] [Cited by in F6Publishing: 30] [Article Influence: 5.0] [Reference Citation Analysis (0)] |
9. | Sengupta N, Feuerstein JD, Jairath V, Shergill AK, Strate LL, Wong RJ, Wan D. Management of Patients With Acute Lower Gastrointestinal Bleeding: An Updated ACG Guideline. Am J Gastroenterol. 2023;118:208-231. [PubMed] [DOI] [Cited in This Article: ] [Cited by in F6Publishing: 36] [Reference Citation Analysis (0)] |
10. | Wedi E, von Renteln D, Gonzalez S, Tkachenko O, Jung C, Orkut S, Roth V, Tumay S, Hochberger J. Use of the over-the-scope-clip (OTSC) in non-variceal upper gastrointestinal bleeding in patients with severe cardiovascular comorbidities: a retrospective study. Endosc Int Open. 2017;5:E875-E882. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 23] [Cited by in F6Publishing: 25] [Article Influence: 3.6] [Reference Citation Analysis (0)] |
11. | Chan SM, Chiu PW, Teoh AY, Lau JY. Use of the Over-The-Scope Clip for treatment of refractory upper gastrointestinal bleeding: a case series. Endoscopy. 2014;46:428-431. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 73] [Cited by in F6Publishing: 81] [Article Influence: 8.1] [Reference Citation Analysis (0)] |
12. | Asokkumar R, Soetikno R, Sanchez-Yague A, Kim Wei L, Salazar E, Ngu JH. Use of over-the-scope-clip (OTSC) improves outcomes of high-risk adverse outcome (HR-AO) non-variceal upper gastrointestinal bleeding (NVUGIB). Endosc Int Open. 2018;6:E789-E796. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 18] [Cited by in F6Publishing: 19] [Article Influence: 3.2] [Reference Citation Analysis (0)] |
13. | Yang E, Chang MA, Savides TJ. New Techniques to Control Gastrointestinal Bleeding. Gastroenterol Hepatol (N Y). 2019;15:471-479. [PubMed] [Cited in This Article: ] |
14. | Kwek BEA, Ang TL, Ong PLJ, Tan YLJ, Ang SWD, Law NM, Thurairajah PH, Fock KM. TC-325 versus the conventional combined technique for endoscopic treatment of peptic ulcers with high-risk bleeding stigmata: A randomized pilot study. J Dig Dis. 2017;18:323-329. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 18] [Cited by in F6Publishing: 22] [Article Influence: 3.1] [Reference Citation Analysis (0)] |
15. | Barkun AN, Almadi M, Kuipers EJ, Laine L, Sung J, Tse F, Leontiadis GI, Abraham NS, Calvet X, Chan FKL, Douketis J, Enns R, Gralnek IM, Jairath V, Jensen D, Lau J, Lip GYH, Loffroy R, Maluf-Filho F, Meltzer AC, Reddy N, Saltzman JR, Marshall JK, Bardou M. Management of Nonvariceal Upper Gastrointestinal Bleeding: Guideline Recommendations From the International Consensus Group. Ann Intern Med. 2019;171:805-822. [PubMed] [DOI] [Cited in This Article: ] [Cited by in Crossref: 209] [Cited by in F6Publishing: 281] [Article Influence: 56.2] [Reference Citation Analysis (16)] |